Daytime Idle Energy Reduction in Radiotherapy Simulation Computed Tomography: A Prospective Workflow‐Based Evaluation

ABSTRACT Introduction Previous computed tomography (CT) energy‐saving studies have primarily focused on after‐hours shutdown or idle energy use in diagnostic CT. Radiotherapy simulation, however, includes preparation and immobilisation steps that may prolong daytime non‐acquisition periods. This study quantified daytime energy use in radiotherapy simulation CT and evaluated the contribution of power‐saving operation. Methods In this prospective single‐centre study, electrical power consumption was directly measured using a three‐phase power logger during five baseline and five power‐saving weekdays (08:00–18:00). Power measurements were integrated with patient in‐room intervals and simulation procedure categories. Staff perceptions of workflow impact were assessed using an anonymous questionnaire, and carbon implications were estimated from the measured energy reduction. Results Mean directly measured clinical‐hour energy consumption decreased from 38.7 to 36.5 kWh/day with power‐saving operation, a 5.8% reduction. Non‐acquisition time during patient presence varied markedly by procedure, from 56.5% for respiratory‐gated four‐dimensional CT to 97.1% for head‐and‐neck simulation. Power‐saving operation shifted daytime power use towards a lower‐power state, corresponding to an estimated reduction of 0.84–1.01 kg‐CO 2 per clinical day across the emission‐factor sensitivity range. Twenty of 21 staff (95.2%) reported no impact on patient throughput. Conclusion Power‐saving operation reduced daytime energy consumption without meaningful workflow disruption. Radiotherapy simulation CT has prolonged, procedure‐dependent idle periods arising from radiotherapy‐specific preparation workflows, highlighting a practical and previously undercharacterised opportunity for energy reduction.

Authors

Institutions

Publication Details

Journal
Journal of Medical Radiation Sciences
Published
2026-11-02
DOI
https://doi.org/10.1002/jmrs.70133
Primary Topic
Energy Efficiency and Management
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Daytime Idle Energy Reduction in Radiotherapy Simulation Computed Tomography: A Prospective Workflow‐Based Evaluation

Kouta Hirotaki, Yoshiaki Honda, Masashi Ito
3 links
Journal of Medical Radiation Sciences
Energy Efficiency and Management
article

Daytime Idle Energy Reduction in Radiotherapy Simulation Computed Tomography: A Prospective Workflow‐Based Evaluation

Kouta Hirotaki, Yoshiaki Honda, Masashi Ito
article en

Abstract

ABSTRACT Introduction Previous computed tomography (CT) energy‐saving studies have primarily focused on after‐hours shutdown or idle energy use in diagnostic CT. Radiotherapy simulation, however, includes preparation and immobilisation steps that may prolong daytime non‐acquisition periods. This study quantified daytime energy use in radiotherapy simulation CT and evaluated the contribution of power‐saving operation. Methods In this prospective single‐centre study, electrical power consumption was directly measured using a three‐phase power logger during five baseline and five power‐saving weekdays (08:00–18:00). Power measurements were integrated with patient in‐room intervals and simulation procedure categories. Staff perceptions of workflow impact were assessed using an anonymous questionnaire, and carbon implications were estimated from the measured energy reduction. Results Mean directly measured clinical‐hour energy consumption decreased from 38.7 to 36.5 kWh/day with power‐saving operation, a 5.8% reduction. Non‐acquisition time during patient presence varied markedly by procedure, from 56.5% for respiratory‐gated four‐dimensional CT to 97.1% for head‐and‐neck simulation. Power‐saving operation shifted daytime power use towards a lower‐power state, corresponding to an estimated reduction of 0.84–1.01 kg‐CO 2 per clinical day across the emission‐factor sensitivity range. Twenty of 21 staff (95.2%) reported no impact on patient throughput. Conclusion Power‐saving operation reduced daytime energy consumption without meaningful workflow disruption. Radiotherapy simulation CT has prolonged, procedure‐dependent idle periods arising from radiotherapy‐specific preparation workflows, highlighting a practical and previously undercharacterised opportunity for energy reduction.

Journal of Medical Radiation Sciences
National Cancer Center Hospital East (JP)
Good health and well-being, Affordable and clean energy
Openalex Percentile: Top 37%
Energy Efficiency and Management
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon