Drug–drug adsorption-based miniaturized solid phase extraction of amlodipine from biological samples before HPLC–PDA analysis
. The relative recovery values ranged between 88% and 96% and indicated negligible matrix effects, which confirms the suitability of the method for reliable determination of amlodipine in biological matrices.