Automated verification of source dwell position in lunar ovoid applicators

BACKGROUND: Accurate reconstruction of the source dwell positions in high-dose-rate (HDR) brachytherapy is essential for treatment quality. The Venezia™ gynaecological applicator, with its lunar-shaped ovoids, poses challenges because conventional dummy markers do not match the actual source path, introducing systematic uncertainties. PURPOSE: The aim of this work was to develop and demonstrate an automated, reproducible, and open-source software for autoradiographic verification of dwell positions in Venezia lunar ovoids, and to compare measured positions with manufacturer-provided applicator library data. This pipeline established a standardized approach for clinical commissioning and quality assurance (QA) workflows. METHODS: A custom 3D-printed holder was designed to standardize autoradiograph acquisition using radiochromic films. An open-source Python pipeline, referred to as SPICA (Source Position Identification in Circular Applicators), was developed to process scanned films, isolate dwell position, and compute their coordinates with associated uncertainties. The code allows direct comparison between measured positions and manufacturer-provided applicator library data. We performed preliminary validation of the software by measuring the trajectory of an Ir-192 source for the 22 and 26 mm ovoids with a step size of 2.5 mm. RESULTS: Mean discrepancies between measured and reference positions were 1.2 ± 0.5 and 1.5 ± 0.4 mm for the 22 and 26 mm ovoids, respectively, with a maximum deviation of 2.6 mm. Distal positions showed better agreement. These findings confirm that, in the tested applicator-afterloader combination, vendor libraries provide a generally reliable baseline but may exhibit systematic differences. CONCLUSIONS: The proposed methodology offers a practical, reproducible solution to improve source path verification while reducing manual intervention. It supports more robust commissioning and QA of complex applicators and using the coordinates calculated in the treatment planning software can potentially reduce geometric uncertainties in clinical cases. Individualized commissioning remains essential for applicators with complex geometries.

Authors

Institutions

Publication Details

Journal
Journal of Applied Clinical Medical Physics
Published
2026-10-30
DOI
https://doi.org/10.1002/acm2.70887
Primary Topic
Advanced Radiotherapy Techniques
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Automated verification of source dwell position in lunar ovoid applicators

Gerard Trias, Cristian Candela‐Juan, Adrià Casamitjana, Eduard Agustí-Camprubí et al.
Journal of Applied Clinical Medical Physics
Advanced Radiotherapy Techniques
article

Automated verification of source dwell position in lunar ovoid applicators

Gerard Trias, Cristian Candela‐Juan, Adrià Casamitjana, Eduard Agustí-Camprubí, Antonio Herreros, Miquel Gubau‐Besalú, Gislè Segura‐Roca
article en

Abstract

BACKGROUND: Accurate reconstruction of the source dwell positions in high-dose-rate (HDR) brachytherapy is essential for treatment quality. The Venezia™ gynaecological applicator, with its lunar-shaped ovoids, poses challenges because conventional dummy markers do not match the actual source path, introducing systematic uncertainties. PURPOSE: The aim of this work was to develop and demonstrate an automated, reproducible, and open-source software for autoradiographic verification of dwell positions in Venezia lunar ovoids, and to compare measured positions with manufacturer-provided applicator library data. This pipeline established a standardized approach for clinical commissioning and quality assurance (QA) workflows. METHODS: A custom 3D-printed holder was designed to standardize autoradiograph acquisition using radiochromic films. An open-source Python pipeline, referred to as SPICA (Source Position Identification in Circular Applicators), was developed to process scanned films, isolate dwell position, and compute their coordinates with associated uncertainties. The code allows direct comparison between measured positions and manufacturer-provided applicator library data. We performed preliminary validation of the software by measuring the trajectory of an Ir-192 source for the 22 and 26 mm ovoids with a step size of 2.5 mm. RESULTS: Mean discrepancies between measured and reference positions were 1.2 ± 0.5 and 1.5 ± 0.4 mm for the 22 and 26 mm ovoids, respectively, with a maximum deviation of 2.6 mm. Distal positions showed better agreement. These findings confirm that, in the tested applicator-afterloader combination, vendor libraries provide a generally reliable baseline but may exhibit systematic differences. CONCLUSIONS: The proposed methodology offers a practical, reproducible solution to improve source path verification while reducing manual intervention. It supports more robust commissioning and QA of complex applicators and using the coordinates calculated in the treatment planning software can potentially reduce geometric uncertainties in clinical cases. Individualized commissioning remains essential for applicators with complex geometries.

Journal of Applied Clinical Medical PhysicsVol. 27(11)
Universitat de Girona (ES), Hospital Clínic de Barcelona (ES), Consorci Institut D'Investigacions Biomediques August Pi I Sunyer (ES), Fundació de Recerca Clínic Barcelona-Institut d’Investigacions Biomèdiques August Pi i Sunyer (ES), Universitat de Barcelona (ES)
Openalex Percentile: Top 14%
Advanced Radiotherapy Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon