The Need for Device-Specific Training Before Handheld Ultrasound Use: Agreement and Repeatability of Simulated Aortic Aneurysm Diameter Measurements by Senior Sonography Students on a Familiar Conventional and an Unfamiliar Handheld System
Background: Ultrasound measurement of abdominal aortic aneurysm (AAA) diameter guides screening and surveillance, but whether students’ measurement competence on a conventional system transfers to an unfamiliar handheld device has rarely been assessed. Methods: Twenty-seven senior sonography students made three anteroposterior outer-to-outer measurements of two water-filled balloons embedded in gelatin (reference diameters 3.53 and 5.88 cm, calculated from the circumference before embedding), first on the familiar cart-based conventional system and then, in a fixed, non-randomised order, on a handheld system with which they had no prior hands-on experience. The primary outcomes were bias against the calculated reference, with equivalence testing (±5%), and the paired between-system difference, with Bland–Altman analysis; repeatability was assessed with repeatability coefficients. Results: Conventional measurements were equivalent to the calculated reference within ±5% (bias −0.09 cm, 95% CI −0.16 to −0.02; +0.07 cm, 95% CI −0.05 to +0.19). Handheld measurements were higher (+0.17 and +0.33 cm; both p < 0.001) and failed equivalence; the mean between-system difference was +0.26 cm (95% CI 0.18–0.34), without proportional bias. Single-reading repeatability coefficients were 0.31–0.67 cm. Conclusions: Students’ conventional-system measurements agreed closely with the calculated reference, although this reference was not verified after embedding. Because device type was confounded with operator inexperience and with a fixed testing order, which introduces an unquantifiable learning or fatigue effect, the 0.26 cm value is an observed between-system difference under the study conditions and cannot be attributed to the handheld device itself. Measurement competence did not transfer automatically to a new device, underlining the critical need for device-specific training before handheld systems are used for aortic measurement. These implications are educational and technical; phantom findings do not support clinical surveillance recommendations without confirmation in patients.
Authors
- Mahasin G. Hassan (ORCID: https://orcid.org/0000-0002-9965-8915)
- Mohammed J. Alsaadi (ORCID: https://orcid.org/0000-0003-0560-0403)
- Alanoud Alotaibi (ORCID: https://orcid.org/0009-0002-5274-7731)
- Dema Alenazi
- Salsabil Aldosari
- Lamis Al-Sultan
- Rema Alrajhi (ORCID: https://orcid.org/0009-0007-9267-6305)
Institutions
- Princess Nourah bint Abdulrahman University (SA)
- Prince Sattam Bin Abdulaziz University (SA)
Publication Details
- Journal
- Diagnostics
- Published
- 2026-10-27
- DOI
- https://doi.org/10.3390/diagnostics16213524
- Primary Topic
- Ultrasound in Clinical Applications
- Type
- article
- Field-Weighted Citation Impact
- 0.00
Funders
- Princess Nourah Bint Abdulrahman University