Effect of Scan Body Geometry and CAD Software on the Accuracy of Implant-Position Transfer: An In Vitro Study

Accurate transfer of implant position to the virtual model is essential for digital implant prosthodontic workflows. Although scan body geometry and CAD software may affect this process, evidence regarding their combined effect remains limited. This in vitro study evaluated the effects of two scan body systems and three CAD software programs on implant-position transfer accuracy and assessed their interaction. An edentulous maxillary titanium model with three implants was digitized ten times for each of two scan body systems. Each data set was processed with exocad, 3shape, and DWOS. Implant-position transfer was assessed as the three-dimensional Euclidean deviation of standardized scan body reference points. Data were analyzed using a two-way mixed repeated-measures ANOVA and Holm-adjusted pairwise comparisons (α = 0.05). Scan body system, CAD software, and their interaction significantly affected transfer accuracy (p < 0.001). Medentika showed lower mean deviations than NT-Trading with all software programs, although the difference was not significant with 3shape. For NT-Trading, 3shape produced significantly lower deviations than exocad and DWOS, whereas for Medentika, exocad produced significantly lower deviations than DWOS, with no significant differences between 3shape and either exocad or DWOS. Implant-position transfer accuracy depended on the specific combination of scan body system and CAD software. These findings suggest that scan body system and CAD software should be considered jointly when evaluating implant-position transfer accuracy.

Authors

Institutions

Publication Details

Journal
Bioengineering
Published
2026-10-26
DOI
https://doi.org/10.3390/bioengineering13111266
Primary Topic
Dental Implant Techniques and Outcomes
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Effect of Scan Body Geometry and CAD Software on the Accuracy of Implant-Position Transfer: An In Vitro Study

Ragai Edward Matta, Manfred Gerhard Wichmann, Felix Förtsch, Fynn-Ole Petersen et al.
Bioengineering
Dental Implant Techniques and Outcomes
article

Effect of Scan Body Geometry and CAD Software on the Accuracy of Implant-Position Transfer: An In Vitro Study

Ragai Edward Matta, Manfred Gerhard Wichmann, Felix Förtsch, Fynn-Ole Petersen, Max-Michael Blechschmidt
article en

Abstract

Accurate transfer of implant position to the virtual model is essential for digital implant prosthodontic workflows. Although scan body geometry and CAD software may affect this process, evidence regarding their combined effect remains limited. This in vitro study evaluated the effects of two scan body systems and three CAD software programs on implant-position transfer accuracy and assessed their interaction. An edentulous maxillary titanium model with three implants was digitized ten times for each of two scan body systems. Each data set was processed with exocad, 3shape, and DWOS. Implant-position transfer was assessed as the three-dimensional Euclidean deviation of standardized scan body reference points. Data were analyzed using a two-way mixed repeated-measures ANOVA and Holm-adjusted pairwise comparisons (α = 0.05). Scan body system, CAD software, and their interaction significantly affected transfer accuracy (p < 0.001). Medentika showed lower mean deviations than NT-Trading with all software programs, although the difference was not significant with 3shape. For NT-Trading, 3shape produced significantly lower deviations than exocad and DWOS, whereas for Medentika, exocad produced significantly lower deviations than DWOS, with no significant differences between 3shape and either exocad or DWOS. Implant-position transfer accuracy depended on the specific combination of scan body system and CAD software. These findings suggest that scan body system and CAD software should be considered jointly when evaluating implant-position transfer accuracy.

BioengineeringVol. 13(11)
Friedrich-Alexander-Universität Erlangen-Nürnberg (DE), Universitätsklinikum Erlangen (DE)
Openalex Percentile: Top 10%
Dental Implant Techniques and Outcomes
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon