Equine chorionic gonadotropin (eCG/PMSG) is widely used for female-cycle synchronization and estrus induction in livestock and in biomedical research, where reliable quantitative assessment is increasingly important. However, its structural and glycosylation heterogeneity complicates antibody-based measurement and assay standardization. This Technical Note evaluated six commercially available quantitative PMSG ELISA kits representing three distinct analytical profile groups. Kit-provided calibration standards were characterized by Western blotting, Coomassie staining, and LC-MS/MS, while quantitative performance was assessed using serial dilutions of two commercial PMSG preparations. DRG, Creative Diagnostics, and Demeditec showed similar calibration-standard profiles and concentration-dependent responses, although measured concentrations were consistently below nominal values. Biomatik and MyBiosource displayed closely related analytical profiles but little meaningful concentration-dependent response, whereas Krishgen measurements remained within a narrow range despite changes in nominal PMSG input. The calibration standards differed substantially in protein composition, and no eCG-derived peptides were identified by LC-MS/MS under the conditions used. These findings demonstrate that quantitative results obtained with commercially available PMSG ELISAs are not necessarily comparable and may be strongly influenced by calibration-material composition, assay format, and matrix effects. The findings highlight the limitations of current quantitative PMSG assays and the need for a well-characterized, commutable reference material, together with assay-specific validation of dilution parallelism and analyte recovery, to improve the reliability and comparability of eCG/PMSG measurements.

Authors

Institutions

Publication Details

Journal
bioRxiv (Cold Spring Harbor Laboratory)
Published
2026-10-30
DOI
https://doi.org/10.64898/2026.10.29.762762
Primary Topic
Clinical Laboratory Practices and Quality Control
Type
preprint
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
preprint

Standardisation Challenges in ELISA-Based Quantification of Equine Chorionic Gonadotropin (eCG/PMSG)

Martina Lösle, Thorsten Buch, Polina A Mishchenko, Ricardo Matiz Martinez
2 links
bioRxiv (Cold Spring Harbor Laboratory)
Clinical Laboratory Practices and Quality Control
preprint

Standardisation Challenges in ELISA-Based Quantification of Equine Chorionic Gonadotropin (eCG/PMSG)

Martina Lösle, Thorsten Buch, Polina A Mishchenko, Ricardo Matiz Martinez
preprint en

Abstract

Equine chorionic gonadotropin (eCG/PMSG) is widely used for female-cycle synchronization and estrus induction in livestock and in biomedical research, where reliable quantitative assessment is increasingly important. However, its structural and glycosylation heterogeneity complicates antibody-based measurement and assay standardization. This Technical Note evaluated six commercially available quantitative PMSG ELISA kits representing three distinct analytical profile groups. Kit-provided calibration standards were characterized by Western blotting, Coomassie staining, and LC-MS/MS, while quantitative performance was assessed using serial dilutions of two commercial PMSG preparations. DRG, Creative Diagnostics, and Demeditec showed similar calibration-standard profiles and concentration-dependent responses, although measured concentrations were consistently below nominal values. Biomatik and MyBiosource displayed closely related analytical profiles but little meaningful concentration-dependent response, whereas Krishgen measurements remained within a narrow range despite changes in nominal PMSG input. The calibration standards differed substantially in protein composition, and no eCG-derived peptides were identified by LC-MS/MS under the conditions used. These findings demonstrate that quantitative results obtained with commercially available PMSG ELISAs are not necessarily comparable and may be strongly influenced by calibration-material composition, assay format, and matrix effects. The findings highlight the limitations of current quantitative PMSG assays and the need for a well-characterized, commutable reference material, together with assay-specific validation of dilution parallelism and analyte recovery, to improve the reliability and comparability of eCG/PMSG measurements.

bioRxiv (Cold Spring Harbor Laboratory)
University of Zurich (CH)
Clinical Laboratory Practices and Quality Control
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon