The Need for Device-Specific Training Before Handheld Ultrasound Use: Agreement and Repeatability of Simulated Aortic Aneurysm Diameter Measurements by Senior Sonography Students on a Familiar Conventional and an Unfamiliar Handheld System

Background: Ultrasound measurement of abdominal aortic aneurysm (AAA) diameter guides screening and surveillance, but whether students’ measurement competence on a conventional system transfers to an unfamiliar handheld device has rarely been assessed. Methods: Twenty-seven senior sonography students made three anteroposterior outer-to-outer measurements of two water-filled balloons embedded in gelatin (reference diameters 3.53 and 5.88 cm, calculated from the circumference before embedding), first on the familiar cart-based conventional system and then, in a fixed, non-randomised order, on a handheld system with which they had no prior hands-on experience. The primary outcomes were bias against the calculated reference, with equivalence testing (±5%), and the paired between-system difference, with Bland–Altman analysis; repeatability was assessed with repeatability coefficients. Results: Conventional measurements were equivalent to the calculated reference within ±5% (bias −0.09 cm, 95% CI −0.16 to −0.02; +0.07 cm, 95% CI −0.05 to +0.19). Handheld measurements were higher (+0.17 and +0.33 cm; both p < 0.001) and failed equivalence; the mean between-system difference was +0.26 cm (95% CI 0.18–0.34), without proportional bias. Single-reading repeatability coefficients were 0.31–0.67 cm. Conclusions: Students’ conventional-system measurements agreed closely with the calculated reference, although this reference was not verified after embedding. Because device type was confounded with operator inexperience and with a fixed testing order, which introduces an unquantifiable learning or fatigue effect, the 0.26 cm value is an observed between-system difference under the study conditions and cannot be attributed to the handheld device itself. Measurement competence did not transfer automatically to a new device, underlining the critical need for device-specific training before handheld systems are used for aortic measurement. These implications are educational and technical; phantom findings do not support clinical surveillance recommendations without confirmation in patients.

Authors

Institutions

Publication Details

Journal
Diagnostics
Published
2026-10-27
DOI
https://doi.org/10.3390/diagnostics16213524
Primary Topic
Ultrasound in Clinical Applications
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

The Need for Device-Specific Training Before Handheld Ultrasound Use: Agreement and Repeatability of Simulated Aortic Aneurysm Diameter Measurements by Senior Sonography Students on a Familiar Conventional and an Unfamiliar Handheld System

Mahasin G. Hassan, Mohammed J. Alsaadi, Alanoud Alotaibi, Dema Alenazi et al.
1 links
Diagnostics
Ultrasound in Clinical Applications
article

The Need for Device-Specific Training Before Handheld Ultrasound Use: Agreement and Repeatability of Simulated Aortic Aneurysm Diameter Measurements by Senior Sonography Students on a Familiar Conventional and an Unfamiliar Handheld System

Mahasin G. Hassan, Mohammed J. Alsaadi, Alanoud Alotaibi, Dema Alenazi, Salsabil Aldosari, Lamis Al-Sultan, Rema Alrajhi
article en

Abstract

Background: Ultrasound measurement of abdominal aortic aneurysm (AAA) diameter guides screening and surveillance, but whether students’ measurement competence on a conventional system transfers to an unfamiliar handheld device has rarely been assessed. Methods: Twenty-seven senior sonography students made three anteroposterior outer-to-outer measurements of two water-filled balloons embedded in gelatin (reference diameters 3.53 and 5.88 cm, calculated from the circumference before embedding), first on the familiar cart-based conventional system and then, in a fixed, non-randomised order, on a handheld system with which they had no prior hands-on experience. The primary outcomes were bias against the calculated reference, with equivalence testing (±5%), and the paired between-system difference, with Bland–Altman analysis; repeatability was assessed with repeatability coefficients. Results: Conventional measurements were equivalent to the calculated reference within ±5% (bias −0.09 cm, 95% CI −0.16 to −0.02; +0.07 cm, 95% CI −0.05 to +0.19). Handheld measurements were higher (+0.17 and +0.33 cm; both p < 0.001) and failed equivalence; the mean between-system difference was +0.26 cm (95% CI 0.18–0.34), without proportional bias. Single-reading repeatability coefficients were 0.31–0.67 cm. Conclusions: Students’ conventional-system measurements agreed closely with the calculated reference, although this reference was not verified after embedding. Because device type was confounded with operator inexperience and with a fixed testing order, which introduces an unquantifiable learning or fatigue effect, the 0.26 cm value is an observed between-system difference under the study conditions and cannot be attributed to the handheld device itself. Measurement competence did not transfer automatically to a new device, underlining the critical need for device-specific training before handheld systems are used for aortic measurement. These implications are educational and technical; phantom findings do not support clinical surveillance recommendations without confirmation in patients.

DiagnosticsVol. 16(21)
Princess Nourah bint Abdulrahman University (SA), Prince Sattam Bin Abdulaziz University (SA)
Princess Nourah Bint Abdulrahman University
Good health and well-being, Quality education
Openalex Percentile: Top 11%
Ultrasound in Clinical Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon