Digital handgrip assessment in musculoskeletal health: reliability, measurement precision, interchangeability, and clinical implications for rehabilitation and sports medicine

Abstract Handgrip strength is a widely used indicator of upper extremity function and a critical field-based tool in sports medicine for injury screening, rehabilitation monitoring, and return-to-sport decision-making. Despite the growing application of digital dynamometers, comparative data on their measurement precision and interchangeability remain limited. This study aimed to assess the test-retest reliability, measurement precision, and inter-device agreement of three digital handgrip dynamometers: GripAble, K-Force Grip, and JTech Commander Echo Grip. Twenty-five healthy adults (15 females, 10 males; mean age 36.2 ± 11.8 years) with no history of upper extremity injury participated in two identical testing sessions held 7 to 18 days apart. Intraclass correlation coefficients (ICC), standard error of measurement (SEM), and minimal detectable change (MDC 95 ) were calculated to evaluate reliability and measurement precision. Inter-device agreement and interchangeability were analyzed using repeated-measures analyses, fixed and proportional bias assessment, and Bland-Altman limits of agreement. All three devices demonstrated excellent test-retest reliability (ICC = 0.88–0.95). K-Force showed the lowest SEM (1.36–1.63 kg) and MDC 95 values, indicating greater measurement precision and sensitivity to change. Inter-device agreement was good (ICC = 0.851–0.873); however, GripAble consistently produced substantially lower force values than JTech and K-Force (≈ 26.37–30.47% bias), with wide limits of agreement and evidence of both fixed and proportional bias. JTech and K-Force showed comparatively narrower limits of agreement and minimal bias, suggesting closer relative measurement alignment between these two devices, although the magnitude of these limits remains clinically non-negligible in the absence of predefined acceptance criteria. Although all three digital dynamometers demonstrated high test-retest reliability, the observed systematic differences in absolute force values suggest limited interchangeability between devices. We recommend that clinicians and researchers use the same device consistently across longitudinal assessments. The device-specific MDC values reported here, ranging from 3.77 to 5.87 kg, provide useful benchmarks for interpreting measurement change in healthy adults. As the current sample consisted of healthy adults, further studies are needed to establish device applicability in athletic and clinical populations.

Authors

Institutions

Publication Details

Journal
BMC Sports Science Medicine and Rehabilitation
Published
2026-11-06
DOI
https://doi.org/10.1186/s13102-026-02225-3
Primary Topic
Orthopedic Surgery and Rehabilitation
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

Digital handgrip assessment in musculoskeletal health: reliability, measurement precision, interchangeability, and clinical implications for rehabilitation and sports medicine

İrem Akgün, Corey Weston McGee, Halil İbrahim Ergen, Adriana Torres-Pardo et al.
4 links
BMC Sports Science Medicine and Rehabilitation
Orthopedic Surgery and Rehabilitation
article

Digital handgrip assessment in musculoskeletal health: reliability, measurement precision, interchangeability, and clinical implications for rehabilitation and sports medicine

İrem Akgün, Corey Weston McGee, Halil İbrahim Ergen, Adriana Torres-Pardo, Zarife Pancar
article en

Abstract

Abstract Handgrip strength is a widely used indicator of upper extremity function and a critical field-based tool in sports medicine for injury screening, rehabilitation monitoring, and return-to-sport decision-making. Despite the growing application of digital dynamometers, comparative data on their measurement precision and interchangeability remain limited. This study aimed to assess the test-retest reliability, measurement precision, and inter-device agreement of three digital handgrip dynamometers: GripAble, K-Force Grip, and JTech Commander Echo Grip. Twenty-five healthy adults (15 females, 10 males; mean age 36.2 ± 11.8 years) with no history of upper extremity injury participated in two identical testing sessions held 7 to 18 days apart. Intraclass correlation coefficients (ICC), standard error of measurement (SEM), and minimal detectable change (MDC 95 ) were calculated to evaluate reliability and measurement precision. Inter-device agreement and interchangeability were analyzed using repeated-measures analyses, fixed and proportional bias assessment, and Bland-Altman limits of agreement. All three devices demonstrated excellent test-retest reliability (ICC = 0.88–0.95). K-Force showed the lowest SEM (1.36–1.63 kg) and MDC 95 values, indicating greater measurement precision and sensitivity to change. Inter-device agreement was good (ICC = 0.851–0.873); however, GripAble consistently produced substantially lower force values than JTech and K-Force (≈ 26.37–30.47% bias), with wide limits of agreement and evidence of both fixed and proportional bias. JTech and K-Force showed comparatively narrower limits of agreement and minimal bias, suggesting closer relative measurement alignment between these two devices, although the magnitude of these limits remains clinically non-negligible in the absence of predefined acceptance criteria. Although all three digital dynamometers demonstrated high test-retest reliability, the observed systematic differences in absolute force values suggest limited interchangeability between devices. We recommend that clinicians and researchers use the same device consistently across longitudinal assessments. The device-specific MDC values reported here, ranging from 3.77 to 5.87 kg, provide useful benchmarks for interpreting measurement change in healthy adults. As the current sample consisted of healthy adults, further studies are needed to establish device applicability in athletic and clinical populations.

BMC Sports Science Medicine and Rehabilitation
University of Minnesota (US), Centre for Automation and Robotics (ES), Gaziantep University (TR), Universidad Politécnica de Madrid (ES)
Good health and well-being
Openalex Percentile: Top 12%
Orthopedic Surgery and Rehabilitation
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon