Three-Dimensional Fluid-Structure Interaction Reveals Trade-offs Among Oxygen Delivery, Shear, and Washout in Pediatric Roller-Pump Cardiopulmonary Bypass

Objective: To determine how roller speed, tubing occlusion, and hematocrit jointly regulate oxygen delivery, shear, and washout in pediatric roller-pump cardiopulmonary bypass. Methods: A three-dimensional, closed-loop fluid-structure interaction model resolved roller-tube contact, tubing deformation, and blood flow. Forty simulations combined 4 roller speeds (38-95 revolutions per minute), 2 occlusion levels (77% and 83%), and 5 hematocrit levels (15%-50%). Cycle-averaged flow, oxygen delivery, peak shear stress, and viscosity-normalized relative residence time were quantified. Results: Roller compression produced retrograde leakage through the occluded gap, localized peak shear stress exceeding 100 pascals, and adjacent stagnation and recirculation. Flow increased nearly linearly with roller speed and was higher at greater occlusion, with little hematocrit dependence. Higher speed and occlusion increased peak shear stress while reducing relative residence time; higher hematocrit further increased peak shear stress. Similar oxygen delivery could arise from substantially different flow environments: among 10 conditions providing oxygen delivery of approximately 9-12 milliliters per minute, peak shear stress ranged from approximately 25 to 70 pascals. Within conditions satisfying an illustrative minimum oxygen-delivery threshold, lower hematocrit permitted speed and occlusion settings yielding lower peak shear stress and relative residence time. Conclusions: Pediatric roller-pump settings that provide similar oxygen delivery can expose blood to markedly different shear and washout environments. Joint evaluation of these metrics may guide selection of speed, occlusion, and hematocrit to meet a target oxygen delivery while reducing shear and improving washout. Keywords: Cardiopulmonary bypass; roller pump; pediatric cardiac surgery; fluid-structure interaction; shear stress; relative residence time

Authors

Institutions

Publication Details

Journal
bioRxiv (Cold Spring Harbor Laboratory)
Published
2026-10-28
DOI
https://doi.org/10.64898/2026.10.26.761989
Primary Topic
Mechanical Circulatory Support Devices
Type
preprint
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
preprint

Three-Dimensional Fluid-Structure Interaction Reveals Trade-offs Among Oxygen Delivery, Shear, and Washout in Pediatric Roller-Pump Cardiopulmonary Bypass

Yunpeng Tu, Christina L. Greene, Juan C. del Álamo, Kevin A. Charette et al.
1 links
bioRxiv (Cold Spring Harbor Laboratory)
Mechanical Circulatory Support Devices
preprint

Three-Dimensional Fluid-Structure Interaction Reveals Trade-offs Among Oxygen Delivery, Shear, and Washout in Pediatric Roller-Pump Cardiopulmonary Bypass

Yunpeng Tu, Christina L. Greene, Juan C. del Álamo, Kevin A. Charette, Yi‐Ting Yeh, Vishal Nigam, Lyubomyr Bohuta, Alex De Lecea
preprint en

Abstract

Objective: To determine how roller speed, tubing occlusion, and hematocrit jointly regulate oxygen delivery, shear, and washout in pediatric roller-pump cardiopulmonary bypass. Methods: A three-dimensional, closed-loop fluid-structure interaction model resolved roller-tube contact, tubing deformation, and blood flow. Forty simulations combined 4 roller speeds (38-95 revolutions per minute), 2 occlusion levels (77% and 83%), and 5 hematocrit levels (15%-50%). Cycle-averaged flow, oxygen delivery, peak shear stress, and viscosity-normalized relative residence time were quantified. Results: Roller compression produced retrograde leakage through the occluded gap, localized peak shear stress exceeding 100 pascals, and adjacent stagnation and recirculation. Flow increased nearly linearly with roller speed and was higher at greater occlusion, with little hematocrit dependence. Higher speed and occlusion increased peak shear stress while reducing relative residence time; higher hematocrit further increased peak shear stress. Similar oxygen delivery could arise from substantially different flow environments: among 10 conditions providing oxygen delivery of approximately 9-12 milliliters per minute, peak shear stress ranged from approximately 25 to 70 pascals. Within conditions satisfying an illustrative minimum oxygen-delivery threshold, lower hematocrit permitted speed and occlusion settings yielding lower peak shear stress and relative residence time. Conclusions: Pediatric roller-pump settings that provide similar oxygen delivery can expose blood to markedly different shear and washout environments. Joint evaluation of these metrics may guide selection of speed, occlusion, and hematocrit to meet a target oxygen delivery while reducing shear and improving washout. Keywords: Cardiopulmonary bypass; roller pump; pediatric cardiac surgery; fluid-structure interaction; shear stress; relative residence time

bioRxiv (Cold Spring Harbor Laboratory)
Seattle Children's Hospital (US), Children's Hospital of Orange County (US), University of California, Los Angeles (US), University of Washington (US)
Good health and well-being
Mechanical Circulatory Support Devices
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon