His Bundle Pacing via the Atrial Port of a Dual‐Chamber ICD: Programming Solution for Undersensing

BACKGROUND: His-bundle pacing connected to the atrial port of a dual-chamber ICD may lead to sensing failure and unsafe pacing. CASE SUMMARY: A 67-year-old man with permanent atrial fibrillation, alternating rapid and slow ventricular rates despite medication, symptomatic pauses longer than 6s, and a dual chamber ICD (His lead in the RA port; DF4 lead in the RV port) presented with palpitations. ECG and 24-hour Holter demonstrated intermittent His-channel undersensing with pacing on the T-wave peak. Interrogation showed small His electrograms followed by a prominent, constant ventricular far-field signal on the atrial channel. We turned off far-field protection after V-sense, intentionally used the ventricular far-field signal for sensing of His lead, and disabled related discrimination algorithms. CONCLUSION: A dual-chamber ICD with a His-bundle pacing lead may be associated with complex programming challenges, and targeted reprogramming can avoid lead revision or reimplantation.

Authors

Institutions

Publication Details

Journal
Pacing and Clinical Electrophysiology
Published
2026-10-27
DOI
https://doi.org/10.1111/pace.70481
Primary Topic
Cardiac pacing and defibrillation studies
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

His Bundle Pacing via the Atrial Port of a Dual‐Chamber ICD: Programming Solution for Undersensing

Ngoc Thai Kieu, Le Uyen Phuong Tran, Thị Vân Khánh Nguyễn, Diane T. Vo
Pacing and Clinical Electrophysiology
Cardiac pacing and defibrillation studies
article

His Bundle Pacing via the Atrial Port of a Dual‐Chamber ICD: Programming Solution for Undersensing

Ngoc Thai Kieu, Le Uyen Phuong Tran, Thị Vân Khánh Nguyễn, Diane T. Vo
article en

Abstract

BACKGROUND: His-bundle pacing connected to the atrial port of a dual-chamber ICD may lead to sensing failure and unsafe pacing. CASE SUMMARY: A 67-year-old man with permanent atrial fibrillation, alternating rapid and slow ventricular rates despite medication, symptomatic pauses longer than 6s, and a dual chamber ICD (His lead in the RA port; DF4 lead in the RV port) presented with palpitations. ECG and 24-hour Holter demonstrated intermittent His-channel undersensing with pacing on the T-wave peak. Interrogation showed small His electrograms followed by a prominent, constant ventricular far-field signal on the atrial channel. We turned off far-field protection after V-sense, intentionally used the ventricular far-field signal for sensing of His lead, and disabled related discrimination algorithms. CONCLUSION: A dual-chamber ICD with a His-bundle pacing lead may be associated with complex programming challenges, and targeted reprogramming can avoid lead revision or reimplantation.

Pacing and Clinical Electrophysiology
Can Tho University (VN), Cho Ray Hospital (VN), Trường ĐH Nguyễn Tất Thành (VN), Can Tho University of Medicine and Pharmacy (VN)
Good health and well-being
Openalex Percentile: Top 11%
Cardiac pacing and defibrillation studies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon