Biomechanical effects of different suture pretension levels in suture-bridge repair of supraspinatus tendon tears: a finite element analysis

Objective The optimal initial suture pretension during suture-bridge repair of full-thickness supraspinatus tendon tears remains uncertain and is largely determined by the surgeon’s experience. This study aimed to evaluate the effects of different suture pretensions on the mechanical environment of the tendon-bone interface and to provide a biomechanical reference for intraoperative tension control. Methods A three-dimensional finite element model of the shoulder joint was reconstructed from computed tomography data of a healthy adult volunteer. A full-thickness supraspinatus tear was simulated and repaired using a double-row suture-bridge configuration. Seven total pretension levels (30–90 N, in 10 N increments) and three shoulder abduction positions (10°, 20°, and 30°) were combined to generate 21 loading conditions. The mean contact pressure, effective contact area, relative interfacial displacement, and tendon-side von Mises stress were analyzed. Results Increasing pretension increased the mean tendon-bone contact pressure at all abduction angles. At 20° of abduction, the mean contact pressure increased from 0.382 MPa to 0.856 MPa. The peak tendon-side von Mises stress also increased with pretension, reaching a maximum of 4.533 MPa in the 30°/90 N condition. The effective contact area changed only slightly at 10° of abduction, decreased from 181.4 mm 2 to 169.5 mm 2 at 20°, and reached relatively high values at 50–60 N at 30°. High-stress regions were mainly concentrated around the medial-row suture perforation sites and expanded as pretension increased. Conclusion Suture pretension improved tendon-bone compression and initial interfacial stability, but higher pretension increased local pressure and tendon stress concentration. Within the tested configurations, 50–60 N represented an intermediate, condition-dependent range rather than a universal clinical target.

Authors

Institutions

Publication Details

Journal
Frontiers in Bioengineering and Biotechnology
Published
2026-11-02
DOI
https://doi.org/10.3389/fbioe.2026.1935045
Primary Topic
Shoulder Injury and Treatment
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Biomechanical effects of different suture pretension levels in suture-bridge repair of supraspinatus tendon tears: a finite element analysis

Chun Ge, SUN Rongxin, Limin Mou, 黄异飞 et al.
1 links
Frontiers in Bioengineering and Biotechnology
Shoulder Injury and Treatment
article

Biomechanical effects of different suture pretension levels in suture-bridge repair of supraspinatus tendon tears: a finite element analysis

Chun Ge, SUN Rongxin, Limin Mou, 黄异飞, Yitong Li, Hongdong Wang
article en

Abstract

Objective The optimal initial suture pretension during suture-bridge repair of full-thickness supraspinatus tendon tears remains uncertain and is largely determined by the surgeon’s experience. This study aimed to evaluate the effects of different suture pretensions on the mechanical environment of the tendon-bone interface and to provide a biomechanical reference for intraoperative tension control. Methods A three-dimensional finite element model of the shoulder joint was reconstructed from computed tomography data of a healthy adult volunteer. A full-thickness supraspinatus tear was simulated and repaired using a double-row suture-bridge configuration. Seven total pretension levels (30–90 N, in 10 N increments) and three shoulder abduction positions (10°, 20°, and 30°) were combined to generate 21 loading conditions. The mean contact pressure, effective contact area, relative interfacial displacement, and tendon-side von Mises stress were analyzed. Results Increasing pretension increased the mean tendon-bone contact pressure at all abduction angles. At 20° of abduction, the mean contact pressure increased from 0.382 MPa to 0.856 MPa. The peak tendon-side von Mises stress also increased with pretension, reaching a maximum of 4.533 MPa in the 30°/90 N condition. The effective contact area changed only slightly at 10° of abduction, decreased from 181.4 mm 2 to 169.5 mm 2 at 20°, and reached relatively high values at 50–60 N at 30°. High-stress regions were mainly concentrated around the medial-row suture perforation sites and expanded as pretension increased. Conclusion Suture pretension improved tendon-bone compression and initial interfacial stability, but higher pretension increased local pressure and tendon stress concentration. Within the tested configurations, 50–60 N represented an intermediate, condition-dependent range rather than a universal clinical target.

Frontiers in Bioengineering and BiotechnologyVol. 14
Xinjiang Medical University (CN), Sixth Affiliated Hospital of Xinjiang Medical University (CN)
Good health and well-being
Openalex Percentile: Top 11%
Shoulder Injury and Treatment
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon