Validity Assessment of Wrist-Worn Electrodermal Activity Measurement During Relaxation and Acute Stressor Tasks: Comparing the EmbracePlus and Shimmer3 GSR+

Wearable electrodermal activity (EDA) monitoring enables assessment of stress-related physiological changes in daily life, but new devices require validation. This study evaluated the newly introduced EmbracePlus against the Shimmer3 GSR+, used as a research-grade wrist-based comparator. Healthy volunteers (n = 53; n = 41 valid paired recordings) completed breathing exercises, a Stroop task, and a virtual reality (VR) height stressor while wearing both devices simultaneously on the non-dominant wrist. Agreement was assessed at signal, parameter, and event levels for mean phasic activity, skin conductance level (SCL), summed skin conductance response (SCR) amplitude/min, and SCR count/min. Mean signal-level cross-correlation was r = 0.30. At the parameter level, only 44% of SCL, 24% of SCR amplitude/min, and 5% of SCR count/min differences fell within a priori acceptable boundaries, and regression analyses indicated proportional bias for mean phasic activity, summed SCR amplitude/min and SCR count/min. At the event level, Shimmer3 detected increases across all four EDA metrics during both stress tasks (all p < 0.001). EmbracePlus showed increases across all four metrics during VR (all p < 0.001), whereas during Stroop only mean phasic activity (p = 0.001) and SCL (p < 0.001) increased. EmbracePlus and Shimmer3 are therefore not interchangeable for absolute EDA assessment, and EmbracePlus validity depends on stressor intensity and metric selection.

Authors

Institutions

Publication Details

Journal
Sensors
Published
2026-10-27
DOI
https://doi.org/10.3390/s26216795
Primary Topic
Emotion and Mood Recognition
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Validity Assessment of Wrist-Worn Electrodermal Activity Measurement During Relaxation and Acute Stressor Tasks: Comparing the EmbracePlus and Shimmer3 GSR+

Manon W. H. Peeters, Petra Heck, Gerard Schouten, Bo Dirks et al.
1 links
Sensors
Emotion and Mood Recognition
article

Validity Assessment of Wrist-Worn Electrodermal Activity Measurement During Relaxation and Acute Stressor Tasks: Comparing the EmbracePlus and Shimmer3 GSR+

Manon W. H. Peeters, Petra Heck, Gerard Schouten, Bo Dirks, Sjoerd Kesteloo, Bart van Boxtel
article en

Abstract

Wearable electrodermal activity (EDA) monitoring enables assessment of stress-related physiological changes in daily life, but new devices require validation. This study evaluated the newly introduced EmbracePlus against the Shimmer3 GSR+, used as a research-grade wrist-based comparator. Healthy volunteers (n = 53; n = 41 valid paired recordings) completed breathing exercises, a Stroop task, and a virtual reality (VR) height stressor while wearing both devices simultaneously on the non-dominant wrist. Agreement was assessed at signal, parameter, and event levels for mean phasic activity, skin conductance level (SCL), summed skin conductance response (SCR) amplitude/min, and SCR count/min. Mean signal-level cross-correlation was r = 0.30. At the parameter level, only 44% of SCL, 24% of SCR amplitude/min, and 5% of SCR count/min differences fell within a priori acceptable boundaries, and regression analyses indicated proportional bias for mean phasic activity, summed SCR amplitude/min and SCR count/min. At the event level, Shimmer3 detected increases across all four EDA metrics during both stress tasks (all p < 0.001). EmbracePlus showed increases across all four metrics during VR (all p < 0.001), whereas during Stroop only mean phasic activity (p = 0.001) and SCL (p < 0.001) increased. EmbracePlus and Shimmer3 are therefore not interchangeable for absolute EDA assessment, and EmbracePlus validity depends on stressor intensity and metric selection.

SensorsVol. 26(21)
Fontys University of Applied Sciences (NL), Eindhoven University of Technology (NL)
Nationaal Regieorgaan Praktijkgericht Onderzoek SIA
Good health and well-being
Openalex Percentile: Top 8%
Emotion and Mood Recognition
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon