In vitro targeted microbubble adhesion exhibits a maximum at intermediate acoustic pressures

Introduction Ultrasound molecular imaging (USMI) relies on the adhesion of targeted microbubbles (MBs) to endothelial biomarkers. Acoustic radiation force ( F rad ) enhances this process by translating circulating MBs toward the vessel wall. Size- isolated MB populations can be designed so that their resonance frequency matches the transmitted ultrasound frequency, thereby increasing F rad -induced displacement under physiological flow. However, increasing peak negative pressure (PNP) may also destabilize MBs and reduce their ability to establish firm adhesion.We therefore investigated how PNP and pulse duration influence the balance among MB oscillation, translation, stability and targeted adhesion. Methods A modified Rayleigh–Plesset equation with the Marmottant shell model was coupled to the Cattaneo–Supponen scaling relationships to predict radial oscillation, displacement and dissolution of monodisperse 5-µm diameter lipid- coated MBs exposed to 4 MHz pulses. Model predictions were evaluated using in vitro flow channel experiments with αvβ3-targeted MBs at PNPs of 10, 50 and 133 kPa and pulse lengths of 27 and 54 cycles. Results The model predicted increasing radial expansion and displacement with increasing PNP, but the sequential model predicted that repeated exposure at 133 kPa could reduce MB size and thereby decrease cumulative transport. Experimentally, PNP significantly affected targeted adhesion, with the highest mean adhesion observed at 50 kPa for both pulse lengths. Discussion These results demonstrate that maximizing F rad -induced displacement does not necessarily maximize targeted adhesion. Instead, USMI protocols should balance MB transport, stability and wall interactions, with an intermediate PNP providing the most favorable F rad -induced targeting conditions.

Authors

Institutions

Publication Details

Journal
Frontiers in Acoustics
Published
2026-10-26
DOI
https://doi.org/10.3389/facou.2026.1988001
Primary Topic
Ultrasound and Hyperthermia Applications
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

In vitro targeted microbubble adhesion exhibits a maximum at intermediate acoustic pressures

J. Ángel Navarro-Becerra, Mark Andrew Borden, Jair I. Castillo
Frontiers in Acoustics
Ultrasound and Hyperthermia Applications
article

In vitro targeted microbubble adhesion exhibits a maximum at intermediate acoustic pressures

J. Ángel Navarro-Becerra, Mark Andrew Borden, Jair I. Castillo
article en

Abstract

Introduction Ultrasound molecular imaging (USMI) relies on the adhesion of targeted microbubbles (MBs) to endothelial biomarkers. Acoustic radiation force ( F rad ) enhances this process by translating circulating MBs toward the vessel wall. Size- isolated MB populations can be designed so that their resonance frequency matches the transmitted ultrasound frequency, thereby increasing F rad -induced displacement under physiological flow. However, increasing peak negative pressure (PNP) may also destabilize MBs and reduce their ability to establish firm adhesion.We therefore investigated how PNP and pulse duration influence the balance among MB oscillation, translation, stability and targeted adhesion. Methods A modified Rayleigh–Plesset equation with the Marmottant shell model was coupled to the Cattaneo–Supponen scaling relationships to predict radial oscillation, displacement and dissolution of monodisperse 5-µm diameter lipid- coated MBs exposed to 4 MHz pulses. Model predictions were evaluated using in vitro flow channel experiments with αvβ3-targeted MBs at PNPs of 10, 50 and 133 kPa and pulse lengths of 27 and 54 cycles. Results The model predicted increasing radial expansion and displacement with increasing PNP, but the sequential model predicted that repeated exposure at 133 kPa could reduce MB size and thereby decrease cumulative transport. Experimentally, PNP significantly affected targeted adhesion, with the highest mean adhesion observed at 50 kPa for both pulse lengths. Discussion These results demonstrate that maximizing F rad -induced displacement does not necessarily maximize targeted adhesion. Instead, USMI protocols should balance MB transport, stability and wall interactions, with an intermediate PNP providing the most favorable F rad -induced targeting conditions.

Frontiers in AcousticsVol. 4
University of Colorado Boulder (US), Centro de Investigación y de Estudios Avanzados del Instituto Politécnico Nacional (MX)
National Institutes of Health
Openalex Percentile: Top 24%
Ultrasound and Hyperthermia Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon