Optimizing patient safety: an evaluation of computed tomography dose-length product variations across hospitals in Abuja, Nigeria

The increasing use of computed tomography (CT) as a diagnostic tool in Nigeria raises concerns about high radiation doses and potential cancer risk. To address this concern, CT dose-length product (DLP) was evaluated at ten hospitals in Abuja, Nigeria, using polymethyl methacrylate phantoms, a 100 mm pencil ionization chamber, an electrometer, and a CT probe. The study assessed variations in radiation dose across the facilities by measuring DLP in mGy.cm for chest and head phantoms. Chest phantom DLP values ranged from 79.36 ± 0.001 mGy.cm at Hospital H to 677.40 ± 0.01 mGy.cm at Hospital D. Head phantom values ranged from 230.14 ± 0.004 mGy.cm at Hospital A to 1165.73 ± 0.3 mGy.cm at Hospital I. Most facilities recorded DLP values below the European diagnostic reference levels (DRLs) used for comparison. However, Hospital D exceeded the chest reference level of 650 mGy.cm, while Hospitals E and I exceeded the head reference level of 1050 mGy.cm. These findings indicate substantial variation in scanning protocols among the facilities. The study recommends routine quality control, equipment calibration, standardized protocols, and refresher training for CT operators to ensure doses administered to patients are As Low as Reasonably Achievable (ALARA) principle and reduce unnecessary radiation exposure.

Authors

Institutions

Publication Details

Journal
Recent Advances in Natural Sciences
Published
2026-11-02
DOI
https://doi.org/10.61298/rans.2026.4.2.415
Primary Topic
Radiation Dose and Imaging
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Optimizing patient safety: an evaluation of computed tomography dose-length product variations across hospitals in Abuja, Nigeria

Joseph Teryima Iortile, John Kahugu Simon
4 links
Recent Advances in Natural Sciences
Radiation Dose and Imaging
article

Optimizing patient safety: an evaluation of computed tomography dose-length product variations across hospitals in Abuja, Nigeria

Joseph Teryima Iortile, John Kahugu Simon
article en

Abstract

The increasing use of computed tomography (CT) as a diagnostic tool in Nigeria raises concerns about high radiation doses and potential cancer risk. To address this concern, CT dose-length product (DLP) was evaluated at ten hospitals in Abuja, Nigeria, using polymethyl methacrylate phantoms, a 100 mm pencil ionization chamber, an electrometer, and a CT probe. The study assessed variations in radiation dose across the facilities by measuring DLP in mGy.cm for chest and head phantoms. Chest phantom DLP values ranged from 79.36 ± 0.001 mGy.cm at Hospital H to 677.40 ± 0.01 mGy.cm at Hospital D. Head phantom values ranged from 230.14 ± 0.004 mGy.cm at Hospital A to 1165.73 ± 0.3 mGy.cm at Hospital I. Most facilities recorded DLP values below the European diagnostic reference levels (DRLs) used for comparison. However, Hospital D exceeded the chest reference level of 650 mGy.cm, while Hospitals E and I exceeded the head reference level of 1050 mGy.cm. These findings indicate substantial variation in scanning protocols among the facilities. The study recommends routine quality control, equipment calibration, standardized protocols, and refresher training for CT operators to ensure doses administered to patients are As Low as Reasonably Achievable (ALARA) principle and reduce unnecessary radiation exposure.

Recent Advances in Natural Sciences
National Open University of Nigeria (NG)
Good health and well-being
Openalex Percentile: Top 15%
Radiation Dose and Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon