Concomitant Aveir AR and Micra AV2 Hybrid Leadless Implantation During Infected Pacemaker Extraction
Leadless pacing is an important option for patients with device-related infections. In those requiring dual-chamber support, reduced atrial battery longevity due to continuous implant-to-implant (i2i) communication remains a concern. While recent software updates now allow for the disablement of i2i communication when clinically appropriate, this mode does not provide atrioventricular synchronous pacing. We describe a case of a hybrid leadless system combining an atrial leadless pacemaker (Aveir AR, Abbott) and a second leadless pacemaker capable of atrial mechanical-sensed ventricular pacing (Micra AV2, Medtronic) in the setting of extraction of an infected transvenous pacemaker, providing atrial pacing with ventricular backup capability while preserving atrial device longevity.
Authors
- Laurence Mark Epstein (ORCID: https://orcid.org/0000-0001-5410-0291)
- Sri Nuvvula (ORCID: https://orcid.org/0009-0007-3154-8238)
- Michael Alston (ORCID: https://orcid.org/0000-0001-7631-0070)
- James Kiernan Gabriels (ORCID: https://orcid.org/0000-0002-1095-6626)
- Catherine Pham
- Julie Thomas
Institutions
- Northwell Health (US)
- North Shore University Hospital (US)
Publication Details
- Journal
- Pacing and Clinical Electrophysiology
- Published
- 2026-10-28
- DOI
- https://doi.org/10.1111/pace.70487
- Primary Topic
- Cardiac pacing and defibrillation studies
- Type
- article
- Field-Weighted Citation Impact
- 0.00