Gold–Titanium Nanofilm Textile Dry Electrodes for Electrocardiogram Monitoring: Electrical, Electromechanical and Skin‐Phantom Evaluation

Ag/AgCl gel electrodes are the clinical standard for electrocardiogram (ECG) recording, but at the skin–electrode biointerface, gel drying and skin irritation limit accurate recording over several days. To explore gel-free alternatives, we sputtered gold over a titanium adhesion layer onto serpentine textile dry electrodes and examined how Au:Ti thickness affects electrical, electromechanical and simulated ECG performance. Four variants (50:10, 80:15, 100:15 and 100:20 nm) were characterised by scanning electron microscopy, collinear four-probe measurement and mandrel bending, and the textile geometry by tensile testing. The two best coatings were then compared with commercial Ag/AgCl electrodes on a two-layer skin phantom using impedance measurements and simulated ECG acquisition. All coatings lowered sheet resistance relative to the uncoated textile, with the 80:15 nm variant giving the lowest value. The 100:20 nm variant was the most stable under bending and showed the least post-bend surface damage. The serpentine track did not significantly reduce bend-induced resistance change. On a porous phantom, both dry electrodes recorded identifiable ECG waveforms; mean signal-to-noise ratio was 24.4 ± 0.2 dB for Ag/AgCl, 15.9 ± 3.9 dB for 80:15 nm and 18.4 ± 0.5 dB for 100:20 nm (p = 0.061). Au:Ti-coated textiles are a feasible, gel-free biointerface for conformable ECG electrodes, but motion, durability and on-skin testing are required.

Authors

Publication Details

Journal
Preprints.org
Published
2026-11-04
DOI
https://doi.org/10.20944/preprints202611.0280.v1
Primary Topic
Advanced Sensor and Energy Harvesting Materials
Type
preprint
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
preprint

Gold–Titanium Nanofilm Textile Dry Electrodes for Electrocardiogram Monitoring: Electrical, Electromechanical and Skin‐Phantom Evaluation

Stuart MacVeigh, Sulab Manandhar, Sharon Man, Om Prakash Singh et al.
Preprints.org
Advanced Sensor and Energy Harvesting Materials
preprint

Gold–Titanium Nanofilm Textile Dry Electrodes for Electrocardiogram Monitoring: Electrical, Electromechanical and Skin‐Phantom Evaluation

Stuart MacVeigh, Sulab Manandhar, Sharon Man, Om Prakash Singh, David Jenkins, Alexande Besinis, E. Bamgboye
preprint en

Abstract

Ag/AgCl gel electrodes are the clinical standard for electrocardiogram (ECG) recording, but at the skin–electrode biointerface, gel drying and skin irritation limit accurate recording over several days. To explore gel-free alternatives, we sputtered gold over a titanium adhesion layer onto serpentine textile dry electrodes and examined how Au:Ti thickness affects electrical, electromechanical and simulated ECG performance. Four variants (50:10, 80:15, 100:15 and 100:20 nm) were characterised by scanning electron microscopy, collinear four-probe measurement and mandrel bending, and the textile geometry by tensile testing. The two best coatings were then compared with commercial Ag/AgCl electrodes on a two-layer skin phantom using impedance measurements and simulated ECG acquisition. All coatings lowered sheet resistance relative to the uncoated textile, with the 80:15 nm variant giving the lowest value. The 100:20 nm variant was the most stable under bending and showed the least post-bend surface damage. The serpentine track did not significantly reduce bend-induced resistance change. On a porous phantom, both dry electrodes recorded identifiable ECG waveforms; mean signal-to-noise ratio was 24.4 ± 0.2 dB for Ag/AgCl, 15.9 ± 3.9 dB for 80:15 nm and 18.4 ± 0.5 dB for 100:20 nm (p = 0.061). Au:Ti-coated textiles are a feasible, gel-free biointerface for conformable ECG electrodes, but motion, durability and on-skin testing are required.

Preprints.org
Advanced Sensor and Energy Harvesting Materials
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon