Daily Multi‐Leaf Collimator positional error prediction using machine learning: A Dynalog‐based approach for patient specific quality assurance

Abstract Background Accurate positioning of multi‐leaf collimators (MLCs) is critical for ensuring the dosimetric precision of intensity‐modulated radiation therapy (IMRT). Nonetheless, daily MLC positional variations can degrade treatment accuracy, emphasizing the need for robust methodologies to detect and predict these deviations. Purpose This study proposes a machine learning (ML) approach to predict daily MLC positional deviations using factors contributing to positional errors, enabling daily patient‐specific quality assurance (PSQA) to assess the dosimetric impact of these discrepancies on patient treatment. Methods DICOM‐RT and Dynalog files from 29 IMRT patients were collected, and 78 features were extracted, including aperture‐based features, mechanical parameters, and MLC check data. A LightGBM machine learning model was constructed to predict daily positional errors between planned and delivered MLC positions at each control point. For dosimetric analysis, nine independent IMRT plans were evaluated using dose volume histogram (DVH) parameters such as D 98% , and V 100% for targets and D max , D mean for Organs at Risk (OARs). Moreover, the Gamma Pass Rate (GPR) was calculated for both the delivered and predicted dose distributions, with regarding the primary planned dose distribution as the reference. Results The proposed LightGBM model demonstrated accurate predictions, achieving Mean Absolute Error (MAE) and Root Mean Square Error (RMSE) of 0.035 mm and 0.078 mm for the training dataset, 0.070 mm and 0.164 mm for the validation dataset, and 0.054 mm and 0.144 mm for the test dataset, respectively. Leaf velocity, with a Spearman correlation coefficient of 0.703, was the most influential feature. Evaluation on nine independent clinical cases demonstrated good agreement between delivered and predicted dosimetric deviations from the planned values, with mean differences below 0.200 Gy and approximately 1.5% for target structures, and even smaller discrepancies for OARs. The GPR (3%/3 mm criterion) was about 100% for delivered doses and 99.99% for predicted doses. Conclusions The proposed ML‐based LightGBM model offers an efficient daily PSQA tool, with the potential to enhance the treatment accuracy and support proactive clinical decision‐making in IMRT.

Authors

Institutions

Publication Details

Journal
Journal of Applied Clinical Medical Physics
Published
2026-10-27
DOI
https://doi.org/10.1002/acm2.70882
Primary Topic
Advanced Radiotherapy Techniques
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Daily Multi‐Leaf Collimator positional error prediction using machine learning: A Dynalog‐based approach for patient specific quality assurance

Zahra Siavashpour, Seyed Mahmoud Reza Aghamiri, Reza Ghaderi, Samira Aghaei Akizi et al.
Journal of Applied Clinical Medical Physics
Advanced Radiotherapy Techniques
article

Daily Multi‐Leaf Collimator positional error prediction using machine learning: A Dynalog‐based approach for patient specific quality assurance

Zahra Siavashpour, Seyed Mahmoud Reza Aghamiri, Reza Ghaderi, Samira Aghaei Akizi, Soharb Barimani
article en

Abstract

Abstract Background Accurate positioning of multi‐leaf collimators (MLCs) is critical for ensuring the dosimetric precision of intensity‐modulated radiation therapy (IMRT). Nonetheless, daily MLC positional variations can degrade treatment accuracy, emphasizing the need for robust methodologies to detect and predict these deviations. Purpose This study proposes a machine learning (ML) approach to predict daily MLC positional deviations using factors contributing to positional errors, enabling daily patient‐specific quality assurance (PSQA) to assess the dosimetric impact of these discrepancies on patient treatment. Methods DICOM‐RT and Dynalog files from 29 IMRT patients were collected, and 78 features were extracted, including aperture‐based features, mechanical parameters, and MLC check data. A LightGBM machine learning model was constructed to predict daily positional errors between planned and delivered MLC positions at each control point. For dosimetric analysis, nine independent IMRT plans were evaluated using dose volume histogram (DVH) parameters such as D 98% , and V 100% for targets and D max , D mean for Organs at Risk (OARs). Moreover, the Gamma Pass Rate (GPR) was calculated for both the delivered and predicted dose distributions, with regarding the primary planned dose distribution as the reference. Results The proposed LightGBM model demonstrated accurate predictions, achieving Mean Absolute Error (MAE) and Root Mean Square Error (RMSE) of 0.035 mm and 0.078 mm for the training dataset, 0.070 mm and 0.164 mm for the validation dataset, and 0.054 mm and 0.144 mm for the test dataset, respectively. Leaf velocity, with a Spearman correlation coefficient of 0.703, was the most influential feature. Evaluation on nine independent clinical cases demonstrated good agreement between delivered and predicted dosimetric deviations from the planned values, with mean differences below 0.200 Gy and approximately 1.5% for target structures, and even smaller discrepancies for OARs. The GPR (3%/3 mm criterion) was about 100% for delivered doses and 99.99% for predicted doses. Conclusions The proposed ML‐based LightGBM model offers an efficient daily PSQA tool, with the potential to enhance the treatment accuracy and support proactive clinical decision‐making in IMRT.

Journal of Applied Clinical Medical PhysicsVol. 27(11)
Amirkabir University of Technology (IR), Shahid Beheshti University (IR), Shahid Beheshti University of Medical Sciences (IR), Tehran University of Medical Sciences (IR)
Openalex Percentile: Top 14%
Advanced Radiotherapy Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon