Irradiance of Phototherapy Devices in General Hospitals in the Northern Netherlands

Background Sufficient irradiance levels of phototherapy (PT) devices are essential for the efficacy of PT. Yet, irradiance levels are infrequently measured, are sometimes lower than recommended, and little is known about whether improvement over time occurs. Aim To investigate the irradiance levels of PT devices in pediatric departments of all general hospitals in the Northern Netherlands in two different time epochs. Methods Irradiance levels of all available PT devices were measured in eight pediatric departments in 2023, and median levels were compared with levels measured in 2017. The PT devices were positioned as they were routinely used. Two calibrated radiometers were used: the Dale 40 Phototherapy Radiometer (Fluke Biomedical, Everett, WA) and the Ohmeda Medical BiliBlanket Meter II (BBII, GE Healthcare, Chicago, IL). Results Irradiance levels of overhead devices increased from 20 µW/cm²/nm in 2017 to 24 µW/cm²/nm in 2023 ( P = .021). The mean distance between the PT device and the silhouette decreased by 19 cm in 2023 ( P < .001). Irradiance of the devices underneath was similar in 2017 and 2023, that is, 25 (18-28) and 24 (11-34) µW/cm²/nm, respectively. There were no differences in median irradiance between hospitals. High-intensity irradiance levels >30 µW/cm²/nm were not possible for all devices, but more (66% in 2017; 97% in 2023) PT devices met the recommended minimum irradiance. Irradiance measurements of the BBII were significantly higher than those of the Dale 40 for underneath PT devices. Conclusions Irradiance levels of PT devices in general hospitals increased over time, but high-intensity PT was not always possible. The observed improvement may be explained by the shorter distance between the PT device and the silhouette and the use of newer PT devices. Finally, the radiometer-specific irradiance measurements underline the importance of measuring irradiance with an appropriate radiometer.

Authors

Institutions

Publication Details

Journal
Journal of Neonatology
Published
2026-11-01
DOI
https://doi.org/10.1177/09732179261499038
Primary Topic
Neonatal Health and Biochemistry
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Irradiance of Phototherapy Devices in General Hospitals in the Northern Netherlands

Christian Victor Hulzebos, Peter H. Dijk, A.C. Hagendoorn, P. Feenstra et al.
Journal of Neonatology
Neonatal Health and Biochemistry
article

Irradiance of Phototherapy Devices in General Hospitals in the Northern Netherlands

Christian Victor Hulzebos, Peter H. Dijk, A.C. Hagendoorn, P. Feenstra, C. B. Heierman-Hendriksen, R.N. Garlich
article en

Abstract

Background Sufficient irradiance levels of phototherapy (PT) devices are essential for the efficacy of PT. Yet, irradiance levels are infrequently measured, are sometimes lower than recommended, and little is known about whether improvement over time occurs. Aim To investigate the irradiance levels of PT devices in pediatric departments of all general hospitals in the Northern Netherlands in two different time epochs. Methods Irradiance levels of all available PT devices were measured in eight pediatric departments in 2023, and median levels were compared with levels measured in 2017. The PT devices were positioned as they were routinely used. Two calibrated radiometers were used: the Dale 40 Phototherapy Radiometer (Fluke Biomedical, Everett, WA) and the Ohmeda Medical BiliBlanket Meter II (BBII, GE Healthcare, Chicago, IL). Results Irradiance levels of overhead devices increased from 20 µW/cm²/nm in 2017 to 24 µW/cm²/nm in 2023 ( P = .021). The mean distance between the PT device and the silhouette decreased by 19 cm in 2023 ( P < .001). Irradiance of the devices underneath was similar in 2017 and 2023, that is, 25 (18-28) and 24 (11-34) µW/cm²/nm, respectively. There were no differences in median irradiance between hospitals. High-intensity irradiance levels >30 µW/cm²/nm were not possible for all devices, but more (66% in 2017; 97% in 2023) PT devices met the recommended minimum irradiance. Irradiance measurements of the BBII were significantly higher than those of the Dale 40 for underneath PT devices. Conclusions Irradiance levels of PT devices in general hospitals increased over time, but high-intensity PT was not always possible. The observed improvement may be explained by the shorter distance between the PT device and the silhouette and the use of newer PT devices. Finally, the radiometer-specific irradiance measurements underline the importance of measuring irradiance with an appropriate radiometer.

Journal of Neonatology
University Medical Center Groningen (NL), Beatrix Kinderziekenhuis (NL)
Openalex Percentile: Top 7%
Neonatal Health and Biochemistry
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon