Dual-channel surface-enhanced Raman immunoassay for simultaneous determination of atrial-fibrillation-associated NT-proBNP and high-sensitivity cardiac troponin I in artificial serum
. Across matrix lots and concentration levels, recoveries were 93.1-107.8% and total coefficients of variation were 4.8-9.4%. Thirty-six independently prepared, blinded artificial-serum specimens representing four dual-biomarker phenotypes were compared with electrochemiluminescence and chemiluminescent microparticle immunoassays. Under an equal-error-variance assumption, Deming slopes were 0.963 and 1.012, concordance correlation coefficients were 0.9951 and 0.9930, and concentration-quadrant agreement was 35/36. Grouped validation and specimen-level comparison assessed the dual readout without treating spatial spectra as independent replicates. This architecture provides a controlled route to simultaneous cardiac-biomarker measurement in artificial serum.
Authors
- Zhengke Ma
- Lijun Qin (ORCID: https://orcid.org/0009-0007-8006-7115)
- Yan Guo
- Yanhong Mou
Institutions
- Lanzhou University Second Hospital (CN)
Publication Details
- Journal
- Analytical Methods
- Published
- 2026-11-26
- DOI
- https://doi.org/10.1039/d6ay02032a
- Primary Topic
- Gold and Silver Nanoparticles Synthesis and Applications
- Type
- article
- Field-Weighted Citation Impact
- 0.00
Funders
- Lanzhou University