Feasibility of real-time 1.5-T MR-guided percutaneous irreversible electroporation in a porcine model

Abstract This exploratory study assessed the procedural feasibility of in‑bore, real‑time magnetic resonance (MR)‑guided percutaneous irreversible electroporation (IRE) using a clinical 1.5-T system in a porcine model. Three healthy anesthetized juvenile swine underwent IRE in normal renal tissue using a prototype pulse generator and MR-conditional coaxial cannula as electrodes. One swine was ablated twice in the same kidney, resulting in statistically non-independent data. The procedure was guided by a balanced steady-state free precession sequence with orthogonal slices and intermittent position verifications using a radial three-dimensional gradient-echo sequence. Preprocedure radiofrequency noise and signal-to-noise ratio measurements showed slightly increased noise levels in generator standby mode without apparent imaging artifacts. Median procedure time was 48 min with an interquartile range of 17 min. Four hours post-procedure, the treatment zone appeared T1-hypointense in delayed contrast-enhanced imaging. For one treatment zone, hematoxylin and eosin staining (H&E), fluorescent terminal deoxynucleotidyl transferase-mediated dUTP nick-end labeling staining (TUNEL), and coregistration to MR imaging were performed. H&E showed minor hemorrhaging at the site of electrode insertion. TUNEL demonstrated a strong TUNEL-positive signal correlating with the T1-hypointense signal in delayed contrast-enhanced imaging, with a Dice-Sørensen coefficient of 67%. These findings demonstrate the procedural feasibility of MR-guided percutaneous IRE in clinical 1.5-T MR systems. Key Points Question Can MR-guided percutaneous irreversible electroporation be performed in a porcine model despite the lack of MR-compliant devices? Findings In four irreversible electroporation procedures in three juvenile pigs, electrodes were successfully placed in the kidney using real-time guidance in a clinical 1.5-T scanner. Relevance statement Real-time MR-guided percutaneous irreversible electroporation is procedurally feasible in a clinical 1.5-T MRI environment, while preserving image quality and enabling arbitrary slice positioning. Further research is required to confirm the reproducibility and enable the translation to oncological applications.

Authors

Publication Details

Journal
European Radiology Experimental
Published
2026-11-03
DOI
https://doi.org/10.1186/s41747-026-00838-4
Primary Topic
Microbial Inactivation Methods
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Feasibility of real-time 1.5-T MR-guided percutaneous irreversible electroporation in a porcine model

Philipp Felgendreff, Marcel Eisenmann, Julian Glandorf, Max Joris Hubmann et al.
European Radiology Experimental
Microbial Inactivation Methods
article

Feasibility of real-time 1.5-T MR-guided percutaneous irreversible electroporation in a porcine model

Philipp Felgendreff, Marcel Eisenmann, Julian Glandorf, Max Joris Hubmann, Jörg Mengwasser, Sahar Arbabzadah, Frank Wacker, Thomas H. Gerlach, Bennet Hensen, Simon Schröer, Othmar Belker, Marcel Gutberlet, Seben Sena Yerdelen, Georg Rose, Björn Hartleben, Inga Bruesch, Christoph Fritsch, Daniel Düx
article en

Abstract

Abstract This exploratory study assessed the procedural feasibility of in‑bore, real‑time magnetic resonance (MR)‑guided percutaneous irreversible electroporation (IRE) using a clinical 1.5-T system in a porcine model. Three healthy anesthetized juvenile swine underwent IRE in normal renal tissue using a prototype pulse generator and MR-conditional coaxial cannula as electrodes. One swine was ablated twice in the same kidney, resulting in statistically non-independent data. The procedure was guided by a balanced steady-state free precession sequence with orthogonal slices and intermittent position verifications using a radial three-dimensional gradient-echo sequence. Preprocedure radiofrequency noise and signal-to-noise ratio measurements showed slightly increased noise levels in generator standby mode without apparent imaging artifacts. Median procedure time was 48 min with an interquartile range of 17 min. Four hours post-procedure, the treatment zone appeared T1-hypointense in delayed contrast-enhanced imaging. For one treatment zone, hematoxylin and eosin staining (H&E), fluorescent terminal deoxynucleotidyl transferase-mediated dUTP nick-end labeling staining (TUNEL), and coregistration to MR imaging were performed. H&E showed minor hemorrhaging at the site of electrode insertion. TUNEL demonstrated a strong TUNEL-positive signal correlating with the T1-hypointense signal in delayed contrast-enhanced imaging, with a Dice-Sørensen coefficient of 67%. These findings demonstrate the procedural feasibility of MR-guided percutaneous IRE in clinical 1.5-T MR systems. Key Points Question Can MR-guided percutaneous irreversible electroporation be performed in a porcine model despite the lack of MR-compliant devices? Findings In four irreversible electroporation procedures in three juvenile pigs, electrodes were successfully placed in the kidney using real-time guidance in a clinical 1.5-T scanner. Relevance statement Real-time MR-guided percutaneous irreversible electroporation is procedurally feasible in a clinical 1.5-T MRI environment, while preserving image quality and enabling arbitrary slice positioning. Further research is required to confirm the reproducibility and enable the translation to oncological applications.

European Radiology ExperimentalVol. 10(1)
Openalex Percentile: Top 20%
Microbial Inactivation Methods
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon