A multicenter end‐to‐end postal audit of skin markerless surface‐guided radiation therapy

BACKGROUND: Surface-guided radiation therapy (SGRT) enables skin markerless positioning, but its accuracy may depend on the reference surface, hardware configuration, commissioning, and institutional workflow. Two reference surfaces are commonly used: a DICOM reference generated from planning CT images and an SGRT reference acquired in the treatment room. Multicenter end-to-end data across commercial systems remain limited. PURPOSE: To evaluate static irradiation accuracy and setup time with two reference surfaces in a nationwide multicenter skin markerless SGRT postal audit. METHODS: A prospective end-to-end postal audit was conducted from October 2023 to July 2025 at 40 Japanese institutions using five commercial SGRT systems (AlignRT, Catalyst, ExacTrac Dynamic, IDENTIFY, and VOXELAN), eight per system. Each institution received a common phantom package: a breast-shaped surface phantom, two cubic phantoms containing gold markers, and radiochromic films. Markerless setup and irradiation were performed with both the DICOM and SGRT references; ExacTrac Dynamic institutions performed the DICOM reference setup only. Static irradiation accuracy was the difference between the film-derived radiation center and the intended center in the DICOM RT Plan; the primary metric was the three-dimensional translational vector error (3D vector error). Paired comparisons used the Wilcoxon signed-rank test, and multivariable linear regression evaluated factors associated with both endpoints. RESULTS: Critical setup errors exceeding 5 mm occurred at four institutions initially (institution-level error rate, 10.0%; three with the DICOM and one with the SGRT reference); one DICOM reference setup was not remeasured and remained out of tolerance, and was excluded as an outlier, leaving 39 DICOM and 31 SGRT reference setups for analysis. Excluding ExacTrac Dynamic, the DICOM reference showed a significantly larger 3D vector error than the SGRT reference (1.54 ± 0.91 vs. 1.20 ± 0.61 mm, p = 0.043). Setup time did not differ between references (86.19 ± 43.57 vs. 81.61 ± 37.09 s, p = 0.804). Axis-specifically, the difference between references was confined to the anterior-posterior direction (mean absolute difference 0.36 mm, p = 0.002), whereas a caudal superior-inferior offset was present with both (-0.50 ± 1.29 and -0.37 ± 0.87 mm; both p < 0.01) and did not differ between them (p = 0.854). The 1-camera condition was independently associated with a larger 3D vector error than the 3-camera condition (β = 0.90 mm; 95% CI, 0.17-1.63 mm; p = 0.017); no factor was associated with setup time, which showed no correlation with 3D vector error. CONCLUSIONS: Skin markerless SGRT achieved millimeter-level static irradiation accuracy but carried a non-negligible risk of critical setup errors. The difference between the two reference workflows was confined to the anterior-posterior direction, whereas a superior-inferior offset common to both indicates a residual uncertainty in the audit chain itself. Reference surface selection and camera configuration influenced accuracy, whereas setup time was not a reliable surrogate. These findings highlight the importance of commissioning, standardized reference surface acquisition, routine QA, and independent third-party audits.

Authors

Institutions

Publication Details

Journal
Journal of Applied Clinical Medical Physics
Published
2026-10-29
DOI
https://doi.org/10.1002/acm2.70893
Primary Topic
Advanced Radiotherapy Techniques
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

A multicenter end‐to‐end postal audit of skin markerless surface‐guided radiation therapy

Yasuhiro Ueshima, Masahiko Kurooka, Tomoki Kitagawa, Ryohei Yamauchi et al.
Journal of Applied Clinical Medical Physics
Advanced Radiotherapy Techniques
article

A multicenter end‐to‐end postal audit of skin markerless surface‐guided radiation therapy

Yasuhiro Ueshima, Masahiko Kurooka, Tomoki Kitagawa, Ryohei Yamauchi, Naoki Tohyama, Yoshifumi Oku, Yoshinobu Shimohigashi, Takayuki Kanai, Makoto Sasaki, Hidekazu Suzuki, Chie Kurokawa, Shinobu Kumagai, Masahide Saito, Hiroshi Onishi, Hironori Kojima, Naoki Hayashi, Eriko Saito, Masataka Sakamoto, Tatsunori Saito, Takeshi Ohno, Hiroki Katayama
article en

Abstract

BACKGROUND: Surface-guided radiation therapy (SGRT) enables skin markerless positioning, but its accuracy may depend on the reference surface, hardware configuration, commissioning, and institutional workflow. Two reference surfaces are commonly used: a DICOM reference generated from planning CT images and an SGRT reference acquired in the treatment room. Multicenter end-to-end data across commercial systems remain limited. PURPOSE: To evaluate static irradiation accuracy and setup time with two reference surfaces in a nationwide multicenter skin markerless SGRT postal audit. METHODS: A prospective end-to-end postal audit was conducted from October 2023 to July 2025 at 40 Japanese institutions using five commercial SGRT systems (AlignRT, Catalyst, ExacTrac Dynamic, IDENTIFY, and VOXELAN), eight per system. Each institution received a common phantom package: a breast-shaped surface phantom, two cubic phantoms containing gold markers, and radiochromic films. Markerless setup and irradiation were performed with both the DICOM and SGRT references; ExacTrac Dynamic institutions performed the DICOM reference setup only. Static irradiation accuracy was the difference between the film-derived radiation center and the intended center in the DICOM RT Plan; the primary metric was the three-dimensional translational vector error (3D vector error). Paired comparisons used the Wilcoxon signed-rank test, and multivariable linear regression evaluated factors associated with both endpoints. RESULTS: Critical setup errors exceeding 5 mm occurred at four institutions initially (institution-level error rate, 10.0%; three with the DICOM and one with the SGRT reference); one DICOM reference setup was not remeasured and remained out of tolerance, and was excluded as an outlier, leaving 39 DICOM and 31 SGRT reference setups for analysis. Excluding ExacTrac Dynamic, the DICOM reference showed a significantly larger 3D vector error than the SGRT reference (1.54 ± 0.91 vs. 1.20 ± 0.61 mm, p = 0.043). Setup time did not differ between references (86.19 ± 43.57 vs. 81.61 ± 37.09 s, p = 0.804). Axis-specifically, the difference between references was confined to the anterior-posterior direction (mean absolute difference 0.36 mm, p = 0.002), whereas a caudal superior-inferior offset was present with both (-0.50 ± 1.29 and -0.37 ± 0.87 mm; both p < 0.01) and did not differ between them (p = 0.854). The 1-camera condition was independently associated with a larger 3D vector error than the 3-camera condition (β = 0.90 mm; 95% CI, 0.17-1.63 mm; p = 0.017); no factor was associated with setup time, which showed no correlation with 3D vector error. CONCLUSIONS: Skin markerless SGRT achieved millimeter-level static irradiation accuracy but carried a non-negligible risk of critical setup errors. The difference between the two reference workflows was confined to the anterior-posterior direction, whereas a superior-inferior offset common to both indicates a residual uncertainty in the audit chain itself. Reference surface selection and camera configuration influenced accuracy, whereas setup time was not a reliable surrogate. These findings highlight the importance of commissioning, standardized reference surface acquisition, routine QA, and independent third-party audits.

Journal of Applied Clinical Medical PhysicsVol. 27(11)
Hamamatsu University (JP), Fujita Health University (JP), Kagawa University (JP), Hamamatsu University School of Medicine (JP), Juntendo University (JP), St. Luke's International Hospital (JP), Teikyo University Hospital (JP), Kanazawa University Hospital (JP), Tokyo Medical University Hospital (JP), University of Yamanashi Hospital (JP), Kagoshima University Hospital (JP), Kumamoto University Hospital (JP), Seirei Hamamatsu General Hospital (JP), Aichi Cancer Center (JP), Kagawa University Hospital (JP), Tokyo Women's Medical University (JP), Kyoto University Hospital (JP), Chiba Tokushukai Hospital (JP), Aichi Cancer Center Hospital (JP), University of Yamanashi (JP), Komazawa University (JP), Teikyo University (JP), Kumamoto University (JP)
Openalex Percentile: Top 14%
Advanced Radiotherapy Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon