Could calprotectin and pentraxin 3 be measured in cerebrospinal fluid – a verification study

Abstract Objectives Calprotectin and pentraxin 3 (PTX3) have emerged as potential biomarkers of central nervous system inflammatory diseases, yet evidence based analytical verifications in cerebrospinal fluid (CSF) is lacking. This study aimed to verify the manufacturer’s claims for precision, limit of quantitation (LoQ) and stability for calprotectin using fCAL turbo reagent (Bühlmann Laboratories AG, Schonenbuch, Switzerland); precision and linearity for PTX3 using Human PTX3 ELISA Kit (BioVendor Group, Brno, Czech Republic) in CSF. Methods Samples were collected in additive-free tubes and centrifuged at 400 g, 10 min. Calprotectin was measured on Atellica Solution analyser (Siemens), PTX3 was determined by ELISA method on the Thunderbolt analyser (Gold Standards Diagnostics). The verification protocol was carried out according to Clinical and Laboratory Standards Institute (CLSI) guidelines: precision by EP15-A3, LoQ verification by EP17-A2, and linearity by EP6-A. Calprotectin stability at −20 °C for 7 days was assessed in three CSF samples. Results The CV%s for calprotectin precision were 4.52 % and 1.66 %; for PTX3 were 6.0 % and 8.9 %, respectively. The declared LoQ for fCAL turbo of 50.0 μg/L (CV%<20 %) was verified. The stability equation for calprotectin did not exceed the MPD (41.6 %), and it was y=−0.15× (95 % Cl −0.49 to 0.19), indicating stability at −20 °C for 7 days. The method for PTX3 was linear over the manufacturer declared ranges (3,212.0–406.0 and 3,315.0–411.0 ng/L). Conclusions This study showed that Bühlmann fCAL turbo reagent and the Biovendor Human PTX3 ELISA kit support the determination of calprotectin and PTX3 in CSF for research purposes, with potential future clinical applications.

Authors

Institutions

Publication Details

Journal
Advances in Laboratory Medicine / Avances en Medicina de Laboratorio
Published
2026-11-02
DOI
https://doi.org/10.1515/almed-2026-0017
Primary Topic
Clinical Laboratory Practices and Quality Control
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Could calprotectin and pentraxin 3 be measured in cerebrospinal fluid – a verification study

Edvard Galić, Ivana Zobić, Latica Friedrich, Helena Čičak et al.
3 links
Advances in Laboratory Medicine / Avances en Medicina de Laboratorio
Clinical Laboratory Practices and Quality Control
article

Could calprotectin and pentraxin 3 be measured in cerebrospinal fluid – a verification study

Edvard Galić, Ivana Zobić, Latica Friedrich, Helena Čičak, Lora Dukić, Andrea Saračević, Igor Alfirević, Marko Kutleša, Ana Sruk, Renata Laškaj, Sanja Tomasović, Marija Santini, Marija Grdić Rajković, Hrvoje Galic, Željka Šarčević
article en

Abstract

Abstract Objectives Calprotectin and pentraxin 3 (PTX3) have emerged as potential biomarkers of central nervous system inflammatory diseases, yet evidence based analytical verifications in cerebrospinal fluid (CSF) is lacking. This study aimed to verify the manufacturer’s claims for precision, limit of quantitation (LoQ) and stability for calprotectin using fCAL turbo reagent (Bühlmann Laboratories AG, Schonenbuch, Switzerland); precision and linearity for PTX3 using Human PTX3 ELISA Kit (BioVendor Group, Brno, Czech Republic) in CSF. Methods Samples were collected in additive-free tubes and centrifuged at 400 g, 10 min. Calprotectin was measured on Atellica Solution analyser (Siemens), PTX3 was determined by ELISA method on the Thunderbolt analyser (Gold Standards Diagnostics). The verification protocol was carried out according to Clinical and Laboratory Standards Institute (CLSI) guidelines: precision by EP15-A3, LoQ verification by EP17-A2, and linearity by EP6-A. Calprotectin stability at −20 °C for 7 days was assessed in three CSF samples. Results The CV%s for calprotectin precision were 4.52 % and 1.66 %; for PTX3 were 6.0 % and 8.9 %, respectively. The declared LoQ for fCAL turbo of 50.0 μg/L (CV%<20 %) was verified. The stability equation for calprotectin did not exceed the MPD (41.6 %), and it was y=−0.15× (95 % Cl −0.49 to 0.19), indicating stability at −20 °C for 7 days. The method for PTX3 was linear over the manufacturer declared ranges (3,212.0–406.0 and 3,315.0–411.0 ng/L). Conclusions This study showed that Bühlmann fCAL turbo reagent and the Biovendor Human PTX3 ELISA kit support the determination of calprotectin and PTX3 in CSF for research purposes, with potential future clinical applications.

Advances in Laboratory Medicine / Avances en Medicina de Laboratorio
University of Zagreb (HR), Klinička bolnica "Sveti Duh" (HR), University North (HR), Catholic University of Croatia (HR), University Hospital for Infectious Diseases "Dr Fran Mihaljevic" (HR), University of Osijek (HR)
Good health and well-being
Openalex Percentile: Top 15%
Clinical Laboratory Practices and Quality Control
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon