Coronal alignment-related mediolateral load distribution and polyethylene wear in a medial-pivot total knee arthroplasty design

INTRODUCTION: While mechanical alignment in total knee arthroplasty aims for a neutral hip-knee-ankle angle, kinematic alignment preserves the native joint line, often allowing residual varus or valgus. Clinically, constitutional variations in alignment affect the coronal joint moments and the mediolateral load distribution, but the impact on polyethylene wear remains unclear. MATERIALS AND METHODS: , implantcast GmbH, Germany) was evaluated in a knee wear simulator replicating the walking cycle. Inserts were tested under three conditions representing overall lower-limb coronal alignment and the associated mediolateral load distribution: neutral (60% medial / 40% lateral), varus (70% / 30%), and valgus (50% / 50%), with three (n = 3) dynamically loaded inserts and one load-soak control per condition. Testing followed ISO 14243-1:2009 over 5 million cycles. Gravimetric wear and wear patterns were assessed. RESULTS: A numerical trend was observed among the groups: The varus-related load distribution demonstrated the lowest overall wear (4.83 ± 1.06 mg/million cycles), whereas the valgus group produced the highest wear (9.43 ± 1.47 mg/million cycles); the neutral group was intermediate (6.81 ± 3.07 mg/million cycles). However, these differences did not reach statistical significance. Visual inspection revealed medial-dominant wear in all groups, with increased lateral wear under valgus-related load distribution. CONCLUSIONS: Overall lower-limb coronal alignment and the associated mediolateral load distribution appear to influence polyethylene wear in total knee arthroplasty. The observed numerical trend suggests that alignment-related changes in mediolateral load distribution may influence polyethylene wear patterns in this medial-pivot TKA design. Further studies are required to determine the clinical relevance of these observations.

Authors

Institutions

Publication Details

Journal
Archives of Orthopaedic and Trauma Surgery
Published
2026-10-30
DOI
https://doi.org/10.1007/s00402-026-06558-z
Primary Topic
Total Knee Arthroplasty Outcomes
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Coronal alignment-related mediolateral load distribution and polyethylene wear in a medial-pivot total knee arthroplasty design

Manuel Kistler, Thomas Richard Niethammer, Johanna‐Maria Simon, Boris Michael Holzapfel et al.
Archives of Orthopaedic and Trauma Surgery
Total Knee Arthroplasty Outcomes
article

Coronal alignment-related mediolateral load distribution and polyethylene wear in a medial-pivot total knee arthroplasty design

Manuel Kistler, Thomas Richard Niethammer, Johanna‐Maria Simon, Boris Michael Holzapfel, Matthias Woiczinski, Christoph Thorwächter, Celina Lesnicar, Irem Tural
article en

Abstract

INTRODUCTION: While mechanical alignment in total knee arthroplasty aims for a neutral hip-knee-ankle angle, kinematic alignment preserves the native joint line, often allowing residual varus or valgus. Clinically, constitutional variations in alignment affect the coronal joint moments and the mediolateral load distribution, but the impact on polyethylene wear remains unclear. MATERIALS AND METHODS: , implantcast GmbH, Germany) was evaluated in a knee wear simulator replicating the walking cycle. Inserts were tested under three conditions representing overall lower-limb coronal alignment and the associated mediolateral load distribution: neutral (60% medial / 40% lateral), varus (70% / 30%), and valgus (50% / 50%), with three (n = 3) dynamically loaded inserts and one load-soak control per condition. Testing followed ISO 14243-1:2009 over 5 million cycles. Gravimetric wear and wear patterns were assessed. RESULTS: A numerical trend was observed among the groups: The varus-related load distribution demonstrated the lowest overall wear (4.83 ± 1.06 mg/million cycles), whereas the valgus group produced the highest wear (9.43 ± 1.47 mg/million cycles); the neutral group was intermediate (6.81 ± 3.07 mg/million cycles). However, these differences did not reach statistical significance. Visual inspection revealed medial-dominant wear in all groups, with increased lateral wear under valgus-related load distribution. CONCLUSIONS: Overall lower-limb coronal alignment and the associated mediolateral load distribution appear to influence polyethylene wear in total knee arthroplasty. The observed numerical trend suggests that alignment-related changes in mediolateral load distribution may influence polyethylene wear patterns in this medial-pivot TKA design. Further studies are required to determine the clinical relevance of these observations.

Archives of Orthopaedic and Trauma SurgeryVol. 146(1)
Waldkrankenhaus Rudolf Elle (DE), Jena University Hospital (DE), Friedrich Schiller University Jena (DE), Ludwig-Maximilians-Universität München (DE)
Good health and well-being
Openalex Percentile: Top 10%
Total Knee Arthroplasty Outcomes
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon