Diagnostic accuracy of three CBCT protocols and Metal Artifact Reduction tool for detecting the second canal of mandibular incisors adjacent to a zirconia implant

OBJECTIVE: To compare the diagnostic accuracy of three cone beam computed tomography (CBCT) acquisition protocols, one low-resolution (LR) and two high-resolution (HR) protocols, using a metal artifact reduction tool (MAR) for detecting a second canal in mandibular incisors adjacent to a zirconia implant. MATERIAL AND METHODS: Twenty single-rooted mandibular incisors were included, twelve with one canal and eight with two canals confirmed by micro-CT. The roots and a zirconia implant were placed into the alveolar sockets of the right and left mandibular central incisors of a dry human mandible, respectively. In total, 240 CBCT scans were obtained using FOV 5x5 cm, 90 kVp, with and without implant, for each root. The variable parameters included the MAR tool, one LR-protocol (0.2 mm voxel, 5 mA), and two HR-protocols (0.085 mm voxel, 4 and 10 mA). Five calibrated examiners independently evaluated the scans. An ordinal cumulative link mixed model (CLMM) with crossed random intercepts for roots and examiners was used for data analysis (α = 5%). RESULTS: The CLMM revealed a significant main effect for implant presence (p = 0.049) and canal status (p < 0.001), but not for protocol (p = 0.135) or MAR (p = 0.392). Significant protocol x canal status (p < 0.001) and implant x canal status (p < 0.001) interactions were found, whereas MAR interactions were not significant (p > 0.05). In two-canal roots, the zirconia implant significantly induced false-negative shifts under HR-4 (p < 0.0001) and LR (p < 0.0001) protocols, whereas HR-10 preserved sensitivity (p = 0.7760). In single-canal roots, implants induced false-positive shifts in HR-10 (p = 0.0011) and HR-4 (p = 0.0011). CONCLUSION: Zirconia implants impair second canal detection via false-negative and false-positive errors. High-resolution CBCT (HR-10) should be prioritized to preserve diagnostic sensitivity, whereas MAR offers no clinical benefit and cannot compensate for reduced resolution or exposure. ADVANCES IN KNOWLEDGE: CBCT diagnostic performance near zirconia implants relies on acquisition protocol rather than MAR activation. High-resolution CBCT (HR-10) mitigates implant-induced false negatives, whereas lower-resolution or lower-milliamperage protocols compromise diagnostic accuracy.

Authors

Institutions

Publication Details

Journal
Dentomaxillofacial Radiology
Published
2026-10-30
DOI
https://doi.org/10.1093/dmfr/twag070
Primary Topic
Dental Radiography and Imaging
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Diagnostic accuracy of three CBCT protocols and Metal Artifact Reduction tool for detecting the second canal of mandibular incisors adjacent to a zirconia implant

Deborah Queiroz Freitas, Débora Costa Ruiz, Renata Maíra de Souza Leal, Thaísa Pinheiro Silva et al.
3 links
Dentomaxillofacial Radiology
Dental Radiography and Imaging
article

Diagnostic accuracy of three CBCT protocols and Metal Artifact Reduction tool for detecting the second canal of mandibular incisors adjacent to a zirconia implant

Deborah Queiroz Freitas, Débora Costa Ruiz, Renata Maíra de Souza Leal, Thaísa Pinheiro Silva, Maria Fernanda Silva Andrade‐Bortoletto, Marcela Tarosso Réa
article en

Abstract

OBJECTIVE: To compare the diagnostic accuracy of three cone beam computed tomography (CBCT) acquisition protocols, one low-resolution (LR) and two high-resolution (HR) protocols, using a metal artifact reduction tool (MAR) for detecting a second canal in mandibular incisors adjacent to a zirconia implant. MATERIAL AND METHODS: Twenty single-rooted mandibular incisors were included, twelve with one canal and eight with two canals confirmed by micro-CT. The roots and a zirconia implant were placed into the alveolar sockets of the right and left mandibular central incisors of a dry human mandible, respectively. In total, 240 CBCT scans were obtained using FOV 5x5 cm, 90 kVp, with and without implant, for each root. The variable parameters included the MAR tool, one LR-protocol (0.2 mm voxel, 5 mA), and two HR-protocols (0.085 mm voxel, 4 and 10 mA). Five calibrated examiners independently evaluated the scans. An ordinal cumulative link mixed model (CLMM) with crossed random intercepts for roots and examiners was used for data analysis (α = 5%). RESULTS: The CLMM revealed a significant main effect for implant presence (p = 0.049) and canal status (p < 0.001), but not for protocol (p = 0.135) or MAR (p = 0.392). Significant protocol x canal status (p < 0.001) and implant x canal status (p < 0.001) interactions were found, whereas MAR interactions were not significant (p > 0.05). In two-canal roots, the zirconia implant significantly induced false-negative shifts under HR-4 (p < 0.0001) and LR (p < 0.0001) protocols, whereas HR-10 preserved sensitivity (p = 0.7760). In single-canal roots, implants induced false-positive shifts in HR-10 (p = 0.0011) and HR-4 (p = 0.0011). CONCLUSION: Zirconia implants impair second canal detection via false-negative and false-positive errors. High-resolution CBCT (HR-10) should be prioritized to preserve diagnostic sensitivity, whereas MAR offers no clinical benefit and cannot compensate for reduced resolution or exposure. ADVANCES IN KNOWLEDGE: CBCT diagnostic performance near zirconia implants relies on acquisition protocol rather than MAR activation. High-resolution CBCT (HR-10) mitigates implant-induced false negatives, whereas lower-resolution or lower-milliamperage protocols compromise diagnostic accuracy.

Dentomaxillofacial Radiology
Universidade Estadual de Campinas (UNICAMP) (BR), Universidade Federal do Paraná (BR)
Good health and well-being
Openalex Percentile: Top 10%
Dental Radiography and Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon