Fluid-solid coupled biomechanical analysis of implants and the microenvironment in Pauwels type III fractures

Pauwels type III femoral neck fractures are characterized by a steep fracture orientation and substantial shear loading. Increasing knee flexion during loaded activities alters hip joint loading and may affect both fixation mechanics and the poroelastic behavior of the interfragmentary healing tissue. How these effects differ among fixation constructs and postoperative tissue states remains unclear. This study compared cannulated compression screws (CCS), the dynamic hip screw (DHS), and the femoral neck system (FNS) using poroelastic finite element models based on Biot's theory, with material properties representing two postoperative tissue states nominally corresponding to 1 and 12 weeks postoperatively. Hip joint reaction forces obtained from musculoskeletal simulations of squatting for the study subject were applied under five loading conditions corresponding to prescribed knee flexion targets of 60°, 75°, 90°, 105°, and 120°. The results showed that under the condition representing 1 week after surgery and at a prescribed knee flexion target of 120°, implant stresses were 1175.38, 928.77, and 634.35 MPa for CCS, DHS, and FNS, respectively, with FNS exhibiting lower stress than CCS and DHS. The early tissue state also produced large localized principal strain and pronounced pore pressure gradients. Under the 12-week condition, implant stresses decreased below 162 MPa and tissue strain remained below 0.013, although peak pore pressure was higher. The higher assigned tissue stiffness enhanced load sharing and reduced deformation. Overall, knee flexion and tissue state jointly affected fixation mechanics and the local poroelastic behavior, providing a biomechanical insight for postoperative load evaluation.

Authors

Institutions

Publication Details

Journal
Proceedings of the Institution of Mechanical Engineers Part H Journal of Engineering in Medicine
Published
2026-11-02
DOI
https://doi.org/10.1177/09544119261500559
Primary Topic
Hip and Femur Fractures
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Fluid-solid coupled biomechanical analysis of implants and the microenvironment in Pauwels type III fractures

Yafeng Li, Jiang Liu, Feng Zhang, Xi Zhang et al.
3 links
Proceedings of the Institution of Mechanical Engineers Part H Journal of Engineering in Medicine
Hip and Femur Fractures
article

Fluid-solid coupled biomechanical analysis of implants and the microenvironment in Pauwels type III fractures

Yafeng Li, Jiang Liu, Feng Zhang, Xi Zhang, Xiaotian Li, Xinbo Li, Lei Wang
article en

Abstract

Pauwels type III femoral neck fractures are characterized by a steep fracture orientation and substantial shear loading. Increasing knee flexion during loaded activities alters hip joint loading and may affect both fixation mechanics and the poroelastic behavior of the interfragmentary healing tissue. How these effects differ among fixation constructs and postoperative tissue states remains unclear. This study compared cannulated compression screws (CCS), the dynamic hip screw (DHS), and the femoral neck system (FNS) using poroelastic finite element models based on Biot's theory, with material properties representing two postoperative tissue states nominally corresponding to 1 and 12 weeks postoperatively. Hip joint reaction forces obtained from musculoskeletal simulations of squatting for the study subject were applied under five loading conditions corresponding to prescribed knee flexion targets of 60°, 75°, 90°, 105°, and 120°. The results showed that under the condition representing 1 week after surgery and at a prescribed knee flexion target of 120°, implant stresses were 1175.38, 928.77, and 634.35 MPa for CCS, DHS, and FNS, respectively, with FNS exhibiting lower stress than CCS and DHS. The early tissue state also produced large localized principal strain and pronounced pore pressure gradients. Under the 12-week condition, implant stresses decreased below 162 MPa and tissue strain remained below 0.013, although peak pore pressure was higher. The higher assigned tissue stiffness enhanced load sharing and reduced deformation. Overall, knee flexion and tissue state jointly affected fixation mechanics and the local poroelastic behavior, providing a biomechanical insight for postoperative load evaluation.

Proceedings of the Institution of Mechanical Engineers Part H Journal of Engineering in Medicine
Tiangong University (CN), Tianjin Hospital (CN), Tianjin Beichen Hospital (CN)
Good health and well-being
Openalex Percentile: Top 11%
Hip and Femur Fractures
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon