Evaluation of Temporal Bone Image Quality with Low-Dose Photon-Counting CT in Cadaveric Heads

BACKGROUND AND PURPOSE: Photon-counting computed tomography (PCCT) is an emerging technology that offers improved spatial resolution and noise reduction at lower radiation doses than conventional CT; however, the effect of dose reduction on clinically significant image quality has not been well established. Following the ALARA (As Low As Reasonably Achievable) principle, this study evaluated temporal bone image quality across decreasing radiation doses to identify the minimum dose required to maintain diagnostic integrity. MATERIALS AND METHODS: Seven cadaveric heads underwent temporal bone PCCT imaging at four tube current-time products: 290, 75, 25, and 11 mAs. All scans were acquired at 120 kV, 0.55 pitch, and a 0.5 second rotation time and reconstructed with 0.2-mm slice thickness in axial and coronal planes for left and right temporal bones. Three board-certified radiologists independently assessed prespecified, randomized temporal bone landmarks using a 5-point Likert scale to determine clinical utility (1-nondiagnostic; 5-excellent). Mixed-effects linear regression with random intercepts for patient, reader, and landmark estimated mean scores by dose. Pairwise comparisons were Tukey-adjusted, and one-sided FDR-adjusted t-tests evaluated whether predicted mean scores exceeded 3 (clinically sufficient). RESULTS: At 290 and 75 mAs, predicted mean scores were 4.81 (95% CI: 4.42-5.21) and 4.21 (95% CI: 3.82-4.61), respectively, demonstrating good/excellent image quality for all evaluated landmarks and exceeding threshold for clinical sufficiency (FDR-adjusted p<0.0001). At 25 mAs, predicted mean scores were 2.80 (95% CI: 2.41-3.19), falling in the poor/sufficient range, and at 11 mAs, were rated the lowest at 2.09 (95% CI: 1.70-2.48), scoring below the clinical sufficiency threshold (p=0.99). Overall, mean Likert image quality scores decreased with progressive mAs reduction, remaining technically interpretable but below the threshold for clinical sufficiency at 25 mAs. All pairwise comparisons were statistically significant (Tukey-adjusted p<0.0001). Interreader agreement (Krippendorff's α) ranged from 0.301 to 0.657 across landmarks. CONCLUSION: PCCT maintains clinically sufficient temporal bone image quality at tube current-time products as low as 75 mAs, while doses of 25 mAs and below fall below the threshold for reliable clinical use. Per ALARA, this threshold optimizes safety via significant dose reduction while maintaining the high image quality necessary for complex temporal bone anatomy.

Authors

Publication Details

Journal
American Journal of Neuroradiology
Published
2026-11-02
DOI
https://doi.org/10.3174/ajnr.a9701
Primary Topic
Advanced X-ray and CT Imaging
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Evaluation of Temporal Bone Image Quality with Low-Dose Photon-Counting CT in Cadaveric Heads

Yueh Z. Lee, Georgia R.R.S.S. Silva, Thad Benefield, Adam J. Kimple et al.
2 links
American Journal of Neuroradiology
Advanced X-ray and CT Imaging
article

Evaluation of Temporal Bone Image Quality with Low-Dose Photon-Counting CT in Cadaveric Heads

Yueh Z. Lee, Georgia R.R.S.S. Silva, Thad Benefield, Adam J. Kimple, Kassie L. McCullagh, Willem G. Calderon, Ji Young Lee, Seong-Won Kim
article en

Abstract

BACKGROUND AND PURPOSE: Photon-counting computed tomography (PCCT) is an emerging technology that offers improved spatial resolution and noise reduction at lower radiation doses than conventional CT; however, the effect of dose reduction on clinically significant image quality has not been well established. Following the ALARA (As Low As Reasonably Achievable) principle, this study evaluated temporal bone image quality across decreasing radiation doses to identify the minimum dose required to maintain diagnostic integrity. MATERIALS AND METHODS: Seven cadaveric heads underwent temporal bone PCCT imaging at four tube current-time products: 290, 75, 25, and 11 mAs. All scans were acquired at 120 kV, 0.55 pitch, and a 0.5 second rotation time and reconstructed with 0.2-mm slice thickness in axial and coronal planes for left and right temporal bones. Three board-certified radiologists independently assessed prespecified, randomized temporal bone landmarks using a 5-point Likert scale to determine clinical utility (1-nondiagnostic; 5-excellent). Mixed-effects linear regression with random intercepts for patient, reader, and landmark estimated mean scores by dose. Pairwise comparisons were Tukey-adjusted, and one-sided FDR-adjusted t-tests evaluated whether predicted mean scores exceeded 3 (clinically sufficient). RESULTS: At 290 and 75 mAs, predicted mean scores were 4.81 (95% CI: 4.42-5.21) and 4.21 (95% CI: 3.82-4.61), respectively, demonstrating good/excellent image quality for all evaluated landmarks and exceeding threshold for clinical sufficiency (FDR-adjusted p<0.0001). At 25 mAs, predicted mean scores were 2.80 (95% CI: 2.41-3.19), falling in the poor/sufficient range, and at 11 mAs, were rated the lowest at 2.09 (95% CI: 1.70-2.48), scoring below the clinical sufficiency threshold (p=0.99). Overall, mean Likert image quality scores decreased with progressive mAs reduction, remaining technically interpretable but below the threshold for clinical sufficiency at 25 mAs. All pairwise comparisons were statistically significant (Tukey-adjusted p<0.0001). Interreader agreement (Krippendorff's α) ranged from 0.301 to 0.657 across landmarks. CONCLUSION: PCCT maintains clinically sufficient temporal bone image quality at tube current-time products as low as 75 mAs, while doses of 25 mAs and below fall below the threshold for reliable clinical use. Per ALARA, this threshold optimizes safety via significant dose reduction while maintaining the high image quality necessary for complex temporal bone anatomy.

American Journal of Neuroradiology
Good health and well-being
Openalex Percentile: Top 26%
Advanced X-ray and CT Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon