Revisiting Artifact Suppression in Electrospray Ionization Deconvolution

Charge state assignment is a critical step in many workflows for intact protein analysis by mass spectrometry. Computational deconvolution algorithms are increasingly used to automate data analysis, but conventional approaches can present artifacts. Here, we explore several approaches to remove artifacts by limiting the charge state assignments for each data point. One approach removes satellite charge states surrounding local maxima. Another removes harmonics at double a more probable charge assignment. The last two limit either the total number of allowed charge state assignments or the minimum relative charge state assignment probability for a given data point. These new features are as effective as the prior SoftMax artifact removal in UniDec, but they enable easier tuning by the user. Although the new set of features introduces additional complexity and an expanded number of parameters, these tools enable users to more precisely reduce artifacts in deconvolution results.

Authors

Institutions

Publication Details

Journal
ChemRxiv
Published
2026-10-29
DOI
https://doi.org/10.26434/chemrxiv.15009546/v3
Primary Topic
Mass Spectrometry Techniques and Applications
Type
preprint

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
preprint

Revisiting Artifact Suppression in Electrospray Ionization Deconvolution

Michael Thomas Marty
1 links
ChemRxiv
Mass Spectrometry Techniques and Applications
preprint

Revisiting Artifact Suppression in Electrospray Ionization Deconvolution

Michael Thomas Marty
preprint en

Abstract

Charge state assignment is a critical step in many workflows for intact protein analysis by mass spectrometry. Computational deconvolution algorithms are increasingly used to automate data analysis, but conventional approaches can present artifacts. Here, we explore several approaches to remove artifacts by limiting the charge state assignments for each data point. One approach removes satellite charge states surrounding local maxima. Another removes harmonics at double a more probable charge assignment. The last two limit either the total number of allowed charge state assignments or the minimum relative charge state assignment probability for a given data point. These new features are as effective as the prior SoftMax artifact removal in UniDec, but they enable easier tuning by the user. Although the new set of features introduces additional complexity and an expanded number of parameters, these tools enable users to more precisely reduce artifacts in deconvolution results.

ChemRxiv
The University of Texas at Austin (US)
National Science Foundation, Amgen, Division of Chemistry
Mass Spectrometry Techniques and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon