Construct Validity and Test–Retest Reliability of the Al Balance Stabilometric System in Healthy Young Adults

Each instrumented balance platform requires its own measurement-property evidence. We examined the test–retest reliability of the Al Balance Stabilometric System and its hypothesis-based construct validity relative to the Biodex Balance System SD. Forty healthy adults aged 18–25 years completed four Al Balance modules: four-condition sensory assessment (mCTSIB), Limits of Stability (LoS), unilateral stance (Bilateral Comparison Test), and Postural Stability Test. The matched sensory conditions, unilateral stance, and LoS were also performed on Biodex in randomized device order; Al Balance was repeated after 7 days. Validity was assessed with repeated-measures ANOVAs, repeated-measures correlations (rrm), within-participant profiles and standardized interaction models; reliability was assessed with intraclass correlation coefficients (ICC), standard error of measurement and minimal detectable change. Both systems captured the expected visual and surface effects, and cross-device convergence, rather than numerical agreement, was supported for the sensory conditions (rrm = 0.513–0.598). However, individual-level agreement was weak: between-device correlations of participant-level sensory response magnitudes were non-significant after Holm correction, and individual asymmetry scores in unilateral stance converged weakly. Device × condition interactions indicated differing relative sensitivity. Test–retest reliability was outcome-dependent, moderate to excellent (ICC(A,1) = 0.678–0.946). In healthy young adults, Al Balance provided repeatable scores that are device-specific rather than interchangeable with Biodex.

Authors

Institutions

Publication Details

Journal
Sensors
Published
2026-10-26
DOI
https://doi.org/10.3390/s26216757
Primary Topic
Balance, Gait, and Falls Prevention
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Construct Validity and Test–Retest Reliability of the Al Balance Stabilometric System in Healthy Young Adults

Ayşe Asena YEKDANEH, Elcin Akyurek, Ayşe Zengin Alpözgen, Bülent Bayraktar et al.
Sensors
Balance, Gait, and Falls Prevention
article

Construct Validity and Test–Retest Reliability of the Al Balance Stabilometric System in Healthy Young Adults

Ayşe Asena YEKDANEH, Elcin Akyurek, Ayşe Zengin Alpözgen, Bülent Bayraktar, Nilay Arman, İrem Kurt, Türker Şahinkaya
article en

Abstract

Each instrumented balance platform requires its own measurement-property evidence. We examined the test–retest reliability of the Al Balance Stabilometric System and its hypothesis-based construct validity relative to the Biodex Balance System SD. Forty healthy adults aged 18–25 years completed four Al Balance modules: four-condition sensory assessment (mCTSIB), Limits of Stability (LoS), unilateral stance (Bilateral Comparison Test), and Postural Stability Test. The matched sensory conditions, unilateral stance, and LoS were also performed on Biodex in randomized device order; Al Balance was repeated after 7 days. Validity was assessed with repeated-measures ANOVAs, repeated-measures correlations (rrm), within-participant profiles and standardized interaction models; reliability was assessed with intraclass correlation coefficients (ICC), standard error of measurement and minimal detectable change. Both systems captured the expected visual and surface effects, and cross-device convergence, rather than numerical agreement, was supported for the sensory conditions (rrm = 0.513–0.598). However, individual-level agreement was weak: between-device correlations of participant-level sensory response magnitudes were non-significant after Holm correction, and individual asymmetry scores in unilateral stance converged weakly. Device × condition interactions indicated differing relative sensitivity. Test–retest reliability was outcome-dependent, moderate to excellent (ICC(A,1) = 0.678–0.946). In healthy young adults, Al Balance provided repeatable scores that are device-specific rather than interchangeable with Biodex.

SensorsVol. 26(21)
Fenerbahçe University (TR), Istanbul University-Cerrahpaşa (TR), Istanbul University (TR)
Openalex Percentile: Top 5%
Balance, Gait, and Falls Prevention
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon