Experimental verification of clinical dose and LET maps for LET-informed proton therapy
Abstract Objective: Linear energy transfer (LET)-informed proton therapy has the potential to improve biological treatment planning beyond the conventional, constant-RBE assumption. Its clinical translation, however, requires experimental tools capable of verifying spatially resolved dose and dose-averaged LET (LET D ) distributions under clinically relevant conditions. We recently introduced an optically stimulated luminescence (OSL) film-based detector system that simultaneously measures two-dimensional (2D) absorbed dose and LET D . This work provides a first assessment of the system for verifying dose and LET D distributions from clinical proton therapy plans. Approach: The system was tested at two proton therapy centers, ranging from simple uniform fields to more complex geometries, including Bragg curve measurements and a clinically realistic anthropomorphic head phantom treatment plan. Measured dose and LET D distributions were compared against commercial treatment planning system (TPS) calculations and, for selected fields, to external Monte Carlo particle transport simulations due to the absence of traceability of LET D . For non-uniform irradiations, agreement was assessed using gamma analysis with 3%/2 mm and 5%/2 mm criteria for dose and LET D , respectively. Main results: For uniform fields, measured dose and LET D distributions generally agreed with TPS and independent Monte Carlo simulations across a range of clinically relevant irradiation conditions. In the Bragg curve measurement, gamma pass rates were 96.6% for dose and 99.7% for LET D , with the largest dose discrepancy observed near the Bragg peak, consistent with a current over-correction of detector quenching in high-LET regions to be improved in future work. In the anthropomorphic phantom, gamma pass rates ranged from 75% to 100% for dose and from 45% to 95% for LET D , with a single layer showing a sub-50% pass rate for LET D . Failing regions were confined to high-dose superior layers for dose and distal-edge, high-LET regions for LET D . Significance: The presented results support the detector system’s use as an experimental benchmarking tool for TPS-calculated dose and LET D maps. Given this and the system’s ability to simultaneously measure 2D dose and LET D maps with high spatial resolution and minimal field perturbation, the system shows promise for future commissioning workflows for LET-informed proton therapy.
Authors
- Jeppe Brage Christensen (ORCID: https://orcid.org/0000-0002-6894-381X)
- E.G. Yukihara (ORCID: https://orcid.org/0000-0002-4615-6698)
- Michelle van Heerden
- Liliana Stolarczyk (ORCID: https://orcid.org/0000-0002-6333-3350)
- Artūrs Meijers (ORCID: https://orcid.org/0000-0002-3647-6647)
- Niels Bassler (ORCID: https://orcid.org/0000-0002-4160-1078)
- Michele Togno (ORCID: https://orcid.org/0000-0001-6840-814X)
- Anne Vestergaard (ORCID: https://orcid.org/0000-0003-1398-6377)
- Mads L. Jensen (ORCID: https://orcid.org/0000-0001-9534-9034)
- Francesca Albertini (ORCID: https://orcid.org/0000-0002-2547-1269)
Institutions
- Paul Scherrer Institute (CH)
- Aarhus University Hospital (DK)
Publication Details
- Journal
- Physics in Medicine and Biology
- Published
- 2026-11-06
- DOI
- https://doi.org/10.1088/1361-6560/aebf1d
- Primary Topic
- Radiation Therapy and Dosimetry
- Type
- article
- Field-Weighted Citation Impact
- 0.00