Experimental validation of a newly developed neutron ambient dose H* detector in an accelerator-based BNCT field

Abstract Out-of-field neutron dose evaluation is important for radiation protection during boron neutron capture therapy (BNCT). However, no commercial detector is available for measuring the neutron ambient dose equivalent H* outside the irradiation field. Establishing a reliable H* evaluation method is essential for the international standardization of BNCT. In this study, a prototype neutron detector was evaluated using a cyclotron-based epithermal neutron source (C-BENS). The ratio of the measured to calculated H*, based on ICRP Publication 116, ranged from 0.21–0.56 at lateral distances of 10–150 cm from the beam aperture center. This discrepancy is likely due to the difficulty in accurately reproducing the experimental geometry in the Monte Carlo simulation used to calculate the reference H*. These results provide a basis for improving H* evaluation methods and developing standardized radiation protection procedures for BNCT.

Authors

Institutions

Publication Details

Journal
Japanese Journal of Applied Physics
Published
2026-11-06
DOI
https://doi.org/10.35848/1347-4065/aebf3d
Primary Topic
Radiation Therapy and Dosimetry
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

Experimental validation of a newly developed neutron ambient dose H* detector in an accelerator-based BNCT field

Naonori Hu, Shinji Kawabata, Akihiro Yamaji, Akihiko Masuda et al.
1 links
Japanese Journal of Applied Physics
Radiation Therapy and Dosimetry
article

Experimental validation of a newly developed neutron ambient dose H* detector in an accelerator-based BNCT field

Naonori Hu, Shinji Kawabata, Akihiro Yamaji, Akihiko Masuda, Takushi Takata, Tetsuro MATSUMOTO, Shunsuke Kurosawa, Hiroki Tanaka, Liang Zhao
article en

Abstract

Abstract Out-of-field neutron dose evaluation is important for radiation protection during boron neutron capture therapy (BNCT). However, no commercial detector is available for measuring the neutron ambient dose equivalent H* outside the irradiation field. Establishing a reliable H* evaluation method is essential for the international standardization of BNCT. In this study, a prototype neutron detector was evaluated using a cyclotron-based epithermal neutron source (C-BENS). The ratio of the measured to calculated H*, based on ICRP Publication 116, ranged from 0.21–0.56 at lateral distances of 10–150 cm from the beam aperture center. This discrepancy is likely due to the difficulty in accurately reproducing the experimental geometry in the Monte Carlo simulation used to calculate the reference H*. These results provide a basis for improving H* evaluation methods and developing standardized radiation protection procedures for BNCT.

Japanese Journal of Applied Physics
Tohoku University (JP), Osaka Medical and Pharmaceutical University (JP), Kyoto University (JP), Institute for Integrated Radiation and Nuclear Science, Kyoto University (JP), Nagoya University (JP), National Institute of Advanced Industrial Science and Technology (JP)
Japan Society for the Promotion of Science
Good health and well-being
Openalex Percentile: Top 16%
Radiation Therapy and Dosimetry
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon