Subcutaneous Defibrillator for Primary Prevention in Chronic Kidney Disease Stage 3–5: The Single‐Arm STEADIED Study

BACKGROUND: Prospective data on the subcutaneous implantable cardioverter-defibrillator (S-ICD) in chronic kidney disease (CKD) are lacking. We estimated the rate of appropriate ICD therapy with the S-ICD in CKD stage 3-5 with left ventricular ejection fraction (LVEF) ≤ 35%. METHODS: In this single-arm exploratory study, adults with CKD stage 3-5 and a guideline-based primary-prevention indication received an EMBLEM MRI S-ICD at eight Korean centers. The primary endpoint, appropriate ICD therapy, was hypothesized at 18% versus a historical rate of 7.8% (α = 0.05, 80% power, planned N = 77; exact binomial test). RESULTS: ; 71.8% had CKD stage 5 and 69.2% were on hemodialysis. Over a median follow-up of 361 days, appropriate ICD therapy occurred in 2 patients (5.1%; 95% CI, 1.4-16.9), significantly below the hypothesized 18% (p = 0.036). Inappropriate shocks occurred in 5 patients (12.8%; 95% CI, 5.6-26.7; 8 shocks), a proportion 2.5 times that of appropriate therapy, from external electromagnetic interference (2 patients), multiple counting after automatic SMART Pass deactivation (2), and supraventricular misclassification (1). Three deaths (7.7%) occurred, none device-related or documented as arrhythmic (one undetermined); all hospitalizations were for heart failure, and no systemic device infection occurred. CONCLUSIONS: In this exploratory single-arm study, appropriate ICD therapy was infrequent and exceeded by inappropriate shocks, while non-arrhythmic competing mortality dominated the clinical course. These hypothesis-generating findings support cautious patient selection, verification of filter status during follow-up, and larger comparative studies. TRIAL REGISTRATION: Clinical Research Information Service (CRIS), Republic of Korea; identifier KCT0005596.

Authors

Institutions

Publication Details

Journal
Pacing and Clinical Electrophysiology
Published
2026-10-31
DOI
https://doi.org/10.1111/pace.70489
Primary Topic
Cardiac pacing and defibrillation studies
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Subcutaneous Defibrillator for Primary Prevention in Chronic Kidney Disease Stage 3–5: The Single‐Arm STEADIED Study

Youngjun Wi, Eue‐Keun Choi, Jongmin Hwang, Tae-Wan Chung et al.
Pacing and Clinical Electrophysiology
Cardiac pacing and defibrillation studies
article

Subcutaneous Defibrillator for Primary Prevention in Chronic Kidney Disease Stage 3–5: The Single‐Arm STEADIED Study

Youngjun Wi, Eue‐Keun Choi, Jongmin Hwang, Tae-Wan Chung, Min Soo Cho, Il‐Young Oh, Seongwook Han, Boyoung Joung, Seung‐Jung Park, Hyoung‐Seob Park, Tae‐Hoon Kim, Min-Su Jung, Jong Sung Park, Myung‐Hwan Bae, Jong‐Sung Park
article en

Abstract

BACKGROUND: Prospective data on the subcutaneous implantable cardioverter-defibrillator (S-ICD) in chronic kidney disease (CKD) are lacking. We estimated the rate of appropriate ICD therapy with the S-ICD in CKD stage 3-5 with left ventricular ejection fraction (LVEF) ≤ 35%. METHODS: In this single-arm exploratory study, adults with CKD stage 3-5 and a guideline-based primary-prevention indication received an EMBLEM MRI S-ICD at eight Korean centers. The primary endpoint, appropriate ICD therapy, was hypothesized at 18% versus a historical rate of 7.8% (α = 0.05, 80% power, planned N = 77; exact binomial test). RESULTS: ; 71.8% had CKD stage 5 and 69.2% were on hemodialysis. Over a median follow-up of 361 days, appropriate ICD therapy occurred in 2 patients (5.1%; 95% CI, 1.4-16.9), significantly below the hypothesized 18% (p = 0.036). Inappropriate shocks occurred in 5 patients (12.8%; 95% CI, 5.6-26.7; 8 shocks), a proportion 2.5 times that of appropriate therapy, from external electromagnetic interference (2 patients), multiple counting after automatic SMART Pass deactivation (2), and supraventricular misclassification (1). Three deaths (7.7%) occurred, none device-related or documented as arrhythmic (one undetermined); all hospitalizations were for heart failure, and no systemic device infection occurred. CONCLUSIONS: In this exploratory single-arm study, appropriate ICD therapy was infrequent and exceeded by inappropriate shocks, while non-arrhythmic competing mortality dominated the clinical course. These hypothesis-generating findings support cautious patient selection, verification of filter status during follow-up, and larger comparative studies. TRIAL REGISTRATION: Clinical Research Information Service (CRIS), Republic of Korea; identifier KCT0005596.

Pacing and Clinical Electrophysiology
Seoul National University (KR), Yonsei University (KR), Severance Hospital (KR), Dong-A University Hospital (KR), Kyungpook National University Hospital (KR), Asan Medical Center (KR), Samsung Medical Center (KR), Seoul National University Hospital (KR), Seoul National University Bundang Hospital (KR), University of Ulsan (KR), Kyungpook National University Medical Center (KR), Keimyung University Daegu Dongsan Hospital (KR), Keimyung University Dongsan Hospital (KR), Dong-A University (KR), Sungkyunkwan University (KR)
Openalex Percentile: Top 11%
Cardiac pacing and defibrillation studies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon