Commutability assessment of candidate reference materials for total bilirubin assays in neonates

OBJECTIVES: Neonatal total bilirubin (TBIL) measurements are the cornerstone for treatment decisions in neonatal hyperbilirubinemia. However, neonatal TBIL assays suffer from substantial inter-assay differences and standardization is hampered by the lack of commutable secondary reference materials. Therefore, the aim of this study was to assess commutability of candidate reference materials (RMs) to support standardization efforts. METHODS: Three types of materials were investigated as candidate RMs: bilirubin-spiked serum materials prepared by the Dutch EQA provider SKML, residual material from neonatal exchange transfusions, and NIST SRM916b material dissolved in adult serum. Commutability of these materials was assessed using the prediction interval and difference in bias approaches. To this end, all candidate RMs and clinical samples were measured on ten different routine methods and compared to the reference measurement procedure. RESULTS: For three methods, commutability could not be assessed due to too high non-selectivity. The exchange transfusion material demonstrated the highest degree of commutability, the highest TBIL concentration being commutable for all methods. Also, the two SRM-spiked samples with the highest TBIL concentrations showed to be commutable for six out of seven selective methods, except for the Ortho Vitros XT3400 platform. CONCLUSIONS: Exchange transfusion materials provide a robust neonatal serum-based material which can be used as secondary reference material and as EQA material to support and monitor future efforts to improve standardization of neonatal TBIL assays.

Authors

Institutions

Publication Details

Journal
Clinical Chemistry and Laboratory Medicine (CCLM)
Published
2026-10-28
DOI
https://doi.org/10.1515/cclm-2026-1261
Primary Topic
Neonatal Health and Biochemistry
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Commutability assessment of candidate reference materials for total bilirubin assays in neonates

Miranda van Berkel, Antje Staaden, Marieke A. M. Frasa, Lauren E H Westenberg et al.
1 links
Clinical Chemistry and Laboratory Medicine (CCLM)
Neonatal Health and Biochemistry
article

Commutability assessment of candidate reference materials for total bilirubin assays in neonates

Miranda van Berkel, Antje Staaden, Marieke A. M. Frasa, Lauren E H Westenberg, Jenny E. Kootstra‐Ros, Andrei N. Tintu, Oostendorp Marlies, Vincent Delatour, Cas W. Weykamp, Paul W. Schenk, Denis Grote-Koska, Eline A.E. van der Hagen, Jolande Y. Vis, Stanley F. Lo, Moniek M. Bioch, Christian V. Hulzebos
article en

Abstract

OBJECTIVES: Neonatal total bilirubin (TBIL) measurements are the cornerstone for treatment decisions in neonatal hyperbilirubinemia. However, neonatal TBIL assays suffer from substantial inter-assay differences and standardization is hampered by the lack of commutable secondary reference materials. Therefore, the aim of this study was to assess commutability of candidate reference materials (RMs) to support standardization efforts. METHODS: Three types of materials were investigated as candidate RMs: bilirubin-spiked serum materials prepared by the Dutch EQA provider SKML, residual material from neonatal exchange transfusions, and NIST SRM916b material dissolved in adult serum. Commutability of these materials was assessed using the prediction interval and difference in bias approaches. To this end, all candidate RMs and clinical samples were measured on ten different routine methods and compared to the reference measurement procedure. RESULTS: For three methods, commutability could not be assessed due to too high non-selectivity. The exchange transfusion material demonstrated the highest degree of commutability, the highest TBIL concentration being commutable for all methods. Also, the two SRM-spiked samples with the highest TBIL concentrations showed to be commutable for six out of seven selective methods, except for the Ortho Vitros XT3400 platform. CONCLUSIONS: Exchange transfusion materials provide a robust neonatal serum-based material which can be used as secondary reference material and as EQA material to support and monitor future efforts to improve standardization of neonatal TBIL assays.

Clinical Chemistry and Laboratory Medicine (CCLM)
University Medical Center Groningen (NL), Radboud University Nijmegen (NL), Medical College of Wisconsin (US), Leiden University Medical Center (NL), Erasmus MC (NL), Radboud University Medical Center (NL), Medizinische Hochschule Hannover (DE), Laboratoire National de Métrologie et d'Essais (FR), Beatrix Kinderziekenhuis (NL), Streekziekenhuis Koningin Beatrix (NL), Reference Institute for Bioanalytics (DE), Erasmus MC - Sophia Children’s Hospital (NL), Rijnstate Hospital (NL), Jeroen Bosch Ziekenhuis (NL)
Good health and well-being
Openalex Percentile: Top 7%
Neonatal Health and Biochemistry
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon