Optical Calibration of a Frequency-Domain-Multiplexed X-Ray TES Microcalorimeter Array via Photon-Number-Resolved Statistics

Transition-edge sensor (TES) microcalorimeters for future X-ray space observatories require accurate low-energy calibration, a regime where conventional X-ray sources are limited by intrinsic linewidth, count rate, or experimental complexity. In this work, we investigate fiber-coupled optical photons as a compact calibration tool for an Athena-like TES array. Optical pulses at 405 nm were generated with an acousto-optic modulator and delivered through an optical fiber onto a selected TES pixel inside a dilution refrigerator, via a dedicated cryogenic alignment interface. Externally triggered pulses were processed with an optimal-filter pipeline, including baseline subtraction, drift correction, and multi-Gaussian reconstruction of the pulse-height spectrum. The reconstructed spectra show resolved photon-number peaks consistent with the Poissonian statistics expected from an attenuated coherent source, reaching equivalent energies above 400 eV with a mean photon number exceeding 100 per pulse. The detector response was further tested for linearity by comparing the measured photon-number distributions against those predicted from the independently calibrated number of photons sent to the detector and a single, globally estimated detection efficiency, confirming Poissonian, linear behaviour across nearly two orders of magnitude in photon number. These results demonstrate that fiber-delivered optical photons can probe the few-eV to sub-keV response of Athena-like TES microcalorimeters, offering a practical and extensible framework for future low-energy calibration studies.

Authors

Institutions

Publication Details

Journal
Sensors
Published
2026-10-27
DOI
https://doi.org/10.3390/s26216788
Primary Topic
Superconducting and THz Device Technology
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Optical Calibration of a Frequency-Domain-Multiplexed X-Ray TES Microcalorimeter Array via Photon-Number-Resolved Statistics

W. Löffler, Luciano G. Gottardi, K. Ravensberg, K. Nagayoshi et al.
1 links
Sensors
Superconducting and THz Device Technology
article

Optical Calibration of a Frequency-Domain-Multiplexed X-Ray TES Microcalorimeter Array via Photon-Number-Resolved Statistics

W. Löffler, Luciano G. Gottardi, K. Ravensberg, K. Nagayoshi, Mio Poortvliet, E. Taralli, M. de Wit, Davide Vaccaro, Giacomo Gorla, Hugo van Bohemen, Bernard Sohl
article en

Abstract

Transition-edge sensor (TES) microcalorimeters for future X-ray space observatories require accurate low-energy calibration, a regime where conventional X-ray sources are limited by intrinsic linewidth, count rate, or experimental complexity. In this work, we investigate fiber-coupled optical photons as a compact calibration tool for an Athena-like TES array. Optical pulses at 405 nm were generated with an acousto-optic modulator and delivered through an optical fiber onto a selected TES pixel inside a dilution refrigerator, via a dedicated cryogenic alignment interface. Externally triggered pulses were processed with an optimal-filter pipeline, including baseline subtraction, drift correction, and multi-Gaussian reconstruction of the pulse-height spectrum. The reconstructed spectra show resolved photon-number peaks consistent with the Poissonian statistics expected from an attenuated coherent source, reaching equivalent energies above 400 eV with a mean photon number exceeding 100 per pulse. The detector response was further tested for linearity by comparing the measured photon-number distributions against those predicted from the independently calibrated number of photons sent to the detector and a single, globally estimated detection efficiency, confirming Poissonian, linear behaviour across nearly two orders of magnitude in photon number. These results demonstrate that fiber-delivered optical photons can probe the few-eV to sub-keV response of Athena-like TES microcalorimeters, offering a practical and extensible framework for future low-energy calibration studies.

SensorsVol. 26(21)
Leiden University (NL), Huygens Institute for History and Culture of the Netherlands (NL), NWO-I (NL), Space Research Organisation Netherlands (NL), Sapienza University of Rome (IT), National Institute for Astrophysics (IT)
Openalex Percentile: Top 14%
Superconducting and THz Device Technology
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon