Charge Loss of Proteins in vacuo: Implications for Native Mass Spectrometry and Ion Mobility

Abstract Native ion mobility−mass spectrometry (IM−MS) is a powerful biophysical technique for characterizing protein structure. Because individual charge states often exhibit distinct structural dynamics, accurate IM-MS measurements require charge states to be examined independently. Although anecdotal reports among mass spectrometrists suggest that protein ions can lose charges during native IM-MS experiments, a thorough investigation of this phenomenon has been lacking. Here, we demonstrate that protein ions undergo charge loss inside the mass spectrometer as a direct consequence of gas-phase collisions, potentially leading to altered charge state distributions and ion mobility data. This effect was consistently observed across two instrument platforms (Synapt G2-S and Q Exactive UHMR), both ion polarities (positive and negative mode) and for three proteins (bovine serum albumin, myoglobin, and immunoglobulin G). Systematic investigations of instrumental parameters revealed that collision gas pressure has the strongest impact on the magnitude of charge loss, and that higher charge states are more affected than lower charge states. We further show that this process is largely driven by the dissociation of salt adducts and can hence be minimized through desalting in solution or inside the mass spectrometer, prior to the first m/z-analyzer (usually a quadrupole). For collision-induced unfolding experiments, significant differences were found between fingerprints acquired with and without quadrupole selection, although it remains unclear whether this phenomenon is directly related to charge loss. Overall, our study provides fundamental insights into collision-induced charge loss as a source of error in native IM-MS measurements and discusses the relevance of quadrupole m/z-selection for improving confidence in IM studies.

Authors

Institutions

Publication Details

Journal
Analytical Chemistry
Published
2026-10-28
DOI
https://doi.org/10.1021/acs.analchem.6c04928
Primary Topic
Mass Spectrometry Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Charge Loss of Proteins in vacuo: Implications for Native Mass Spectrometry and Ion Mobility

Kevin Pagel, Jacob S. Jordan, Nicklas Österlund, Niklas Geue et al.
Analytical Chemistry
Mass Spectrometry Techniques and Applications
article

Charge Loss of Proteins in vacuo: Implications for Native Mass Spectrometry and Ion Mobility

Kevin Pagel, Jacob S. Jordan, Nicklas Österlund, Niklas Geue, Mathew Bejoy
article en

Abstract

Abstract Native ion mobility−mass spectrometry (IM−MS) is a powerful biophysical technique for characterizing protein structure. Because individual charge states often exhibit distinct structural dynamics, accurate IM-MS measurements require charge states to be examined independently. Although anecdotal reports among mass spectrometrists suggest that protein ions can lose charges during native IM-MS experiments, a thorough investigation of this phenomenon has been lacking. Here, we demonstrate that protein ions undergo charge loss inside the mass spectrometer as a direct consequence of gas-phase collisions, potentially leading to altered charge state distributions and ion mobility data. This effect was consistently observed across two instrument platforms (Synapt G2-S and Q Exactive UHMR), both ion polarities (positive and negative mode) and for three proteins (bovine serum albumin, myoglobin, and immunoglobulin G). Systematic investigations of instrumental parameters revealed that collision gas pressure has the strongest impact on the magnitude of charge loss, and that higher charge states are more affected than lower charge states. We further show that this process is largely driven by the dissociation of salt adducts and can hence be minimized through desalting in solution or inside the mass spectrometer, prior to the first m/z-analyzer (usually a quadrupole). For collision-induced unfolding experiments, significant differences were found between fingerprints acquired with and without quadrupole selection, although it remains unclear whether this phenomenon is directly related to charge loss. Overall, our study provides fundamental insights into collision-induced charge loss as a source of error in native IM-MS measurements and discusses the relevance of quadrupole m/z-selection for improving confidence in IM studies.

Analytical Chemistry
Fritz Haber Institute of the Max Planck Society (DE), Freie Universität Berlin (DE)
Openalex Percentile: Top 27%
Mass Spectrometry Techniques and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon