Beam-width-corrected DOI resolution measurement of dual-ended readout PET detectors

OBJECTIVE: Dual-ended readout PET detectors provide depth-of-interaction (DOI) information for reducing parallax errors in high-resolution PET imaging. This study compared LYSO detectors with different inter-crystal reflectors and investigated the effect of irradiation beam width on DOI resolution measurements under electronic collimation. APPROACH: Two 28 × 28 LYSO arrays with 0.75 × 0.75 × 10 mm³ crystals and BaSO₄ or E60 reflectors were characterized in terms of flood-map quality, energy resolution, and DOI resolution. Part of E60 detector (8×8 crystals) was further measured using three irradiation geometries with beam widths of 0.80 mm, 0.62 mm, and 0.46 mm. A beam-width correction was applied to the measured DOI resolution, and the effects of energy selection and empirical flood-map-based inter-crystal scattering (ICS) filtering were evaluated. MAIN RESULTS: The BaSO₄ detector showed better flood-map quality and energy resolution, whereas the E60 detector showed better DOI resolution. In the 400-650 keV energy window, the measured average DOI resolutions of entire 28 × 28 LYSO pixels with irradiation beam width of 1.00 mm were 1.92 and 1.75 mm for the BaSO₄ and E60 detectors, respectively. For the selected 8 × 8 crystal region of the E60 detector, the corrected DOI resolution by subtracting the irradiation beam width improved from 1.19 mm for all events to 1.01 mm after energy selection, and further to 0.86 mm after additional empirical flood-map-based ICS filtering. SIGNIFICANCE: Irradiation beam width introduces a geometry-dependent contribution to DOI resolution measurements. Beam-width correction enables more consistent DOI characterization under different irradiation geometries.

Authors

Institutions

Publication Details

Journal
Physics in Medicine and Biology
Published
2026-11-03
DOI
https://doi.org/10.1088/1361-6560/aebd34
Primary Topic
Radiation Detection and Scintillator Technologies
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

Beam-width-corrected DOI resolution measurement of dual-ended readout PET detectors

Zhonghua Kuang, Ning Ren, Yongfeng Yang, Zheng Liu et al.
1 links
Physics in Medicine and Biology
Radiation Detection and Scintillator Technologies
article

Beam-width-corrected DOI resolution measurement of dual-ended readout PET detectors

Zhonghua Kuang, Ning Ren, Yongfeng Yang, Zheng Liu, Sen Yang, Hui Huang, Gangxin Hou, Haonan Li
article en

Abstract

OBJECTIVE: Dual-ended readout PET detectors provide depth-of-interaction (DOI) information for reducing parallax errors in high-resolution PET imaging. This study compared LYSO detectors with different inter-crystal reflectors and investigated the effect of irradiation beam width on DOI resolution measurements under electronic collimation. APPROACH: Two 28 × 28 LYSO arrays with 0.75 × 0.75 × 10 mm³ crystals and BaSO₄ or E60 reflectors were characterized in terms of flood-map quality, energy resolution, and DOI resolution. Part of E60 detector (8×8 crystals) was further measured using three irradiation geometries with beam widths of 0.80 mm, 0.62 mm, and 0.46 mm. A beam-width correction was applied to the measured DOI resolution, and the effects of energy selection and empirical flood-map-based inter-crystal scattering (ICS) filtering were evaluated. MAIN RESULTS: The BaSO₄ detector showed better flood-map quality and energy resolution, whereas the E60 detector showed better DOI resolution. In the 400-650 keV energy window, the measured average DOI resolutions of entire 28 × 28 LYSO pixels with irradiation beam width of 1.00 mm were 1.92 and 1.75 mm for the BaSO₄ and E60 detectors, respectively. For the selected 8 × 8 crystal region of the E60 detector, the corrected DOI resolution by subtracting the irradiation beam width improved from 1.19 mm for all events to 1.01 mm after energy selection, and further to 0.86 mm after additional empirical flood-map-based ICS filtering. SIGNIFICANCE: Irradiation beam width introduces a geometry-dependent contribution to DOI resolution measurements. Beam-width correction enables more consistent DOI characterization under different irradiation geometries.

Physics in Medicine and Biology
Southern University of Science and Technology (CN), Shenzhen Institutes of Advanced Technology (CN)
National Natural Science Foundation of China, National Key Research and Development Program of China, Basic and Applied Basic Research Foundation of Guangdong Province
Openalex Percentile: Top 15%
Radiation Detection and Scintillator Technologies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon