The Effects of Distance and Warming Temperature on Intensity of Extremely Low Frequency Electromagnetic Field from Forced-Air Warming System

Purpose. Electronic devices, increasingly used in operating rooms, heavily depend on mains electricity. This may significantly increase occupational exposure to extremely low frequency electromagnetic field. As a possible carcinogen as per International Agency for Research on Cancer, Swedish Confederation of Professional Employee recommended it should be below 2 mG. Forced air warming system is used to prevent hypothermia. This study was to investigate the effect of distance and warming temperature on extremely low frequency electromagnetic field intensity from Bair-Hugger™ Warming Unit, in order to find the minimum safe distance from the device. Materials and methods. The forced-air warming unit was hung on the side of an operating table in an empty operating room, with all other electronic devices turned off. With three different warming temperatures, electromagnetic field was measured at 15, 30, and 45 cm from the anterior surface of the device. Data was collected at a sampling time of 0.5 times per second for 10 minutes. Kruskal-Wallis one-way analysis was used. Results. Increasing warming temperature and reducing the distance from the device resulted in a significant increase in electromagnetic field intensity (p <0.05). At 45 cm in particular, low and medium warming temperatures showed significantly lower intensity compared to high temperature (p<0.05). Exposure below 2.0 mG was achieved only when it was at 45 cm regardless of temperature. Conclusions. Health workers should be aware of occupational exposure to extremely low frequency electromagnetic field and avoid as much as possible. It is suggested to stay at least 45 cm away from 3M Bair Hugger™ Warming Unit.

Authors

Institutions

Publication Details

Journal
Україна Здоров’я нації
Published
2026-10-28
DOI
https://doi.org/10.32782/2077-6594/2026.3/21
Primary Topic
Electromagnetic Fields and Biological Effects
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

The Effects of Distance and Warming Temperature on Intensity of Extremely Low Frequency Electromagnetic Field from Forced-Air Warming System

Soo Jee Lee, Yong Han Kim
1 links
Україна Здоров’я нації
Electromagnetic Fields and Biological Effects
article

The Effects of Distance and Warming Temperature on Intensity of Extremely Low Frequency Electromagnetic Field from Forced-Air Warming System

Soo Jee Lee, Yong Han Kim
article en

Abstract

Purpose. Electronic devices, increasingly used in operating rooms, heavily depend on mains electricity. This may significantly increase occupational exposure to extremely low frequency electromagnetic field. As a possible carcinogen as per International Agency for Research on Cancer, Swedish Confederation of Professional Employee recommended it should be below 2 mG. Forced air warming system is used to prevent hypothermia. This study was to investigate the effect of distance and warming temperature on extremely low frequency electromagnetic field intensity from Bair-Hugger™ Warming Unit, in order to find the minimum safe distance from the device. Materials and methods. The forced-air warming unit was hung on the side of an operating table in an empty operating room, with all other electronic devices turned off. With three different warming temperatures, electromagnetic field was measured at 15, 30, and 45 cm from the anterior surface of the device. Data was collected at a sampling time of 0.5 times per second for 10 minutes. Kruskal-Wallis one-way analysis was used. Results. Increasing warming temperature and reducing the distance from the device resulted in a significant increase in electromagnetic field intensity (p <0.05). At 45 cm in particular, low and medium warming temperatures showed significantly lower intensity compared to high temperature (p<0.05). Exposure below 2.0 mG was achieved only when it was at 45 cm regardless of temperature. Conclusions. Health workers should be aware of occupational exposure to extremely low frequency electromagnetic field and avoid as much as possible. It is suggested to stay at least 45 cm away from 3M Bair Hugger™ Warming Unit.

Україна Здоров’я нації(3)
Inje University (KR), Inje University Haeundae Paik Hospital (KR), Inje University Busan Paik Hospital (KR)
Inje University
Good health and well-being
Openalex Percentile: Top 20%
Electromagnetic Fields and Biological Effects
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon