Model‐based image restoration targeting higher‐dose digital mammography: Impact on radiologists' detection of clustered microcalcifications

Abstract Background Image noise in digital mammography (DM) can obscure image details and reduce the detectability of clinically relevant findings such as microcalcifications. Purpose To present and evaluate an extended model‐based (MB) image restoration pipeline for DM that incorporates quantum, electronic, and structural noise components, as well as spatial noise correlation, and to assess its effect on breast radiologists' detection of simulated microcalcification clusters in clinical mammograms. Methods A total of 120 clinical DM images acquired using a GE Senographe Pristina system were used. Simulated microcalcification clusters were inserted into locally dense fibroglandular regions at controlled contrast levels. The restoration pipeline was configured to target the signal and noise characteristics predicted by the model for an acquisition at approximately twice the original radiation dose. Six breast radiologists with different levels of experience, classified as Junior or Senior, participated in two human observer experiments: (i) a pilot study using original clinical images with simulated clusters and a four‐alternative forced‐choice (4AFC) protocol combined with a three‐down‐one‐up (3D1U) staircase to estimate reader‐specific contrast detection thresholds and achieve comparable baseline task difficulty across readers; and (ii) a main 4AFC observer study using individualized datasets generated according to each reader's estimated contrast threshold, in which microcalcification detection performance was compared between the original and restored image conditions based on the percentage of correct responses (PC). Two generalized linear mixed models (GLMMs) were used: the first evaluated the effect of image condition as a fixed effect, with reader and case as crossed random intercepts, and the second evaluated the effect of reader experience on detection performance in the original image condition. Results At the reader‐specific contrast thresholds estimated in the pilot study, the mean PC for the original images was 78.3%, within the prespecified target performance range. Mean PC increased to 84.3% for the restored images, corresponding to an absolute improvement of 6.0 percentage points. The primary GLMM indicated a significant restoration effect on detection performance (OR = 1.57, 95% CI: 1.15–2.14, ). Baseline mean PC was 78.7% for Junior readers and 78.0% for Senior readers, and the separate GLMM found no evidence of an experience‐group effect under the original image condition (). Conclusions The proposed extended MB restoration pipeline improved the detectability of simulated microcalcification clusters by breast radiologists under the controlled 4AFC conditions investigated. By targeting the signal and noise characteristics predicted by the model for an acquisition at approximately twice the original radiation dose without requiring additional radiation exposure, the method represents a promising approach for improving microcalcification detectability in DM.

Authors

Institutions

Publication Details

Journal
Medical Physics
Published
2026-10-26
DOI
https://doi.org/10.1002/mp.70739
Primary Topic
Digital Radiography and Breast Imaging
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Model‐based image restoration targeting higher‐dose digital mammography: Impact on radiologists' detection of clustered microcalcifications

Silvia Maria Prioli de Souza Sabino, Lucas Rodrigues Azevedo Pinheiro Borges, Predrag R. Bakić, Alessandro Foi et al.
Medical Physics
Digital Radiography and Breast Imaging
article

Model‐based image restoration targeting higher‐dose digital mammography: Impact on radiologists' detection of clustered microcalcifications

Silvia Maria Prioli de Souza Sabino, Lucas Rodrigues Azevedo Pinheiro Borges, Predrag R. Bakić, Alessandro Foi, Renann F. Brandão, Renato França Caron, Marcelo A. C. Vieira, Anders Tingberg
article en

Abstract

Abstract Background Image noise in digital mammography (DM) can obscure image details and reduce the detectability of clinically relevant findings such as microcalcifications. Purpose To present and evaluate an extended model‐based (MB) image restoration pipeline for DM that incorporates quantum, electronic, and structural noise components, as well as spatial noise correlation, and to assess its effect on breast radiologists' detection of simulated microcalcification clusters in clinical mammograms. Methods A total of 120 clinical DM images acquired using a GE Senographe Pristina system were used. Simulated microcalcification clusters were inserted into locally dense fibroglandular regions at controlled contrast levels. The restoration pipeline was configured to target the signal and noise characteristics predicted by the model for an acquisition at approximately twice the original radiation dose. Six breast radiologists with different levels of experience, classified as Junior or Senior, participated in two human observer experiments: (i) a pilot study using original clinical images with simulated clusters and a four‐alternative forced‐choice (4AFC) protocol combined with a three‐down‐one‐up (3D1U) staircase to estimate reader‐specific contrast detection thresholds and achieve comparable baseline task difficulty across readers; and (ii) a main 4AFC observer study using individualized datasets generated according to each reader's estimated contrast threshold, in which microcalcification detection performance was compared between the original and restored image conditions based on the percentage of correct responses (PC). Two generalized linear mixed models (GLMMs) were used: the first evaluated the effect of image condition as a fixed effect, with reader and case as crossed random intercepts, and the second evaluated the effect of reader experience on detection performance in the original image condition. Results At the reader‐specific contrast thresholds estimated in the pilot study, the mean PC for the original images was 78.3%, within the prespecified target performance range. Mean PC increased to 84.3% for the restored images, corresponding to an absolute improvement of 6.0 percentage points. The primary GLMM indicated a significant restoration effect on detection performance (OR = 1.57, 95% CI: 1.15–2.14, ). Baseline mean PC was 78.7% for Junior readers and 78.0% for Senior readers, and the separate GLMM found no evidence of an experience‐group effect under the original image condition (). Conclusions The proposed extended MB restoration pipeline improved the detectability of simulated microcalcification clusters by breast radiologists under the controlled 4AFC conditions investigated. By targeting the signal and noise characteristics predicted by the model for an acquisition at approximately twice the original radiation dose without requiring additional radiation exposure, the method represents a promising approach for improving microcalcification detectability in DM.

Medical PhysicsVol. 53(11)
Tampere University of Applied Sciences (FI), Tampere University (FI), Universidade de São Paulo (BR), Malmö University (SE), Lund University (SE), Real Time Tomography (United States) (US), Hospital de Câncer de Barretos (BR), Tampere University (FI), Skåne University Hospital (SE), University of Tampere (FI), University of Pennsylvania (US)
Fundação de Amparo à Pesquisa do Estado de São Paulo, Academy of Finland, Region Skåne, Siemens Healthineers
Openalex Percentile: Top 12%
Digital Radiography and Breast Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon