Size and Attenuation Measurements for Urinary Stone on Virtual Monoenergetic Images Derived From Dual Layer Spectral CT: A Phantom Study

OBJECTIVE: This study aimed to investigate the optimal energy level for virtual monoenergetic (VM) images with specific window settings for stone size and attenuation measurements, and to determine the effects of radiation dose and reconstructed slice thickness on these measurements. METHODS: Forty-eight urinary calcium oxalate stones (<1 cm in diameter) were placed within a 32-cm phantom, and CT scans were performed on a dual-layer detector scanner at 3 radiation doses (volumetric computed tomography dose index=5, 10, and 15 mGy). Images were reconstructed as conventional 120-kVp and VM images ranging from 40 to 100 keV in 10-keV increments with a slice thickness of 1 mm. In addition, 70-keV images were reconstructed at slice thicknesses of 2 and 5 mm. Length, width, and attenuation were measured at bone and soft tissue window settings by 2 radiologists. The reference stone size was measured with a digital caliper, and the reference attenuation was measured on 120 kVp images. Statistical analyses were performed using the paired t test, the Wilcoxon test, and repeated-measures one-way analysis of variance. RESULTS: The stone size and attenuation decreased with increasing energy level of the VM images, irrespective of window settings. The stone size on the VM 70- to 90-keV images under the soft tissue window and on the VM 40-keV images under the bone window was closest to the true size. Attenuation values measured on the VM 70-keV images were similar to those obtained on the conventional 120-kVp images across both window settings. In addition, measurements of stone size (length and width) and attenuation on VM images with thinner slices were closer to the true value. CONCLUSIONS: For VM images used to measure urinary stone size and attenuation, the optimal energy level required may differ based on the window settings. In addition, reconstructed thinner slice thickness can improve measurement accuracy.

Authors

Institutions

Publication Details

Journal
Journal of Computer Assisted Tomography
Published
2026-10-30
DOI
https://doi.org/10.1097/rct.0000000000001936
Primary Topic
Advanced X-ray and CT Imaging
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Size and Attenuation Measurements for Urinary Stone on Virtual Monoenergetic Images Derived From Dual Layer Spectral CT: A Phantom Study

Yannan Cheng, Jianxin Guo, Le Cao, Lu Zhang et al.
1 links
Journal of Computer Assisted Tomography
Advanced X-ray and CT Imaging
article

Size and Attenuation Measurements for Urinary Stone on Virtual Monoenergetic Images Derived From Dual Layer Spectral CT: A Phantom Study

Yannan Cheng, Jianxin Guo, Le Cao, Lu Zhang, Ganglian Fan, Huan Wang, Xiaoyue Zhang, Yanfei Cheng, Yanan Li, Tianze Wang
article en

Abstract

OBJECTIVE: This study aimed to investigate the optimal energy level for virtual monoenergetic (VM) images with specific window settings for stone size and attenuation measurements, and to determine the effects of radiation dose and reconstructed slice thickness on these measurements. METHODS: Forty-eight urinary calcium oxalate stones (<1 cm in diameter) were placed within a 32-cm phantom, and CT scans were performed on a dual-layer detector scanner at 3 radiation doses (volumetric computed tomography dose index=5, 10, and 15 mGy). Images were reconstructed as conventional 120-kVp and VM images ranging from 40 to 100 keV in 10-keV increments with a slice thickness of 1 mm. In addition, 70-keV images were reconstructed at slice thicknesses of 2 and 5 mm. Length, width, and attenuation were measured at bone and soft tissue window settings by 2 radiologists. The reference stone size was measured with a digital caliper, and the reference attenuation was measured on 120 kVp images. Statistical analyses were performed using the paired t test, the Wilcoxon test, and repeated-measures one-way analysis of variance. RESULTS: The stone size and attenuation decreased with increasing energy level of the VM images, irrespective of window settings. The stone size on the VM 70- to 90-keV images under the soft tissue window and on the VM 40-keV images under the bone window was closest to the true size. Attenuation values measured on the VM 70-keV images were similar to those obtained on the conventional 120-kVp images across both window settings. In addition, measurements of stone size (length and width) and attenuation on VM images with thinner slices were closer to the true value. CONCLUSIONS: For VM images used to measure urinary stone size and attenuation, the optimal energy level required may differ based on the window settings. In addition, reconstructed thinner slice thickness can improve measurement accuracy.

Journal of Computer Assisted Tomography
First Affiliated Hospital of Xi'an Jiaotong University (CN), Neurology, Inc (US), Xi'an Jiaotong University (CN)
Good health and well-being
Openalex Percentile: Top 25%
Advanced X-ray and CT Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon