Shockingly Unreliable: A Case Report of Failure of Synchronized Cardioversion Device to Capture, Resulting in Ventricular Fibrillation and Cardiac Arrest

Synchronized cardioversion is a standard treatment for unstable ventricular tachycardia. While modern defibrillators are designed with safety mechanisms, synchronization failure can result in catastrophic complications. We present the case of a 65-year-old woman with unstable wide-complex tachycardia with a pulse who underwent emergent synchronized cardioversion. Despite activation of the “sync” mode, the device failed to capture appropriately and delivered a shock on the T-wave, precipitating ventricular fibrillation and cardiac arrest. After 26 minutes of resuscitation and multiple defibrillations, return of spontaneous circulation (ROSC) was achieved. The patient ultimately recovered and was discharged home in stable condition. This case underscores the rare but serious risk of synchronization failure, particularly in obese patients where pad placement and capture may be impaired,1 and emphasizes the importance of visually confirming synchronization prior to shock delivery.

Authors

Publication Details

Journal
Journal of Education and Teaching in Emergency Medicine
Published
2026-10-31
DOI
https://doi.org/10.5070/m5.53094
Primary Topic
Cardiac Arrest and Resuscitation
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Shockingly Unreliable: A Case Report of Failure of Synchronized Cardioversion Device to Capture, Resulting in Ventricular Fibrillation and Cardiac Arrest

Andrew L. Juergens, Clifford J. Buckley, Bradley J McClintic, Shea Standley
Journal of Education and Teaching in Emergency Medicine
Cardiac Arrest and Resuscitation
article

Shockingly Unreliable: A Case Report of Failure of Synchronized Cardioversion Device to Capture, Resulting in Ventricular Fibrillation and Cardiac Arrest

Andrew L. Juergens, Clifford J. Buckley, Bradley J McClintic, Shea Standley
article en

Abstract

Synchronized cardioversion is a standard treatment for unstable ventricular tachycardia. While modern defibrillators are designed with safety mechanisms, synchronization failure can result in catastrophic complications. We present the case of a 65-year-old woman with unstable wide-complex tachycardia with a pulse who underwent emergent synchronized cardioversion. Despite activation of the “sync” mode, the device failed to capture appropriately and delivered a shock on the T-wave, precipitating ventricular fibrillation and cardiac arrest. After 26 minutes of resuscitation and multiple defibrillations, return of spontaneous circulation (ROSC) was achieved. The patient ultimately recovered and was discharged home in stable condition. This case underscores the rare but serious risk of synchronization failure, particularly in obese patients where pad placement and capture may be impaired,1 and emphasizes the importance of visually confirming synchronization prior to shock delivery.

Journal of Education and Teaching in Emergency MedicineVol. 11(4)
Openalex Percentile: Top 10%
Cardiac Arrest and Resuscitation
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon
Shockingly Unreliable: A Case Report of Failure of Synchronized Cardioversion Device to Capture, Resulting in Ventricular Fibrillation and Cardiac Arrest — Andrew L. Juergens, Clifford J. Buckley, et al. · Journal of Education and Teaching in Emergency Medicine (2026) | TGRS Research Map | TGRS