A Novel Automated Approach for the Evaluation and Correction of Geometric Distortion in Magnetic Resonance Imaging

Abstract Background: The geometric distortion of magnetic resonance (MR) images, mainly due to the inhomogeneity of the magnetic field, is a limiting factor for MR-only radiotherapy. Aims: This study presents a new automated method to identify and correct MR imaging (MRI) image distortion. Methods: A large 3D home-made phantom with cylindrical water pipes has been used for the geometric distortion measurements of three MR protocols: turbo inversion recovery magnitude (TIRM), turbo spin-echo (TSE), and fast low angle shot (FLASH) sequences. The vascular modeling toolkit, 3D Slicer, and MATLAB software have been used to determine the correction matrix of MR geometric distortion. Results: The maximum mean absolute error (MAE) of geometric distortion has been achieved 8.17, 4.25, and 7.12 in TIRM, TSE, and FLASH MR sequences, respectively, that has been reduced to 2.10, 0.57, and 0.38 after the automated correction of geometric distortion. In TIRM and TSE protocols, the phase-encoding direction showed more distortion than the frequency-encoding direction. The maximum and minimum MAE values of 8.17 and 1.72 mm has been seen in the phase- and frequency-encoding direction of TIRM and TSE sequences, respectively. Conclusion: The observed distortion of MR images depends on the image protocol and the slice position, resulting in a unique conversion matrix for each MR image sequence. The automated method proposed in this research can effectively assess and rectify geometric distortions in MR images for large field of views of MRI-only radiotherapy purposes.

Authors

Institutions

Publication Details

Journal
Journal of Medical Signals & Sensors
Published
2026-11-01
DOI
https://doi.org/10.4103/jmss.jmss_10_25
Primary Topic
Advanced MRI Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

A Novel Automated Approach for the Evaluation and Correction of Geometric Distortion in Magnetic Resonance Imaging

Erfan Saatchian, Azam Eskandari, Atefeh Rostami, Shahrokh Naseri et al.
Journal of Medical Signals & Sensors
Advanced MRI Techniques and Applications
article

A Novel Automated Approach for the Evaluation and Correction of Geometric Distortion in Magnetic Resonance Imaging

Erfan Saatchian, Azam Eskandari, Atefeh Rostami, Shahrokh Naseri, Hosein Mazaheri, Mahdi Momennezhad
article en

Abstract

Abstract Background: The geometric distortion of magnetic resonance (MR) images, mainly due to the inhomogeneity of the magnetic field, is a limiting factor for MR-only radiotherapy. Aims: This study presents a new automated method to identify and correct MR imaging (MRI) image distortion. Methods: A large 3D home-made phantom with cylindrical water pipes has been used for the geometric distortion measurements of three MR protocols: turbo inversion recovery magnitude (TIRM), turbo spin-echo (TSE), and fast low angle shot (FLASH) sequences. The vascular modeling toolkit, 3D Slicer, and MATLAB software have been used to determine the correction matrix of MR geometric distortion. Results: The maximum mean absolute error (MAE) of geometric distortion has been achieved 8.17, 4.25, and 7.12 in TIRM, TSE, and FLASH MR sequences, respectively, that has been reduced to 2.10, 0.57, and 0.38 after the automated correction of geometric distortion. In TIRM and TSE protocols, the phase-encoding direction showed more distortion than the frequency-encoding direction. The maximum and minimum MAE values of 8.17 and 1.72 mm has been seen in the phase- and frequency-encoding direction of TIRM and TSE sequences, respectively. Conclusion: The observed distortion of MR images depends on the image protocol and the slice position, resulting in a unique conversion matrix for each MR image sequence. The automated method proposed in this research can effectively assess and rectify geometric distortions in MR images for large field of views of MRI-only radiotherapy purposes.

Journal of Medical Signals & SensorsVol. 16(11)
Mashhad University of Medical Sciences (IR), Sabzevar University of Medical Sciences (IR)
Openalex Percentile: Top 13%
Advanced MRI Techniques and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon
A Novel Automated Approach for the Evaluation and Correction of Geometric Distortion in Magnetic Resonance Imaging — Erfan Saatchian, Azam Eskandari, et al. · Journal of Medical Signals & Sensors (2026) | TGRS Research Map | TGRS