His Bundle Pacing via the Atrial Port of a Dual‐Chamber ICD: Programming Solution for Undersensing
BACKGROUND: His-bundle pacing connected to the atrial port of a dual-chamber ICD may lead to sensing failure and unsafe pacing. CASE SUMMARY: A 67-year-old man with permanent atrial fibrillation, alternating rapid and slow ventricular rates despite medication, symptomatic pauses longer than 6s, and a dual chamber ICD (His lead in the RA port; DF4 lead in the RV port) presented with palpitations. ECG and 24-hour Holter demonstrated intermittent His-channel undersensing with pacing on the T-wave peak. Interrogation showed small His electrograms followed by a prominent, constant ventricular far-field signal on the atrial channel. We turned off far-field protection after V-sense, intentionally used the ventricular far-field signal for sensing of His lead, and disabled related discrimination algorithms. CONCLUSION: A dual-chamber ICD with a His-bundle pacing lead may be associated with complex programming challenges, and targeted reprogramming can avoid lead revision or reimplantation.
Authors
- Ngoc Thai Kieu (ORCID: https://orcid.org/0009-0005-1464-328X)
- Le Uyen Phuong Tran (ORCID: https://orcid.org/0000-0002-6313-8379)
- Thị Vân Khánh Nguyễn (ORCID: https://orcid.org/0009-0003-8747-2835)
- Diane T. Vo (ORCID: https://orcid.org/0009-0006-8810-4614)
Institutions
- Can Tho University (VN)
- Cho Ray Hospital (VN)
- Trường ĐH Nguyễn Tất Thành (VN)
- Can Tho University of Medicine and Pharmacy (VN)
Publication Details
- Journal
- Pacing and Clinical Electrophysiology
- Published
- 2026-10-27
- DOI
- https://doi.org/10.1111/pace.70481
- Primary Topic
- Cardiac pacing and defibrillation studies
- Type
- article
- Field-Weighted Citation Impact
- 0.00