Comparative diagnostic performance of digital and conventional tests for congenital red–green colour vision deficiency

PURPOSE: To compare the diagnostic performance of the DIVE Colour Test for detecting congenital red-green (RG) colour vision deficiency, using the Colour Assessment and Diagnosis (CAD) test as the reference standard, with established conventional colour vision tests providing within-sample benchmarks. METHODS: Sixty participants were enrolled in an observational study: 30 CAD-confirmed normal trichromats and 30 participants with congenital RG deficiency. Each completed the CAD test, DIVE Colour Test (touch and eye-tracking), Nagel anomaloscope, Ishihara 38 Plates Test and Roth 28-Hue Test. Diagnostic accuracy against CAD was assessed using sensitivity, specificity and Cohen's kappa (k), both in right-eye and binocular testing. RESULTS: In right-eye testing, DIVE Touch achieved a sensitivity of 78.1% and a specificity of 100.0%, demonstrating substantial agreement with CAD (κ = 0.77). In comparison, DIVE eye-tracking showed lower sensitivity (53.1%) while maintaining high specificity (96.4%), yielding moderate agreement with CAD (κ = 0.48). Regarding the binocular analysis: DIVE Touch yielded a sensitivity of 69.0%, a specificity of 100.0% and substantial agreement with CAD (κ = 0.69). Among conventional tests versus CAD, Ishihara showed the highest sensitivity with complete specificity, the Nagel anomaloscope achieved substantial agreement, and the Roth 28-Hue Test showed the lowest sensitivity for RG classification. CONCLUSION: Diagnostic performance varied across colour vision tests. DIVE Touch showed high specificity, moderate sensitivity and substantial agreement with CAD, whereas DIVE Eye-Tracking maintained high specificity but showed lower sensitivity and requires further optimisation.

Authors

Institutions

Publication Details

Journal
Acta Ophthalmologica
Published
2026-11-06
DOI
https://doi.org/10.1111/aos.70265
Primary Topic
Ophthalmology and Visual Impairment Studies
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

Comparative diagnostic performance of digital and conventional tests for congenital red–green colour vision deficiency

Amelia Nieto-Bona, Jesús Carballo, Eduardo Esteban-Ibañez, Marina Vilella et al.
1 links
Acta Ophthalmologica
Ophthalmology and Visual Impairment Studies
article

Comparative diagnostic performance of digital and conventional tests for congenital red–green colour vision deficiency

Amelia Nieto-Bona, Jesús Carballo, Eduardo Esteban-Ibañez, Marina Vilella, Victor I. Fernandez‐Fernandez
article en

Abstract

PURPOSE: To compare the diagnostic performance of the DIVE Colour Test for detecting congenital red-green (RG) colour vision deficiency, using the Colour Assessment and Diagnosis (CAD) test as the reference standard, with established conventional colour vision tests providing within-sample benchmarks. METHODS: Sixty participants were enrolled in an observational study: 30 CAD-confirmed normal trichromats and 30 participants with congenital RG deficiency. Each completed the CAD test, DIVE Colour Test (touch and eye-tracking), Nagel anomaloscope, Ishihara 38 Plates Test and Roth 28-Hue Test. Diagnostic accuracy against CAD was assessed using sensitivity, specificity and Cohen's kappa (k), both in right-eye and binocular testing. RESULTS: In right-eye testing, DIVE Touch achieved a sensitivity of 78.1% and a specificity of 100.0%, demonstrating substantial agreement with CAD (κ = 0.77). In comparison, DIVE eye-tracking showed lower sensitivity (53.1%) while maintaining high specificity (96.4%), yielding moderate agreement with CAD (κ = 0.48). Regarding the binocular analysis: DIVE Touch yielded a sensitivity of 69.0%, a specificity of 100.0% and substantial agreement with CAD (κ = 0.69). Among conventional tests versus CAD, Ishihara showed the highest sensitivity with complete specificity, the Nagel anomaloscope achieved substantial agreement, and the Roth 28-Hue Test showed the lowest sensitivity for RG classification. CONCLUSION: Diagnostic performance varied across colour vision tests. DIVE Touch showed high specificity, moderate sensitivity and substantial agreement with CAD, whereas DIVE Eye-Tracking maintained high specificity but showed lower sensitivity and requires further optimisation.

Acta Ophthalmologica
Universidad Complutense de Madrid (ES), Instituto de Salud Carlos III (ES), Hospital Universitario Miguel Servet (ES), Instituto de Investigación Sanitaria Aragón (ES)
European Commission, Universidad de Zaragoza, Instituto de Salud Carlos III, Agencia Estatal de Investigación
Good health and well-being
Openalex Percentile: Top 10%
Ophthalmology and Visual Impairment Studies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon