Identifying sequencing artifacts in low-diversity, short-insert aptamer libraries on the Illumina MiSeq platform

∼1% in other samples) which resulted in overestimation of sample concentration and, in turn, decreased sequencing yield. Sequencing runs displayed highly variable pass rates (6-93%), likely due to short sequences outcompeting longer ones and assignment errors of neighboring clusters with identical early-cycle bases. Unexpected duplicates in template samples suggest that adapter-ligated daughter strands and optical misidentifications contribute to artificial duplication. These duplication artifacts are of particular importance in amplicon sequencing for which copy-number analysis is crucial. These findings highlight the need for refined ligation conditions, improved purification strategies, and enhanced bioinformatic filtering to ensure sequencing accuracy in aptamer screening library workflows on the Illumina platform.

Authors

Institutions

Publication Details

Journal
The Analyst
Published
2026-11-23
DOI
https://doi.org/10.1039/d6an01029f
Primary Topic
Genomics and Phylogenetic Studies
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

Identifying sequencing artifacts in low-diversity, short-insert aptamer libraries on the Illumina MiSeq platform

Anton V. Bryksin, Valeria Tohver Milam, Naima Djeddar, Mary Catherine Adams
The Analyst
Genomics and Phylogenetic Studies
article

Identifying sequencing artifacts in low-diversity, short-insert aptamer libraries on the Illumina MiSeq platform

Anton V. Bryksin, Valeria Tohver Milam, Naima Djeddar, Mary Catherine Adams
article en

Abstract

∼1% in other samples) which resulted in overestimation of sample concentration and, in turn, decreased sequencing yield. Sequencing runs displayed highly variable pass rates (6-93%), likely due to short sequences outcompeting longer ones and assignment errors of neighboring clusters with identical early-cycle bases. Unexpected duplicates in template samples suggest that adapter-ligated daughter strands and optical misidentifications contribute to artificial duplication. These duplication artifacts are of particular importance in amplicon sequencing for which copy-number analysis is crucial. These findings highlight the need for refined ligation conditions, improved purification strategies, and enhanced bioinformatic filtering to ensure sequencing accuracy in aptamer screening library workflows on the Illumina platform.

The AnalystVol. 151(23)
Georgia Institute of Technology (US)
National Science Foundation, Division of Chemical, Bioengineering, Environmental, and Transport Systems, Division of Molecular and Cellular Biosciences
Genomics and Phylogenetic Studies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon
Identifying sequencing artifacts in low-diversity, short-insert aptamer libraries on the Illumina MiSeq platform — Anton V. Bryksin, Valeria Tohver Milam, et al. · The Analyst (2026) | TGRS Research Map | TGRS