Separating Motion Artifacts from Biomarker Signals in Flexible MoS2 FET Cortisol Sensors: A Coupled Multiphysics Model and Finite-Element Analysis of Strain-Binding Cross-Talk and Mitigation Strategies

Abstract Flexible molybdenum disulfide (MoS2) field-effect transistor (FET) biosensors can monitor cortisol in sweat, but their atomically thin channel is strongly piezoresistive, so body movement produces current changes that resemble hormone signals. We present a six-stage multiphysics model that links wrist and elbow flexion and walking to a strain-equivalent false cortisol concentration. Channel strain is computed by plane-strain finite-element analysis in the open-source solver Elmer, verified against laminate beam theory, and drives compact models of piezoresistance, Schottky-barrier lowering, strain-binding cross-talk, aptamer kinetics and Debye screening. For a 25 µm polyimide patch the model predicts wrist flexion produces a current artifact of 44% of the full-scale cortisol response and walking 5 to 11 ng/mL of false cortisol, with about 70% of the artifact arising at the Schottky contacts. Low-pass filtering leaves the non-zero mean of unipolar flexion (22% of full scale). Twin-sensor subtraction is limited by gauge mismatch and by a cross-talk floor proportional to aptamer occupancy, 1.0 to 2.8% of full scale across the physiological range at the wrist. With a 2.5 µm substrate and matched twin channels the residual falls to 0.9 to 1.3 ng/mL, not consistently below a 1 ng/mL target at the wrist for the assumed affinity. A rigid island under the channel retains 18 to 109% of the baseline strain under end-moment loading and about 300% when the skin imposes the curvature, so it fails to isolate the channel in bending. The predicted magnitudes scale with the assumed threshold shift; design guidelines by body location are given.

Authors

Institutions

Publication Details

Journal
Research Square
Published
2026-11-04
DOI
https://doi.org/10.21203/rs.3.rs-11606589/v1
Primary Topic
Advanced Sensor and Energy Harvesting Materials
Type
preprint
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
preprint

Separating Motion Artifacts from Biomarker Signals in Flexible MoS2 FET Cortisol Sensors: A Coupled Multiphysics Model and Finite-Element Analysis of Strain-Binding Cross-Talk and Mitigation Strategies

Ayomide Adeyemi, Victor Ogunyemi
Research Square
Advanced Sensor and Energy Harvesting Materials
preprint

Separating Motion Artifacts from Biomarker Signals in Flexible MoS2 FET Cortisol Sensors: A Coupled Multiphysics Model and Finite-Element Analysis of Strain-Binding Cross-Talk and Mitigation Strategies

Ayomide Adeyemi, Victor Ogunyemi
preprint en

Abstract

Abstract Flexible molybdenum disulfide (MoS2) field-effect transistor (FET) biosensors can monitor cortisol in sweat, but their atomically thin channel is strongly piezoresistive, so body movement produces current changes that resemble hormone signals. We present a six-stage multiphysics model that links wrist and elbow flexion and walking to a strain-equivalent false cortisol concentration. Channel strain is computed by plane-strain finite-element analysis in the open-source solver Elmer, verified against laminate beam theory, and drives compact models of piezoresistance, Schottky-barrier lowering, strain-binding cross-talk, aptamer kinetics and Debye screening. For a 25 µm polyimide patch the model predicts wrist flexion produces a current artifact of 44% of the full-scale cortisol response and walking 5 to 11 ng/mL of false cortisol, with about 70% of the artifact arising at the Schottky contacts. Low-pass filtering leaves the non-zero mean of unipolar flexion (22% of full scale). Twin-sensor subtraction is limited by gauge mismatch and by a cross-talk floor proportional to aptamer occupancy, 1.0 to 2.8% of full scale across the physiological range at the wrist. With a 2.5 µm substrate and matched twin channels the residual falls to 0.9 to 1.3 ng/mL, not consistently below a 1 ng/mL target at the wrist for the assumed affinity. A rigid island under the channel retains 18 to 109% of the baseline strain under end-moment loading and about 300% when the skin imposes the curvature, so it fails to isolate the channel in bending. The predicted magnitudes scale with the assumed threshold shift; design guidelines by body location are given.

Research Square
Olabisi Onabanjo University (NG), Kwara State University (NG)
Advanced Sensor and Energy Harvesting Materials
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon
Separating Motion Artifacts from Biomarker Signals in Flexible MoS2 FET Cortisol Sensors: A Coupled Multiphysics Model and Finite-Element Analysis of Strain-Binding Cross-Talk and Mitigation Strategies — Ayomide Adeyemi, Victor Ogunyemi · Research Square (2026) | TGRS Research Map | TGRS