VERIFICATION OF THE ANALYTICAL PERFORMANCE OF THE IGE ASSAY ON ABBOTT ARCHITECT CI8200

Measurements of IgE are useful to carry out on a routine basis for diagnosis, postoperative monitoring and assessment of treatment of the atopy-associated diseases such as atopic asthma, rhinitis and dermatitis etc., as well as of the infectious diseases, caused by parasites and of the IgE-myeloma. The objective of this study was to independently evaluate the analytical performance of the IgE assay on the Abbott Architect ci8200 automated system and to ensure the best reliability of the results according to the recommendations of the NF ISO 15,189 standard. Our study was performed according to the Scope A criteria detailed in the Medical Biology Method Verification/Validation Guide. We evaluated the repeatability and the intermediate precision of the IgE assay. The coefficients of variation obtained by our study were compared with those set by SFBC. The results obtained showed a satisfactory repeatability for the 2 levels; low and high with respectively CV1 = 3.85 % and CV2 = 1.56 %, a satisfactory intra-laboratory reproducibility for the 2 levels; low and high with CV1 = 7.48 % and CV2 = 5.19 % % respectively. The results obtained allowed us to verify the performance of the method and compare them to the analytical objectives set in order to meet regulatory and normative requirements. The central laboratory of the University Hospital Mohammed VI of Oujda is committed to a quality policy that includes an approach of mastery of the different analytical systems used.

Authors

Institutions

Publication Details

Journal
World Journal of Biology Pharmacy and Health Sciences
Published
2026-10-26
DOI
https://doi.org/10.30574/wjbphs.2026.28.1.0518
Primary Topic
Clinical Laboratory Practices and Quality Control
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

VERIFICATION OF THE ANALYTICAL PERFORMANCE OF THE IGE ASSAY ON ABBOTT ARCHITECT CI8200

Imane Elmezgueldi, El-houcine Sebbar, Imane Naji, Marouane Kaabouchi et al.
World Journal of Biology Pharmacy and Health Sciences
Clinical Laboratory Practices and Quality Control
article

VERIFICATION OF THE ANALYTICAL PERFORMANCE OF THE IGE ASSAY ON ABBOTT ARCHITECT CI8200

Imane Elmezgueldi, El-houcine Sebbar, Imane Naji, Marouane Kaabouchi, Mohammed Choukri, Oussama REHHAB
article en

Abstract

Measurements of IgE are useful to carry out on a routine basis for diagnosis, postoperative monitoring and assessment of treatment of the atopy-associated diseases such as atopic asthma, rhinitis and dermatitis etc., as well as of the infectious diseases, caused by parasites and of the IgE-myeloma. The objective of this study was to independently evaluate the analytical performance of the IgE assay on the Abbott Architect ci8200 automated system and to ensure the best reliability of the results according to the recommendations of the NF ISO 15,189 standard. Our study was performed according to the Scope A criteria detailed in the Medical Biology Method Verification/Validation Guide. We evaluated the repeatability and the intermediate precision of the IgE assay. The coefficients of variation obtained by our study were compared with those set by SFBC. The results obtained showed a satisfactory repeatability for the 2 levels; low and high with respectively CV1 = 3.85 % and CV2 = 1.56 %, a satisfactory intra-laboratory reproducibility for the 2 levels; low and high with CV1 = 7.48 % and CV2 = 5.19 % % respectively. The results obtained allowed us to verify the performance of the method and compare them to the analytical objectives set in order to meet regulatory and normative requirements. The central laboratory of the University Hospital Mohammed VI of Oujda is committed to a quality policy that includes an approach of mastery of the different analytical systems used.

World Journal of Biology Pharmacy and Health SciencesVol. 28(1)
Mohamed I University (MA), Centre Hospitalo-Universitaire Mohammed VI d'Oujda (MA)
Openalex Percentile: Top 12%
Clinical Laboratory Practices and Quality Control
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon