Air-Scan-Based Point-Spread-Function Estimation for Megavoltage CT with Flat-Panel Detectors

Accurate spatial-resolution characterization of flat-panel detectors at megavoltage energies is difficult due to long-tailed cross-talk. Conventional measurements at high energy using line-pair gauges or knife-edge phantoms require expensive, carefully fabricated objects and precise alignment, and despite these, scatter from the phantom, detector, and imaging room can further confound the measurement. We present a method for estimating the detector point spread function (PSF) from the mean noise power spectrum (NPS) calculated from repeated radiographs in air. The approach is mathematically equivalent to an autocovariance-based method but is more practical for long-tailed PSFs. We validate the method using simulated air data and Monte Carlo N-particle (MCNP6) calculations of energy deposition in a flat-panel detector. We show that deconvolving the experimental PSF increases the accuracy of transmission radiographs of step wedges. The method requires no phantom or alignments and is straightforward to reproduce.

Authors

Publication Details

Journal
arXiv (Cornell University)
Published
2026-11-02
DOI
https://doi.org/10.48550/arxiv.2611.03359
Primary Topic
Advanced Radiotherapy Techniques
Type
preprint
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
preprint

Air-Scan-Based Point-Spread-Function Estimation for Megavoltage CT with Flat-Panel Detectors

Joseph Bendahan, K. Aditya Mohan, Seemeen Karimi
arXiv (Cornell University)
Advanced Radiotherapy Techniques
preprint

Air-Scan-Based Point-Spread-Function Estimation for Megavoltage CT with Flat-Panel Detectors

Joseph Bendahan, K. Aditya Mohan, Seemeen Karimi
preprint en

Abstract

Accurate spatial-resolution characterization of flat-panel detectors at megavoltage energies is difficult due to long-tailed cross-talk. Conventional measurements at high energy using line-pair gauges or knife-edge phantoms require expensive, carefully fabricated objects and precise alignment, and despite these, scatter from the phantom, detector, and imaging room can further confound the measurement. We present a method for estimating the detector point spread function (PSF) from the mean noise power spectrum (NPS) calculated from repeated radiographs in air. The approach is mathematically equivalent to an autocovariance-based method but is more practical for long-tailed PSFs. We validate the method using simulated air data and Monte Carlo N-particle (MCNP6) calculations of energy deposition in a flat-panel detector. We show that deconvolving the experimental PSF increases the accuracy of transmission radiographs of step wedges. The method requires no phantom or alignments and is straightforward to reproduce.

arXiv (Cornell University)
Advanced Radiotherapy Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon