Shape memory and superelastic nitinol wires: A combined mechanical, electrochemical, and ion release study for biomedical use

This work presents a study in detail of commercial shape memory (SM) and superelastic (SE) nitinol wires, including phase transformation analysis, mechanical testing, and electrochemical evaluation in physiological solutions. Phase transformation behavior was investigated by differential scanning calorimetry (DSC) and X-ray diffraction (XRD), confirming a phase transformation B2 → R → B19′ for both alloys. The SE wire exhibited lower transformation temperatures (Ms = −73.01 °C, Af = 53.29 °C) compared to the SM variant (Ms = −45.53 °C, Af = 67.87 °C), with a predominantly B2/austenitic state at room and physiological temperatures under the thermal history relevant to the present experiments enabling stress-induced martensitic transformation. Microstructural characterization by microscopy and XRD revealed acicular microstructures with Ni 4 Ti 3 and Ni 3 Ti 2 precipitates in both materials. Vickers microhardness testing revealed a higher hardness for the SE (429 HV0.2) relative to SM (259 HV0.2). Uniaxial tensile testing at −60 °C, 23 °C, and 100 °C established that SM wires achieve superior ductility in the martensitic state (32% strain at −60 °C), whereas SE wires exhibit complete pseudoelastic recovery with strain-dependent hysteresis. Electrochemical characterization via OCP, potentiodynamic polarization, and EIS in artificial saliva and phosphate-buffered saline demonstrated that the SM alloy presents superior corrosion resistance, characterized by nobler potentials, lower passive current densities, and higher polarization resistance. Ion release quantification by ICP-OES confirmed negligible titanium leaching and limited nickel release (0.90–0.98 µg·cm⁻²·day⁻¹) for SM, with SE below detection limits. These findings suggest that thermomechanical processing influences functional, mechanical, and electrochemical behavior, and that differences in material condition have a stronger effect than the test electrolyte on corrosion resistance. Thus, material selection must balance mechanical functionality with electrochemical stability for biomedical applications.

Authors

Institutions

Publication Details

Journal
Journal of Alloys and Compounds
Published
2026-10-27
DOI
https://doi.org/10.1016/j.jallcom.2026.192021
Primary Topic
Shape Memory Alloy Transformations
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Shape memory and superelastic nitinol wires: A combined mechanical, electrochemical, and ion release study for biomedical use

C. Garcia-Hernandez, B. Merillas, F. Martin-Pedrosa, C. Perez-Gonzalez et al.
Journal of Alloys and Compounds
Shape Memory Alloy Transformations
article

Shape memory and superelastic nitinol wires: A combined mechanical, electrochemical, and ion release study for biomedical use

C. Garcia-Hernandez, B. Merillas, F. Martin-Pedrosa, C. Perez-Gonzalez, C. Garcia-Cabezon
article en

Abstract

This work presents a study in detail of commercial shape memory (SM) and superelastic (SE) nitinol wires, including phase transformation analysis, mechanical testing, and electrochemical evaluation in physiological solutions. Phase transformation behavior was investigated by differential scanning calorimetry (DSC) and X-ray diffraction (XRD), confirming a phase transformation B2 → R → B19′ for both alloys. The SE wire exhibited lower transformation temperatures (Ms = −73.01 °C, Af = 53.29 °C) compared to the SM variant (Ms = −45.53 °C, Af = 67.87 °C), with a predominantly B2/austenitic state at room and physiological temperatures under the thermal history relevant to the present experiments enabling stress-induced martensitic transformation. Microstructural characterization by microscopy and XRD revealed acicular microstructures with Ni 4 Ti 3 and Ni 3 Ti 2 precipitates in both materials. Vickers microhardness testing revealed a higher hardness for the SE (429 HV0.2) relative to SM (259 HV0.2). Uniaxial tensile testing at −60 °C, 23 °C, and 100 °C established that SM wires achieve superior ductility in the martensitic state (32% strain at −60 °C), whereas SE wires exhibit complete pseudoelastic recovery with strain-dependent hysteresis. Electrochemical characterization via OCP, potentiodynamic polarization, and EIS in artificial saliva and phosphate-buffered saline demonstrated that the SM alloy presents superior corrosion resistance, characterized by nobler potentials, lower passive current densities, and higher polarization resistance. Ion release quantification by ICP-OES confirmed negligible titanium leaching and limited nickel release (0.90–0.98 µg·cm⁻²·day⁻¹) for SM, with SE below detection limits. These findings suggest that thermomechanical processing influences functional, mechanical, and electrochemical behavior, and that differences in material condition have a stronger effect than the test electrolyte on corrosion resistance. Thus, material selection must balance mechanical functionality with electrochemical stability for biomedical applications.

Journal of Alloys and CompoundsVol. 1084
Universidad de Valladolid (ES)
Ministerio de Ciencia, Innovación y Universidades, Junta de Castilla y León
Openalex Percentile: Top 29%
Shape Memory Alloy Transformations
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon