Numerical and experimental assessment of the influence of thermocouple orientation on temperature measurements for radiative superficial hyperthermia

BACKGROUND: Radiative oncological hyperthermia (HT) is a medical therapy for cancer treatment aimed at inducing temperatures up to 40-44 °C for 1h in pathological tissues. This aim is achieved through the application of an electromagnetic (EM) field radiated by antennas located on the body surface. According to tumor location, hyperthermia is typically classified in deep or superficial HT. In both cases, temperature measurements are typically performed during treatment for verification purposes and in validation procedures to assess HT device performance. Temperature monitoring is typically carried out using thermocouples or thermistors, both of which transmit the information through a signal along metallic wires; when thermocouples are used, their readings are performed while discontinuing EM radiation to avoid electromagnetic interference. PURPOSE: This work aims to study the effect that temperature sensors systems have on the EM field and temperature distributions produced by the HT device. In particular, the effect of the orientation of the metallic wires was analyzed. METHODS: Experimental measurements and a numerical analysis were performed using two different setups, based on a phantom equipped with catheters for temperature probe insertion and a commercial hyperthermia antenna. RESULTS AND CONCLUSION: Results showed that wires oriented orthogonally to the electric field do not induce a significant change in the EM-temperature distributions. Instead, the presence of wires oriented parallel to the main electric field component induces a significant perturbation of the EM field and temperature distributions, with a reduction of the penetration depth. These results evidence that care must be taken when placing temperature sensors in HT procedures.

Authors

Institutions

Publication Details

Journal
International Journal of Hyperthermia
Published
2026-10-27
DOI
https://doi.org/10.1080/02656736.2026.2732523
Primary Topic
Ultrasound and Hyperthermia Applications
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Numerical and experimental assessment of the influence of thermocouple orientation on temperature measurements for radiative superficial hyperthermia

Peter Manser, Markus Fuerstner, D. Frauchiger, Marta Cavagnaro et al.
International Journal of Hyperthermia
Ultrasound and Hyperthermia Applications
article

Numerical and experimental assessment of the influence of thermocouple orientation on temperature measurements for radiative superficial hyperthermia

Peter Manser, Markus Fuerstner, D. Frauchiger, Marta Cavagnaro, Marco Di Cristofano
article en

Abstract

BACKGROUND: Radiative oncological hyperthermia (HT) is a medical therapy for cancer treatment aimed at inducing temperatures up to 40-44 °C for 1h in pathological tissues. This aim is achieved through the application of an electromagnetic (EM) field radiated by antennas located on the body surface. According to tumor location, hyperthermia is typically classified in deep or superficial HT. In both cases, temperature measurements are typically performed during treatment for verification purposes and in validation procedures to assess HT device performance. Temperature monitoring is typically carried out using thermocouples or thermistors, both of which transmit the information through a signal along metallic wires; when thermocouples are used, their readings are performed while discontinuing EM radiation to avoid electromagnetic interference. PURPOSE: This work aims to study the effect that temperature sensors systems have on the EM field and temperature distributions produced by the HT device. In particular, the effect of the orientation of the metallic wires was analyzed. METHODS: Experimental measurements and a numerical analysis were performed using two different setups, based on a phantom equipped with catheters for temperature probe insertion and a commercial hyperthermia antenna. RESULTS AND CONCLUSION: Results showed that wires oriented orthogonally to the electric field do not induce a significant change in the EM-temperature distributions. Instead, the presence of wires oriented parallel to the main electric field component induces a significant perturbation of the EM field and temperature distributions, with a reduction of the penetration depth. These results evidence that care must be taken when placing temperature sensors in HT procedures.

International Journal of HyperthermiaVol. 43(1)
University of Bern (CH), University Hospital of Bern (CH), Sapienza University of Rome (IT)
Openalex Percentile: Top 24%
Ultrasound and Hyperthermia Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon