Optimization of Leukocytic Flags in the Sysmex‐ XN Series With the Help of Flow Cytometric Differential ( FlowDiff ) Can Significantly Optimize Laboratory Efficiency

ABSTRACT Objectives The aim of this study was to adjust the default white blood cell (WBC) differential count flag settings on the Sysmex XN‐9100 haematology analyser to reduce unnecessary additional leukocyte differential counts without compromising diagnostic sensitivity. Methods Data from 4554 routine blood samples flagged on Sysmex XN‐9100 and corresponding additional leukocyte differential counts performed on the Sysmex XF‐1600 flow cytometer were used to optimise the WBC flagging rules. Results A total of 1148 samples had scattergram‐related flags, which are non‐adjustable. Optimal cut points for Q‐flags were established based on Youden's Index for the remaining 3406 samples, while the diagnostic sensitivity of the new flagging rules for acute leukaemia was assessed in a validation cohort. The optimal cut points for Q‐flags were 120 for ‘Atypical Lymphocytes’, 170 for ‘Abnormal Lymphocytes’ and 160 for ‘Blasts’, while the Q‐flag ‘Left shift’ offered no diagnostic value in our cohort. The count flags ‘eosinophilia’, ‘basophilia’ and ‘monocytopenia’ were deactivated, as the results from Sysmex XN were comparable to the flow cytometric reference method. By implementing the new flagging rules, 1709 (37.5%) of the samples would have been processed without need for additional leukocyte differential count. When evaluated in a separate validation cohort of 90 samples from patients with newly diagnosed acute leukaemia, the new rules showed similar diagnostic sensitivity to the default flagging system. Conclusion Optimisation of Sysmex flagging rules for leukocyte differential counts was feasible and significantly reduced unnecessary reflex testing, while maintaining analytical accuracy and diagnostic safety.

Authors

Institutions

Publication Details

Journal
International Journal of Laboratory Hematology
Published
2026-11-03
DOI
https://doi.org/10.1111/ijlh.70275
Primary Topic
Clinical Laboratory Practices and Quality Control
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Optimization of Leukocytic Flags in the Sysmex‐ XN Series With the Help of Flow Cytometric Differential ( FlowDiff ) Can Significantly Optimize Laboratory Efficiency

Konstantinos Dimopoulos, Ninna Hahn Tougaard, Anne Winther‐Larsen, Anders Mønsted Abildgaard et al.
International Journal of Laboratory Hematology
Clinical Laboratory Practices and Quality Control
article

Optimization of Leukocytic Flags in the Sysmex‐ XN Series With the Help of Flow Cytometric Differential ( FlowDiff ) Can Significantly Optimize Laboratory Efficiency

Konstantinos Dimopoulos, Ninna Hahn Tougaard, Anne Winther‐Larsen, Anders Mønsted Abildgaard, Snorre Kristian Vesten Pedersen
article en

Abstract

ABSTRACT Objectives The aim of this study was to adjust the default white blood cell (WBC) differential count flag settings on the Sysmex XN‐9100 haematology analyser to reduce unnecessary additional leukocyte differential counts without compromising diagnostic sensitivity. Methods Data from 4554 routine blood samples flagged on Sysmex XN‐9100 and corresponding additional leukocyte differential counts performed on the Sysmex XF‐1600 flow cytometer were used to optimise the WBC flagging rules. Results A total of 1148 samples had scattergram‐related flags, which are non‐adjustable. Optimal cut points for Q‐flags were established based on Youden's Index for the remaining 3406 samples, while the diagnostic sensitivity of the new flagging rules for acute leukaemia was assessed in a validation cohort. The optimal cut points for Q‐flags were 120 for ‘Atypical Lymphocytes’, 170 for ‘Abnormal Lymphocytes’ and 160 for ‘Blasts’, while the Q‐flag ‘Left shift’ offered no diagnostic value in our cohort. The count flags ‘eosinophilia’, ‘basophilia’ and ‘monocytopenia’ were deactivated, as the results from Sysmex XN were comparable to the flow cytometric reference method. By implementing the new flagging rules, 1709 (37.5%) of the samples would have been processed without need for additional leukocyte differential count. When evaluated in a separate validation cohort of 90 samples from patients with newly diagnosed acute leukaemia, the new rules showed similar diagnostic sensitivity to the default flagging system. Conclusion Optimisation of Sysmex flagging rules for leukocyte differential counts was feasible and significantly reduced unnecessary reflex testing, while maintaining analytical accuracy and diagnostic safety.

International Journal of Laboratory Hematology
University of Copenhagen (DK), Aarhus University (DK), Frederiksberg Hospital (DK), Bispebjerg Hospital (DK), Aarhus University Hospital (DK)
Openalex Percentile: Top 14%
Clinical Laboratory Practices and Quality Control
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon