Development and commissioning of a LaBr3:Ce, Sr-based collimated hard x-ray spectrometer for MAST-U

A new collimated hard x-ray (HXR) spectrometer has been developed and installed on the Mega Amp Spherical Tokamak Upgrade (MAST-U) to support future studies of runaway electron generation and provide a diagnostic capability to plasma-facing components monitoring. The diagnostic is based on an Sr-codoped LaBr3:Ce scintillator (2″ × 2″) coupled to a fast data acquisition system capable of recording raw scintillation waveforms in both pulse and continuous modes, allowing detailed reconstruction of individual events using algorithms for optimization of spectrometric parameters and pulse-shape discrimination. The detector was energy-calibrated up to 4.8 MeV using standard spectroscopy sources and a 66Ga radioisotope. Its performance was then evaluated at the MC40 cyclotron (University of Birmingham), demonstrating stable operation at 1.1 Mcps, with gain shifts of less than 4%. Following successful validation, the diagnostic was installed using an existing midplane line-of-sight configuration of the MAST-U, passing through both plasma-facing components and the central column region. A neutron flux map was simulated to estimate the radiation environment within the diagnostic field of view defined by a 15 cm diameter penetration through 1.5 m of concrete biological shielding acting as a collimator, together with an additional 10 cm polyethylene layer. Following installation, the collimated HXR diagnostic demonstrated reliable operation by recording gamma-ray events and hard x-ray bursts during both ohmic and neutral beam injection heated plasma discharges, confirming the successful commissioning of the diagnostic on MAST-U.

Authors

Institutions

Publication Details

Journal
Review of Scientific Instruments
Published
2026-11-01
DOI
https://doi.org/10.1063/5.0350509
Primary Topic
Radiation Detection and Scintillator Technologies
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Development and commissioning of a LaBr3:Ce, Sr-based collimated hard x-ray spectrometer for MAST-U

R. Worrall, P. Beaumont, R. Scannell, Z. Mianowska et al.
Review of Scientific Instruments
Radiation Detection and Scintillator Technologies
article

Development and commissioning of a LaBr3:Ce, Sr-based collimated hard x-ray spectrometer for MAST-U

R. Worrall, P. Beaumont, R. Scannell, Z. Mianowska, S. Mianowski, Zamir Ghani, Ben Phoenix, V. Kiptily, M. Beldishevski, MAST Upgrade Team, C. Downing, R. de Chazal, B. Kandan, E. Rose, K. Martens
article en

Abstract

A new collimated hard x-ray (HXR) spectrometer has been developed and installed on the Mega Amp Spherical Tokamak Upgrade (MAST-U) to support future studies of runaway electron generation and provide a diagnostic capability to plasma-facing components monitoring. The diagnostic is based on an Sr-codoped LaBr3:Ce scintillator (2″ × 2″) coupled to a fast data acquisition system capable of recording raw scintillation waveforms in both pulse and continuous modes, allowing detailed reconstruction of individual events using algorithms for optimization of spectrometric parameters and pulse-shape discrimination. The detector was energy-calibrated up to 4.8 MeV using standard spectroscopy sources and a 66Ga radioisotope. Its performance was then evaluated at the MC40 cyclotron (University of Birmingham), demonstrating stable operation at 1.1 Mcps, with gain shifts of less than 4%. Following successful validation, the diagnostic was installed using an existing midplane line-of-sight configuration of the MAST-U, passing through both plasma-facing components and the central column region. A neutron flux map was simulated to estimate the radiation environment within the diagnostic field of view defined by a 15 cm diameter penetration through 1.5 m of concrete biological shielding acting as a collimator, together with an additional 10 cm polyethylene layer. Following installation, the collimated HXR diagnostic demonstrated reliable operation by recording gamma-ray events and hard x-ray bursts during both ohmic and neutral beam injection heated plasma discharges, confirming the successful commissioning of the diagnostic on MAST-U.

Review of Scientific InstrumentsVol. 97(11)
Culham Centre for Fusion Energy (GB), United Kingdom Atomic Energy Authority (GB), University of Birmingham (GB)
Openalex Percentile: Top 14%
Radiation Detection and Scintillator Technologies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon