Single-use versus Reusable flexible rhinolaryngoscopes: A retrospective cohort study

Objectives To describe the real-world clinical application of a single-use flexible rhinolaryngoscope in routine otolaryngology practice. Methods This retrospective cohort study included consecutive patients who underwent flexible rhinolaryngoscopy at a tertiary medical center between August 1 and December 1, 2024. Patients were examined using either a Single-use Flexible Rhinolaryngoscope (SFR) or a Reusable Flexible Rhinolaryngoscope (RFR) as part of standard clinical care. All procedures followed institutional protocols. Outcome measures included examination time, treatment time, operator-rated procedural performance (insertion, rotation, withdrawal, and instrument compatibility), subjective image quality, and procedure-related adverse events. Results A total of 426 patients were analyzed, including 219 in the SFR group and 207 in the RFR group. Baseline demographic and clinical characteristics were comparable between groups. Mean examination time (SFR: 2.20 ± 1.22 min vs. RFR: 2.18 ± 1.02 min, p = 0.431) and treatment time (SFR: 5.80 ± 3.96 min vs. RFR: 5.35 ± 5.18 min, p = 0.639) did not differ significantly. There was no difference in operator ratings between the two devices regarding insertion, rotation, withdrawal, and compatibility with accessory instruments. Image quality scores were lower for SFR compared with RFR (4.26 ± 0.76 vs. 4.82 ± 0.39, p = 0.001). The SFR was successfully used across a wide range of diagnostic and therapeutic indications, with no device-related adverse events observed. Conclusion This single-center retrospective study describes the use of a single-use flexible rhinolaryngoscope in routine otolaryngology practice, with procedures completed without device-related interruptions or adverse events. Level of Evidence Level 3.

Authors

Institutions

Publication Details

Journal
Brazilian Journal of Otorhinolaryngology
Published
2026-11-04
DOI
https://doi.org/10.1016/j.bjorl.2026.101924
Primary Topic
Medical Device Sterilization and Disinfection
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Single-use versus Reusable flexible rhinolaryngoscopes: A retrospective cohort study

Ying Zhang, Yang, Qian, 1973-, Haibo Ou, Bingbing Zhang et al.
Brazilian Journal of Otorhinolaryngology
Medical Device Sterilization and Disinfection
article

Single-use versus Reusable flexible rhinolaryngoscopes: A retrospective cohort study

Ying Zhang, Yang, Qian, 1973-, Haibo Ou, Bingbing Zhang, Ou Peng, Yekai Feng, Shisheng Li
article en

Abstract

Objectives To describe the real-world clinical application of a single-use flexible rhinolaryngoscope in routine otolaryngology practice. Methods This retrospective cohort study included consecutive patients who underwent flexible rhinolaryngoscopy at a tertiary medical center between August 1 and December 1, 2024. Patients were examined using either a Single-use Flexible Rhinolaryngoscope (SFR) or a Reusable Flexible Rhinolaryngoscope (RFR) as part of standard clinical care. All procedures followed institutional protocols. Outcome measures included examination time, treatment time, operator-rated procedural performance (insertion, rotation, withdrawal, and instrument compatibility), subjective image quality, and procedure-related adverse events. Results A total of 426 patients were analyzed, including 219 in the SFR group and 207 in the RFR group. Baseline demographic and clinical characteristics were comparable between groups. Mean examination time (SFR: 2.20 ± 1.22 min vs. RFR: 2.18 ± 1.02 min, p = 0.431) and treatment time (SFR: 5.80 ± 3.96 min vs. RFR: 5.35 ± 5.18 min, p = 0.639) did not differ significantly. There was no difference in operator ratings between the two devices regarding insertion, rotation, withdrawal, and compatibility with accessory instruments. Image quality scores were lower for SFR compared with RFR (4.26 ± 0.76 vs. 4.82 ± 0.39, p = 0.001). The SFR was successfully used across a wide range of diagnostic and therapeutic indications, with no device-related adverse events observed. Conclusion This single-center retrospective study describes the use of a single-use flexible rhinolaryngoscope in routine otolaryngology practice, with procedures completed without device-related interruptions or adverse events. Level of Evidence Level 3.

Brazilian Journal of OtorhinolaryngologyVol. 93(1)
Central South University (CN), Second Xiangya Hospital of Central South University (CN), Xiangya Hospital Central South University (CN)
Openalex Percentile: Top 16%
Medical Device Sterilization and Disinfection
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon