Airway stenosis alters the balance between tidal volume, air entrainment, and oxygen-source delivery during normal frequency and superimposed jet ventilation: a bench model study

Background Severe airway stenosis alters gas mixing and volume delivery during open-system jet ventilation. How high-frequency (HF) superimposition and component titration modify these effects remains incompletely characterized. Objective To compare stenosis-induced changes in distal oxygen concentration (FiO 2 ), tidal volume (VT), pressure, flow, and estimated gas composition among pure normal-frequency jet ventilation (NFJV) and superimposed HF jet ventilation (SHFJV) with titration of either the NF (SHFJV-NF) or HF component (SHFJV-HF). Methods An open rigid-bronchoscopy bench model was tested with and without 90% fixed airway stenosis. Pure NFJV and SHFJV-NF shared 56 NF parameter settings; SHFJV-HF included 60 HF settings with a fixed NF component. Linear mixed-effects models evaluated strategy × airway interactions and Holm-adjusted pairwise contrasts. A separate matched model assessed HF addition. Gas-component fractions and volumes were estimated at each setting using a two-source mixing model. Results Strategy × airway interactions were significant for all five measured outcomes (all P < 0.001). Median stenosis-induced changes were +4.7, +29.1, and +24.7 percentage points for FiO 2 and −454, −577, and −232 mL for VT during pure NFJV, SHFJV-NF, and SHFJV-HF, respectively. At identical NF settings, HF addition modified the stenosis response by +24.1 percentage points for FiO 2 , −111.7 mL for VT (95% CI −173.6 to −49.8), +2.02 cmH 2 O for measured positive end-expiratory pressure (PEEP), and −8.11 L·min −1 for expiratory flow magnitude (all P < 0.001). Estimated entrained-air volume decreased in all strategies, whereas oxygen-source volume did not increase uniformly. Conclusion At identical NF settings, HF addition amplified the stenosis-associated FiO 2 rise, VT loss, and PEEP increase. HF-component titration produced a different response across its tested parameter range. Estimated gas-composition changes were consistent with reduced oxygen dilution. Distal FiO 2 did not establish preserved ventilation or arterial oxygenation, and CO 2 elimination was not measured.

Authors

Institutions

Publication Details

Journal
Frontiers in Medicine
Published
2026-10-28
DOI
https://doi.org/10.3389/fmed.2026.1910316
Primary Topic
Respiratory Support and Mechanisms
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Airway stenosis alters the balance between tidal volume, air entrainment, and oxygen-source delivery during normal frequency and superimposed jet ventilation: a bench model study

Zhuomin Deng, Qinghao Cheng, Mingyuan Yang, Shuwang Yang et al.
1 links
Frontiers in Medicine
Respiratory Support and Mechanisms
article

Airway stenosis alters the balance between tidal volume, air entrainment, and oxygen-source delivery during normal frequency and superimposed jet ventilation: a bench model study

Zhuomin Deng, Qinghao Cheng, Mingyuan Yang, Shuwang Yang, Jing Guo, Hong Li, Yuxue Yao
article en

Abstract

Background Severe airway stenosis alters gas mixing and volume delivery during open-system jet ventilation. How high-frequency (HF) superimposition and component titration modify these effects remains incompletely characterized. Objective To compare stenosis-induced changes in distal oxygen concentration (FiO 2 ), tidal volume (VT), pressure, flow, and estimated gas composition among pure normal-frequency jet ventilation (NFJV) and superimposed HF jet ventilation (SHFJV) with titration of either the NF (SHFJV-NF) or HF component (SHFJV-HF). Methods An open rigid-bronchoscopy bench model was tested with and without 90% fixed airway stenosis. Pure NFJV and SHFJV-NF shared 56 NF parameter settings; SHFJV-HF included 60 HF settings with a fixed NF component. Linear mixed-effects models evaluated strategy × airway interactions and Holm-adjusted pairwise contrasts. A separate matched model assessed HF addition. Gas-component fractions and volumes were estimated at each setting using a two-source mixing model. Results Strategy × airway interactions were significant for all five measured outcomes (all P < 0.001). Median stenosis-induced changes were +4.7, +29.1, and +24.7 percentage points for FiO 2 and −454, −577, and −232 mL for VT during pure NFJV, SHFJV-NF, and SHFJV-HF, respectively. At identical NF settings, HF addition modified the stenosis response by +24.1 percentage points for FiO 2 , −111.7 mL for VT (95% CI −173.6 to −49.8), +2.02 cmH 2 O for measured positive end-expiratory pressure (PEEP), and −8.11 L·min −1 for expiratory flow magnitude (all P < 0.001). Estimated entrained-air volume decreased in all strategies, whereas oxygen-source volume did not increase uniformly. Conclusion At identical NF settings, HF addition amplified the stenosis-associated FiO 2 rise, VT loss, and PEEP increase. HF-component titration produced a different response across its tested parameter range. Estimated gas-composition changes were consistent with reduced oxygen dilution. Distal FiO 2 did not establish preserved ventilation or arterial oxygenation, and CO 2 elimination was not measured.

Frontiers in MedicineVol. 13
Mindray (China) (CN)
Good health and well-being
Openalex Percentile: Top 12%
Respiratory Support and Mechanisms
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon